header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 39 - 39
1 Mar 2021
Jimenez-Cruz D Masterson A Dubey M Board T Williams S
Full Access

Abstract

OBJECTIVES

Abnormal joint mechanics have been proposed as adversely affecting natural hip joint tribology, whereby increased stress on the articular cartilage from abnormal loading leads to joint degeneration. The aim of this project was to assess the damage caused by different loading conditions on the articular surfaces of the porcine hip joint in an experimental simulator.

METHODS

Porcine hip joints were dissected and mounted in a single station hip simulator (SimSol, UK) and tested under loading scenarios (that corresponded to equivalent of different body mass index's’ (BMI) in humans), as follows:“Normal” (n=4), the loading cycle consisted of a simplified gait cycle based on a scaled version of a simplified twin-peak human gait cycle, the peak load was 900N (representative of a healthy BMI). Representative of an “Overweight” BMI (n=3), as the normal cycle with a peak load of 1,130N Representative of an “Obese” BMI (n=1), as the normal cycle with a peak load of 1,340N Tests were conducted at 1Hz for 14,400 cycles in Ringers solution; photogrammetry was used to characterise the appearance of the cartilage and labrum pre, during and post simulation. the appearance and location of damage was recorded.