Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 123 - 123
1 Mar 2009
MASSIN P HEIZMANN J PROVE S PONTHIEUX A
Full Access

Introduction. Using conventional tests, manual computer-assisted methods offer a satisfactory reproducibility and accuracy in wear evaluation in knee arthroplasty. However, repeatability may be underestimated. Various calibration methods were proposed but not tested. Accuracy may decrease with fluoroscopic images.

Method. We recalculated intra- and interobserver variability based on 132 fluoroscopic images of 3 unimplanted specimens, using specific analyses like the Brand and Altman test or the repeatability coefficient. Calibration was tested in comparison to a reference method (based on a circular shape of known diameter). The influence of a radiological marker embedded into the central part of the polyethylene to control the incidence of fluoroscopic guided images was tested. Accuracy was compared between fluoroscopic and digitized plain radiographs.

Results. The coefficient of repeatability indicated a variation of 0.4 mm in repeated measurements by the same observer. Although a high interobserver reproducibility (r=0.99), the Brand-Altman test showed inconsistent values in about 30% of cases with both observers. The method of calibration influenced the results significantly. Under ideal conditions of radiographic incidence, the accuracy was better with a digitized x-ray (0.10 mm) than with fluoroscopic images (0.25 mm), and compared favourably with fully automatized methods. If small variations of radiographic incidence occured, the accuracy droped to 1 mm. A comparison of successive radiographs of the same implant showed that 1° tilting induced an average reduction of 0.36 mm in the apparent thickness of the polyethylene.

Discussion. When comparing wear measurements in TKA using manual, computer-assisted methods, examination by a single observer should be preferred. The difference between the measurements of the thickness of the polyethylene insert on two images taken with exactly the same incidence should be considered significant if it is greater than twice the precision of the method, i.e. greater than 0.20 mm on digitized X-ray images, or greater than 0.50 mm on fluoroscopic images. In the absence of a strict control of the radiographic incidence, it appears that the difference between the measurements of the thickness of the polyethylene on two images should be considered significant only if it is greater than 2 mm. Given the estimated wear rate of conventional TKR, 10 years of use of the prosthesis would be needed before such systems can detect wear with certainty.

Conclusion. A metallic marker of known dimension embedded in the central part of the polyethylene insert appears mandatory to control calibration as well as the angle of incidence of fluoroscopically guided radiographs. This would improve the performance of radiographic measurements enough to offer clinical relevance.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 121 - 121
1 Mar 2006
Prové S Heizmann J Laumonier F
Full Access

This retrospective study of 196 cases of children’s elbow dislocation had for objectives to check literature’s epidemiological data and to compare the various therapeutic protocols used over a 32 years period.

The average age was 11 years, boys (117) were more frequently affected than girls (79). Both sides were almost equally affected. Most dislocations were posterior or postero-lateral (176). An associated fracture has been found in 104 cases, in 64 cases it was a medial epicondyle’s fracture. Nervous or vascular complications have been found in 8 cases and have immediately been reversed following reduction of the dislocation.

153 patients have been reviewed, 134 patients reported good results, 12 results were average and 7 results were poor. Radiological calcifications have been found in 37 cases (24%). We have shown significant statistical link between the occurrence of elbow stiffness and 2 factors: the carrying out of late kinesitherapy and the appearance of posttraumatic calcifications. We haven’t found any factor favoring the appearance of calcification.

This large series has confirmed the epidemiological data found in literature and allowed us to initiate a simple therapeutic stance combining an early reduction under general anesthesia, the carrying out of an osteosynthesis each time it is necessary and a plastered immobilization of 4 weeks followed by a self rehabilitation.