Vol. 86-B, Issue SUPP_III | Pages 323 - 324
1 Mar 2004
Campbell D Muthusamy K Sturdee S Stone M Finlayson D
Full Access

Aims: To review the effectiveness of the Posterior Lip Augmentation Device (P.L.A.D.) in treating recurrent prosthetic total hip dislocation. Method: We reviewed 36 patients from 2 hip revision surgeons from 2 UK centres who underwent cup augmentation using this device since October 1995. Data were collected from case notes, X-rays and clinical review. All dislocations were posterior (minimum of 3 dislocations). In the majority of cases the abductors were poor or detached, and 35 of the 36 patients had undergone at least one major operation on their hip before a P.L.A.D. was þtted (mean 2.2 operations). Results: The mean age at the time of P.L.A.D. was 73 years (range 47–94). The longest P.L.A.D. follow up is 72 months, and is working well without problems reported by the patient or visible changes on x-ray. A total of 26 out of 36 patients (72.2%) have had a successful correction of their dislocation over this mean period of 21 months (range 8 days Ð 72 months). 10 patients (27.8%) developed signiþcant problems of which 7 (19.4%) required removal of the implant. There were 4 unrelated deaths in our cohort. Conclusions: We feel it has a place in the treatment of recurrent dislocation, particularly in the frail patient, with well-þxed and orientated components, where risks of revision surgery are high and there will be less loading postoperatively


Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_II | Pages 154 - 154
1 Jul 2002
Campbell D Muthusamy K Sturdee S Finlayson D Stone M
Full Access

This study reports the outcome of using the Posterior Lip Augmentation Device (PLAD) for recurrent dislocation of total hip replacement. Twenty-seven patients (16 in Inverness, 11 in Leeds) were treated with the device. The indication for its use was recurrent dislocation of the hip in a patient who had a well-orientated and well-fixed cemented acetabular component. The patients had had between 2 and 14 dislocations before using the PLAD. 24 of the 27 patients had satisfactory control of the dislocations after surgery. The operation was found to be straightforward in most cases with few complications, although one case in the Inverness series has a partial sciatic nerve palsy which is recovering. Of the three failures, two had recurrent dislocation after the use of the PLAD. In one of these, in whom the dislocation followed acetabular reconstruction with impaction grafting, the cup also pulled out due to the semi-captive state after use of the PLAD. Both of these cases were treated by a Girdlestone excision arthroplasty. There was one final failure in whom there has been no further dislocation but following an abduction injury some four months after surgery, progressive cup loosening developed and he awaits revision of the cup. One patient complains of an occasional click in the hip joint and a broken screw is visible on x-ray but there has been no loss of position of the device and no further evidence of dislocation. Overall, we have found that this device gives satisfactory control of recurrent dislocations of the hip but the failures have stressed the need to ensure that the cup fixation is sound before inserting the device.