Osteoarthritis (OA), a painful, debilitating joint disease, often caused by excessive joint stress, is a leading cause of disability (World Health Organisation, 2003) and increases with age and obesity. A 5° varus malalignment increases loading in the medial knee compartment from 70% to 90% (Tetsworth and Paley, 1994). Internal unloading implants, placed subcutaneously upon the medial aspect of the knee joint, are designed to offload the medial compartment of the knee without violating natural joint tissues. The To simulate surgical treatment of medial knee OA, a three-dimensional computer-aided design of an Atlas™ knee system was virtually fixed to the medial aspect of a validated finite element knee model (Mootanah, 2014), using CATIA v5 software (Dassault Systèmes, Velizy Villacoublay, France). The construct was meshed and assigned material properties and boundary conditions, using Abaqus finite element software (Dassault Systèmes, Velizy Villacoublay, France). A cartilage defect was simulated by removing elements corresponding to 4.7 mm2. The international cartilage repair society (ICRS) Grade II and III damage were simulated by normalized defect depth of 33% and 67%, respectively. The femur was mechanically grounded and the tibia was subjected to loading conditions corresponding to the stance phase of walking of a healthy 50-year-old 68-Kg male with anthropometrics that matched those of the cadaver. Finite element analyses were run for peak shear and von Mises stress in the medial and lateral tibiofemoral compartments.Introduction
Methods