Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 94 - 94
23 Feb 2023
Grupp T Schierjott R Pfaff A Tozzi G Schwiesau J Giurea A
Full Access

Total knee arthroplasty with a rotating hinge knee with carbon-fibre-reinforced (CFR)-PEEK as an alternative bushing material with enhanced creep, wear and fatigue behaviour has been clinically established [1–4]. The objective of our study was to compare results from in vitro biotribological characterisation to ex vivo findings on a retrievals.

A modified in vitro wear simulation based on ISO 14243-1 was performed for 5 million cycles on rotating hinge knee (RHK) designs (EnduRo®) out of cobalt-chromium and ZrN-multilayer ceramic coating. The rotational & flexion axles-bushings and the flanges are made of CFR-PEEK with 30% polyacrylonitrile fibre content.

Analysis of 12 retrieved EnduRo® RHK systems in cobalt-chromium and ZrN-multilayer in regard to loosening torques, microscopic surface analysis, distinction between different wear modes and classification with a modified HOOD-score has been performed.

For the RHK design with the polyethylene gliding surface and bushings and flanges made out of CFR-PEEK, a cumulative volumetric wear was measured to be 12.9±3.95 mm3 in articulation to cobalt-chromium and 1.3±0.21 mm3 to ZrN-multilayer coating - a significant 9.9-fold decrease (p=0.0072).

For the CFR-PEEK flexion bushing and flanges the volumetric wear rates were 2.3±0.48 mm3/million cycles (cobalt-chromium) and 0.21±0.02 mm3/million cycles (ZrN-multilayer) (p=0.0016). The 5 million cycles of in vitro wear testing reflect a mean in vivo service life of 2.9 years, which is in accordance to the time in vivo of 12-60 months of the retrieved RHK components [5]. The main wear modes were comparable between retrievals and in vitro specimens, whereby the size of affected area on the retrieved components showed a higher variation.

For the EnduRo® RHK design the findings on retrieved implants demonstrate the high suitability of CFR-PEEK as a biomaterial for highly loaded bearings, such as RHK bushings and flanges in articulation to cobalt-chromium and to a ZrN-multilayer coating.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 14 - 14
10 Feb 2023
Vertesich K Staats K Böhler C Koza R Lass R Giurea A
Full Access

The use of rotating hinge (RH) prostheses for severe primary as well as revision arthroplasty is widely established. Aim of this study was to investigate long term results of a new RH prosthesis (EnduRo®, B Braun, Germany), which uses carbon-fiber reinforced poly-ether-ether-ketone (CFR PEEK) as a new bearing material, first time used in knee arthroplasty.

Fifty-six consecutive patients, who received the EnduRo® RH prosthesis were included in this prospective study: 21 patients (37.5%) received the prosthesis as a primary total knee arthroplasty (TKA) and 35 patients (62.5%) underwent revision total knee arthroplasties (rTKA). Clinical and radiographic examinations were performed preoperatively as well as postoperatively after 3 and 12 months and annually thereafter. Min. Follow up was 7 and mean follow up 9,3 years. Clinical examination included Knee Society Score (KSS), Western Ontario and McMaster Osteoarthritis Index (WOMAC), Oxford Knee Score (OKS), and range of motion (ROM). Competing risk analysis was assessed for survival with respect to indication and failure mode.

KSS, WOMAC, OKS, and ROM significantly improved from the preoperative to the follow up investigations (p < 0.0001). There was no difference in clinical outcome between the primary and the revision group. The overall cumulative incidence for revision for any reason was 23.6% and the cumulative incidence for complications associated with failure of the prothesis was 5.6% at 7 years, respectively. Complications occurred more frequently in the revision group (p = 0.002).

The evaluated RH prosthesis provided reliable and durable results with a minimum follow-up of 7 years. Prosthesis survival was successful considering the complexity of cases. The use of this RH system in primary patients showed high survival rates. Long-term functional and clinical results proved to be satisfying in both revision and primary cases. No adverse events were associated with the new bearing material CFR-PEEK.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 13 - 13
10 Feb 2023
Giurea A Fraberger G Kolbitsch P Lass R Kubista B Windhager R
Full Access

Ten to twenty percent of patients are dissatisfied with the clinical result after total knee arthroplasty (TKA). Aim of this study was to investigate the impact of personality traits on patient satisfaction and subjective outcome of TKA.

We investigated 80 patients with 86 computer navigated TKAs (Emotion®, B Braun Aesculap) and asked for patient satisfaction. We divided patients into two groups (satisfied or dissatisfied). 12 personality traits were tested by an independent psychologist, using the Freiburg Personality Inventory (FPI-R). Postoperative examination included Knee Society Score (KSS), Western Ontario and McMaster University Osteoarthritis Index (WOMAC) and the Visual Analogue Scale (VAS). Radiologic investigation was done in all patients.

84% of our patients were satisfied, while 16% were not satisfied with clinical outcome. The FPI-R showed statistically significant influence of four personality traits on patient satisfaction: life satisfaction (ρ = 0.006), performance orientation (ρ =0.015), somatic distress (ρ = 0.001), and emotional stability (ρ = 0.002). All clinical scores (VAS, WOMAC, and KSS) showed significant better results in the satisfied patient group. Radiological examination showed optimal alignment of all TKAs. There were no complications requiring revision surgery in both groups.

The results of our study show that personality traits may influence patient satisfaction and clinical outcome after TKA. Thus, patients personality traits may be a useful predictive factor for postoperative satisfaction after TKA.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_15 | Pages 35 - 35
1 Dec 2021
Sigmund IK Holinka J Sevelda F Staats K Lass R Kubista B Giurea A Windhager R
Full Access

Aim

Although established serum inflammatory biomarkers, such as serum C-reactive protein (CRP) and serum white blood cell count (WBC), showed low accuracies in the literature, they are still commonly used in diagnosing periprosthetic joint infections (PJI). For a sufficient preoperative diagnosis novel more accurate serum parameters are needed. The aim of our study was to evaluate the performances of the established and novel routinely available serum parameters in diagnosing periprosthetic joint infections when using the proposed European Bone and Joint Infection Society (pEBJIS) criteria.

Method

In this retrospective study, 177 patients with an indicated revision surgery after a total joint replacement were included from 2015 to 2019. The easily accessible and routinely available serum parameters CRP, WBC, the percentage of neutrophils (%N), the neutrophils to lymphocytes ratio (NLR), fibrinogen and the platelet count to mean platelet volume ratio (PC/mPV) were evaluated preoperatively. The performances were examined via receiver operating characteristic (ROC) curve analysis (AUC). The curves were compared using the z-test. Seventy-five cases (42%) showed a PJI based on the pEBJIS-criteria.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 50 - 50
1 Feb 2021
Sanchez E Schilling C Grupp T Giurea A Verdonschot N Janssen D
Full Access

Introduction

Cementless total knee arthroplasty (TKA) implants use an interference fit to achieve fixation, which depends on the difference between the inner dimensions of the implant and outer dimensions of the bone. However, the most optimal interference fit is still unclear. A higher interference fit could lead to a superior fixation, but it could also cause bone abrasion and permanent deformation during implantation. Therefore, this study aims to investigate the effect of increasing the interference fit from 350 µm to 700 µm on the primary stability of cementless tibial implants by measuring micromotions and gaps at the bone-implant interface when subjected to two loading conditions.

Methods

Two cementless e.motion® tibial components (Total Knee System, B. Braun) with different interference fit and surface coating were implanted in six pairs of relatively young human cadaver tibias (47–60 years). The Orthoload peak loads of gait (1960N) and squat (1935N) were applied to the specimens with a custom made load applicator (Figure 1A). The micromotions (shear displacement) and opening/closing gaps (normal displacement) were measured with Digital Image Correlation (DIC) in 6 different regions of interest (ROIs - Figure 1B). Two General Linear Mixed Models (GLMMs) were created with micromotions and interfacial gaps as dependent variables, bone quality, loading conditions, ROIs, and interference fit implants as independent variables, and the cadaver specimens as subject variables.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 58 - 58
1 Jan 2017
Grupp T Schierjott R Pfaff A Tozzi G Schwiesau J Giurea A Utzschneider S
Full Access

Knee arthroplasty with a rotating hinge knee (RHK) prosthesis has become an important clinical treatment option for knee revisions and primary patients with severe varus or valgus deformities and instable ligaments. The rotational axle constraints the anterior-posterior shear and varus-valgus moments, but currently used polyethylene bushings may fail in the mid-term due to insufficient creep and wear resistance of the material. Due to that carbon-fibre-reinforced (CFR) PEEK as an alternativ bushing material with enhanced creep, wear and fatigue behaviour has been introduced in a RHK design [Grupp 2011, Giurea 2014]. The objective of our study was to compare results from the pre-clinical biotribological characterisation to ex vivo findings on a series of retrieved implants.

In vitro wear simulation according to ISO 14243-1 was performed on rotating hinge knee devices (EnduRo® Aesculap, Germany) made out of cobalt-chromium and of a ZrN multi-layer ceramic coating for 5 million cycles. The mobile gliding surfaces were made out of polyethylene (GUR 1020, β-irradiated 30 ± 2 kGy). For the bushings of the rotational and flexion axles and the flanges a new bearing material based on CFR-PEEK with 30% PAN fiber content was used.

Analysis of 12 retrieved EnduRo® RHK systems in cobalt-chromium and ZrN multi-layer in regard to

loosening torques in comparison with initial fastening torques

Optical, DSLR camera and stereo light microscope analysis

distinction between different wear modes and classification with a modified HOOD-score

SEM & EDX of representative samples

surface roughness and depth profilometry

with a focus on the four CFR-PEEK components integrated in the EnduRo® RHK design.

For the rotating hinge knee design with flexion bushing and flanges out of CFR-PEEK the volumetric wear rates were 2.3 ± 0.48 mm3/million cycles (cobalt-chromium) and 0.21 ± 0.02 mm3/million cycles (ZrN multi-layer), a 10.9-fold reduction (p = 0.0016). The UHMWPE and CFR-PEEK particles were comparable in size and morphology and predominantly in submicron size [5]. The biological response to representative sub-micron sized CFR-PEEK particles has been demonstrated in vivo based on the leucoyte-endothelian-cell interactions in the synovia of a murine intra-articular knee model by Utzschneider 2010. Schwiesau 2013 extracted the frequency of daily activities in hip and knee replacement patients from literature and estimated an average of 1.76 million gait cycles per year. Thus, the 5 million cycles of in vitro wear testing reflect a mean in vivo service life of 2.9 years, which fits to the time in vivo of 12–60 months of the retrieved RHK devices. The in vitro surface articulation pattern of the wear simulation tests are comparable to findings on retrieved CFR-PEEK components for both types of articulations – cobalt-chromium and ZrN multi-layer coating.

For the rotating hinge knee design the findings on retrieved implants demonstrate the high suitability of CFR-PEEK as a biomaterial for highly loaded bearings, such as RHK bushings and flanges in articulation to cobalt-chromium and to a ZrN multi-layer coating.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 772 - 779
1 Jun 2016
Stihsen C Hipfl C Kubista B Funovics PT Dominkus M Giurea A Windhager R

Aim

Until now, there has been no consensus as to whether stemmed acetabular components are appropriate for use in patients undergoing revision total hip arthroplasty (THA) who have major acetabular defects or pelvic discontinuity. We wished to address this deficiency in the literature.

Patients and Methods

We carried out a retrospective study of 35 patients (six men and 29 women) with a mean age of 68 years (37 to 87), with major acetabular defects who underwent revision THA between 2000 and 2012.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 98 - 98
1 May 2016
Utzschneider S Paulus A Hasselt S Jansson V Giurea A Neuhaus H Grupp T
Full Access

Introduction

The complex process of inflammation and osteolysis due to wear particles still is not understood in detail. So far, Ultra-high-molecular-weight-polyethylene (UHMWPE) is the bearing material of choice in knee arthroplasty and revision knee arthroplasty, but there is a growing demand for alternative bearing materials with improved wear properties. Lately, increasing interest developed in the use of natural and carbon-fiber-reinforced-poly-ether-ether-ketones (CFR-PEEK).

While there is a lack of data concerning the effects of CFR-PEEK particles on human tissue, the effects of such wear debris in vitro and in animal studies is controversially discussed.

The aim of this study was to analyze human tissue containing CFR-PEEK as well as UHMWPE wear debris.

The authors hypothesized no difference between the used biomaterials because of similar size parameters of the wear particles in a prior knee simulator study of this implant.

Methods and Materials

Synovial tissue samples of 10 patients while knee revision surgery of a rotating hinge knee implant design (Enduro®, Aesculap, Germany) were achieved. The tibial inserts of this design were made from UHMWPE (GUR 1020), whereas the bushings and flanges are made of CFR-PEEK containing 30% polyacrylonitrile (PAN) based carbon fibers (PEEK-Optima LT1, Invibio Ltd. Thornton-Cleveleys, UK). In a prior in vitro test most of the released CFR-PEEK particles were in a size range between 0.1 and 2μm. The implant survival until revision surgery was 22 (2.5–48 min.-max.) months. As a control synovial tissue out of a patient also got knee revision surgery without any PEEK components.

The tissue was fixed with 4% paraformaldehyde, embedded in paraffin, sliced into 2 µm thick sections

stained with hematoxylin and eosin in a standard process. A modified panoptical staining (preincubation in propylenglycol; >3h; 35°C) was also done which stained the UHMWPE particles turquoise.

The study was approved by the ethics committee of the local university.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 138 - 138
1 May 2016
Lass R Kubista B Olischar B Frantal S Windhager R Giurea A
Full Access

Material and Methods

In a prospective randomized study of two groups of 65 patients each, we compared the acetabular component position when usingthe imageless navigation system compared to the freehand conventional technique for cementless total hip arthroplasty. The position of the component was determined postoperatively on computed tomographic scans of the pelvis.

Results

There was no significant difference for postoperative mean inclination (p=0.29), but a significant difference for mean postoperativ acetabular component anteversion (p=0.007), for mean deviation of the postoperative anteversion from the target position of 15°(p=0.02) and for the outliers regarding inclination (p=0.02) and anteversion (p<0.05) between the computer-assisted and the freehand-placement group.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 137 - 137
1 May 2016
Lass R Gruebl A Kolb A Stelzeneder D Pilger A Kubista B Giurea A Windhager R
Full Access

Introduction

In a recent study we evaluated the clinical and radiographic long-term results as well as the serum metal concentrations of 105 cementless primary total hip prosthesis, performed between November 1992 and May 1994 with a 28-mm high-carbide-concentration metal-on-metal articulating surfaces. Forty-one patients who had had a total of forty-four arthroplasties were available for follow-up evaluation at a minimum of seventeen years postoperatively.

The median serum cobalt concentration of the patients with their hip replacement as the only source of cobalt was 0.7 µg/L (range 0.4–5.1µg/L), showing no significant difference to the previous study after a minimum of 10 years follow-up. We were investigating the systemic dissemination, which in turn, did not show more severe effects, such as carcinogenicity or renal failure. There are many complex issues associated with the analysis of metal ions, including collecting technique, analysis and reporting of the results. At the AAOS in March 2013, the Hip Society mentioned, that systemic ion levels are just one factor in the evaluation and should not be relied upon solely to determine the need for revision surgery. Furthermore, the correlation between cobalt or chromium serum, urin or synovial fluid levels and adverse local tissue reactions is incompletely understood.

Patients and Methods

In our present study we evaluated the serum, urin as well as the joint aspirate metal concentrations, of cementless total hip arthroplasties with a high-carbon, metal-on-metal bearing (Metasul®) at a mean of eighteen-years follow-up. We performed a correlation analysis to evaluate the relationship between these values and to determine whether elevated serum metal concentrations are associated with elevated and local metal concentrations and with early failure of metal-on-metal articulations.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 536 - 536
1 Oct 2010
Giurea A Holinka J Jenny J Kotz R Kubista B Lass R Miehlke R Pfeiffer M
Full Access

Introduction: Total knee arthroplasty in obese patients remains a challenge to most surgeons. Surgical complication rates as well as perioperative morbidity are higher than total knee arthroplasty in the nonobese. The purpose of this paper is to review our experience with total knee arthroplasty in superobese patients (BMI> 50).

Methods: From 1998–2005, 84 patients underwent 148 knee arthroplasties. Sixty-four patients underwent simultaneous bilateral total knee arthroplasties and 20 patients underwent unilateral knee arthroplasties. They were compared with similar group of nonobese patients who underwent knee arthroplasties during the same time period. All patients received combined regional and general anesthesia.

Results: Mean follow-up was 3.8 years (2–7). Knee society scores improved by 36 points in the superobese (pre-op 47 to 83 post-op) and by 45 points in the non-obese (pre-op 47 to 93 post-op) (p< .05). There was a greater incidence of complications in the superobese group, namely superficial wound infections and deep vein thrombosis. There was late loosening in three tibial components and instability in two patients that required revision in the superobese group. No reoperations in the nonobese group.

Conclusion: Although total knee arthroplasty may be safely performed in the superobese, it may be complicated by infection, loosening, instability, and lower knee scores.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 390 - 391
1 Sep 2009
Jenny J Miehlke R Giurea A
Full Access

Accuracy of implantation is an accepted prognostic factor for the long term survival of total knee replacement (TKR). The use of navigation demonstrated a significant higher accuracy of implant orientation in comparison to conventional methods. However, these systems are often thought to be technically demanding, to increase operating time and to involve a long learning curve. We performed a prospective, multicenter study to compare the accuracy of implantation of a TKR measured on post-operative X-rays in experienced and less experienced centers.

All centers used the same navigation system (Ortho-Pilot ®, Asculap, Tuttlingen, FRG): 4 had already a significant experience with it (group A – 182 cases), 9 centers were considered as beginners with less than 10 cases performed prior to the study (group B – 221 cases). Accuracy of implantation was measured on post-operative antero-posterior and lateral long leg X-rays with five items: mechanical femoro-tibial angle, coronal orientation of the femoral component, sagittal orientation of the femoral component, coronal orientation of the tibial component, sagittal orientation of the tibial component.

When the measured angle was in the expected range, one point was given. The accuracy note was defined as the sum of all points given for each patient, with a maximum of 5 points (all items fulfilled) and a minimum of 0 point (no item fulfilled). The mean accuracy note was compared in the two groups by a Student t-test at a 0.05 level of significance. Power of the study was 0.80.

There were no significant differences in pre-operative parameters between the two groups, except for the clinical KSS. The mean operative time was significantly longer in group B than in group A (110 minutes vs 90 minutes, p=0.01). However this difference occurred mainly during the first twenty cases in the beginner centres where we observed a clear tendency to achieve the same operative time as the experienced centres at the end of the study. The mean accuracy note was 4.3 ± 0.8 (range, 1 to 5) in the control group and 4.3 ± 0.9 (range, 1 to 5) in the study group (p > 0.05). The power of the study to detect a 0.25 point difference in the post-operative accuracy note was retrospectively calculated to be 0.80. There were no significant differences between the two groups for all individual radiographic items.

This study is, to our knowledge, the first one which investigates the learning curve of navigated TKR The used navigation system allowed a very accurate implantation of a TKR in both experienced and less experienced centers. There was no detectable learning curve with respect to accuracy of TKR implantation, clinical outcome and complication rate. The duration of the learning curve when considering the operating time was 30 cases.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 285 - 285
1 Jul 2008
JENNY J MIEHLKE RK GIUREA A
Full Access

Purpose of the study: Navigation systems have proven efficacy for implantation of total knee arthroplasty (TKA). Navigations have been accused of being complex, requiring a long learning curve. We compared the results obtained with the same navigation system in centers with experienced operators and centers with new operators.

Material and methods: Thirteen European centers participated in this prospective consecutive study. Inclusion criteria was indication for a TKA using a gliding prostheis with preservation of the posterior cruciate ligament. Four experienced cents(group A) with a mean experience of four years, and nine new centers (group B) with no prior experience participated in the study. The study concerned 403 TKA (182 in group A and 221 in group B). The main indications were primarily lateralized osteoarthritis. The navigation system was an imageless system based on intaoperative kinematic anatomic and kinematic analysis. A mobile plateau prosthesis was inserted. The following items were compared between the two groups: overall operative time and its variation over time, postoperative HKA, orientation of the femoral and tibial components in the ap and lateral views, complications and revisions.

Results: No significant difference was observed between the two groups for the preoperative items so comparison between the groups was licit. Correction of the frontal mechanical axis was satisfactory in 90% of patients in group A and 88% in group B (p> 0.05). There was no difference between the groups in quality of implantation for each prosthetic element on the ap and lateral views. There was no difference for rate of complications or reoperations. Longer operative time in group B disappeared after 15 implantations.

Discussion: The results from centers using navigation systems for prosthetic implantations shows that the performance in centers starting use is the same as in experienced centers. The only difference is an operative time slightly longer for the first 15 cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1003 - 1005
1 Aug 2006
Grübl A Weissinger M Brodner W Gleiss A Giurea A Gruber M Pöll G Meisinger V Gottsauner-Wolf F Kotz R

In a randomised study, 28 patients with a mean age of 62.2 years (32 to 81) with osteoarthritis or avascular necrosis of the hip received either a ceramic-on-ceramic or a metal-on-metal total hip replacement. Apart from the liners the acetabular and femoral components were made of Ti-Al-Nb alloy. The serum aluminium and cobalt levels were measured before, and at one year after surgery.

The 15 patients in the ceramic-on-ceramic group had a median pre-operative aluminium level of 1.3 μg/l (0.25 to 8.4) and a cobalt level below the detection limit. At one year the aluminium level was 1.1 μg/l (0.25 to 2.3) and the cobalt level was 0.4 μg/l (0.15 to 0.7). The 13 patients in the metal-on-metal group had a median pre-operative aluminium level of 1.9 μg/l (0.25 to 4.4) and a cobalt level below the detection limit. At one year the median aluminium level was 0.9 μg/l (0.25 to 3.9) whereas the cobalt level was 1.4 μg/l (0.5 to 10.5). This increase in the cobalt level at one year was significant (p < 0.001).

Our findings indicate that ceramic-on-ceramic bearings do not cause elevated levels of serum aluminium in the first post-operative year.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 294 - 294
1 Mar 2004
Zehetgruber H Wurnig C Gruebl A Giurea A
Full Access

Aims: Since several authors query the beneþt of surgical repair of rotator cuff tears, especially in the elderly patients, the purpose of our study was to observe patients with asymptomatic rotator cuff tears. Methods: A series of 235 patients were enrolled clinically during the time of hospitalization, additional ultrasound and X-ray examination of the shoulder was done. In 14 patients ultrasound revealed a clinical asymptomatic cuff tear, afþrmed by MR examination. All 14 patients were observed for 5 years with regular clinical and sonographic examination. Additional MRI was done at 5 year follow-up. Results: At the time of the last follow-up, only one patient developed speciþc symptoms, such as pain, nocturnal pain and weakness due to the cuff tear. Operative treatment for rotator cuff repair was scheduled. Active range of motion deceased from 180û to 167û in ßexion and from 178û to 165û in abduction. Also range of external and internal rotation decreased. Radiological aggravation of cuff arthropathy was observed in all cases. Conclusion: Results of this study suggest that a rotator cuff tear can remain asymptomatic for years, although radiological signs of osteoarthrosis increases.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 284 - 287
1 Mar 1998
Giurea A Paternostro T Heinz-Peer G Kaider A Gottsauner-Wolf F

We compared two methods of reconstruction of the abductor mechanism in 15 patients after prosthetic replacement of the upper femur, to assess abductor strength and function.

Six patients in group I had direct fixation of the gluteus medius tendon and a segment of the original bone to the prosthesis. Nine patients in group 2 had the abductor tendon fixed to the iliotibial band. We assessed clinical function, isometric muscle strength and muscle cross-sectional area for each patient.

The patients in group 1 had better clinical and functional results (p = 0.059), with average peak torques for hip abduction of 92% of that in the non-operated leg in group 1, and of 57% in group 2. Group 1 had a mean muscle cross-sectional area of 69% and a mean value of strength per cross-sectional area of 134% when compared with the control side. The respective values for group 2 were 52% and 91%.

Direct fixation of the abductor muscles to the prostheses gave improved function and higher isometric abductor muscle force.