Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 27 - 27
1 Apr 2019
Wang L Beedall D Thompson J
Full Access

INTRODUCTION

Component positioning of an artificial hip joint plays a key role in durability of implant. Despite the fact that a number of numerical, experimental and clinical studies have been carried out to investigate the effects of cup inclination on polyethylene wear, steep inclination has been reported to be associated with both high and low volumetric wear. Moreover, how cup anteversion affects wear and its interaction with inclination are still unclear. To address these knowledge gaps, in this study wear and contact mechanics of a hip joint under various cup positioning has been investigated by using FEA (Finite Element Analysis).

METHOD

A Pinnacle® Marathon neutral liner 36×56mm was chosen to model the wear and creep over 3 million cycles (mc) based on the Archard's law and modified time hardening model in ANSYS, respectively. Central composite design of response surface method was used to generate 9 FEA runs, where the operative inclination angles varied from 35º, 45º to 55º and operative anteversion angles differed amongst 0º, 15º and 30º. The range of cup angles were chosen to be similar to the Lewinnek “golden” safe zone for dislocation. The gait cycle as specified in ISO 14242-1 was applied to the femoral head.