header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 28 - 28
1 Apr 2017
Rastetter B Wright S Gheduzzi S Miles A Clift S
Full Access

Background

Finite element (FE) models are frequently used in biomechanics to predict the behaviour of new implant designs. To increase the stability after severe bone loss tibial components with long stems are used in revision total knee replacements (TKR). A clinically reported complication after revision surgery is the occurrence of pain in the stem-end region. The aim of this analysis was the development of a validated FE-model of a fully cemented implant and to evaluate the effect of different tibial stem orientations.

Methods

A scanned 4th generation synthetic left tibia (Sawbones) was used to develop the FE-model with a virtually implanted fully cemented tibial component. The 500 N load was applied with medial:lateral compartment distributions of 60:40 and 80:20. Different stem positons were simulated by modifying the resection surface angle posterior to the tibias shaft axis. The results were compared with an experimental study which used strain gauges on Sawbones tibias with an implanted tibial TKR component. The locations of the experimental strain gauges were modelled in the FE study.