header advert
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 523 - 523
1 Dec 2013
Latta L Sawardeker P Kam C Milne E Ouellette E
Full Access

INTRODUCTION:

Radial head arthroplasty is a reliable procedure with good functional outcomes when faced with irreparable radial head fractures. Since the first attempt at arthroplasty by Speed in 1941, there have been a variety of different designs created for radial head prostheses. There has been considerable recent interest in bipolar radial head components. However, to date, there have been few biomechanical studies comparing bipolar components to their monopolar counterparts. We examine the effects of alteration of axial length of the radial head prosthesis and force conveyed at the radiocapitellar joint in a head-to-head comparison of bipolar implants to monopolar implants.

METHODS:

Sixteen fresh-frozen, sided cadaveric arms were utilized. Radial heads were resected and either a monopolar, rigid, metal radial head prosthesis (Solar, Stryker, Mahwah, NJ) was implanted or a bipolar metal prosthesis used (Katalyst, Integra, Plainsboro, NJ). Adjustments of radial head length were made in 2 mm increments using radiolucent washers to create an understuffed (−2), neutral (0), and overstuffed (+2, +4) effect, see Fig. 1. Forearms were cyclically loaded in compression from 13N to 130N with the forearm in neutral. Radiocapitellar forces were measured using Tekscan (Tekscan, Inc., Boston, MA) pressure sensors with radial head length set at −2 mm, 0, +2 mm and +4 mm and comparisons were made with the neutral (0) radial head, see Fig. 2. Multivariant ANOVA with Tukey's HSD correction was used for statistical analysis.