The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1327 - 1332
1 Nov 2024
Ameztoy Gallego J Diez Sanchez B Vaquero-Picado A Antuña S Barco R

Aims. In patients with a failed radial head arthroplasty (RHA), simple removal of the implant is an option. However, there is little information in the literature about the outcome of this procedure. The aim of this study was to review the mid-term clinical and radiological results, and the rate of complications and removal of the implant, in patients whose initial RHA was undertaken acutely for trauma involving the elbow. Methods. A total of 11 patients in whom removal of a RHA without reimplantation was undertaken as a revision procedure were reviewed at a mean follow-up of 8.4 years (6 to 11). The range of motion (ROM) and stability of the elbow were recorded. Pain was assessed using a visual analogue scale (VAS). The functional outcome was assessed using the Mayo Elbow Performance Score (MEPS), the Oxford Elbow Score (OES), and the Disabilities of the Arm, Shoulder and Hand questionnaire (DASH). Radiological examination included the assessment of heterotopic ossification (HO), implant loosening, capitellar erosion, overlengthening, and osteoarthritis. Complications and the rate of further surgery were also recorded. Results. The indications for removal of the implant were stiffness in five patients, aseptic loosening in five, and pain attributed to the RHA in three. The mean time interval between RHA for trauma to removal was ten months (7 to 21). Preoperatively, three patients had overlengthening of the implant, three had capitellar erosion, six had HO, and four had radiological evidence of loosening. At the final follow-up, the mean the flexion-extension arc improved significantly by 38.2° (95% CI 20 to 59; p = 0.002) and the mean arc of prono-supination improved significantly by 20° (95% CI 0 to 72.5; p = 0.035). The mean pain VAS score improved significantly by 3.5 (95% CI 2 to 5.5; p = 0.004). The mean MEPS improved significantly by 27.5 (95% CI 17.5 to 42.5; p = 0.002). The mean OES improved significantly by 9 (95% CI 2.5 to 14; p = 0.012), and the mean DASH score improved significantly by 23.5 (95% CI 7.5 to 31.6; p = 0.012). Ten patients (91%) had HO and osteoarthritis. Two patients underwent further surgery due to stiffness and pain, respectively. Conclusion. Simple removal of the implant at revision surgery following a failed RHA introduced following trauma provides satisfactory mid-term results with an acceptable risk of complications. Osteoarthritis, instability, and radioulnar impingement were not problems in this series. Cite this article: Bone Joint J 2024;106-B(11):1327–1332


Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims

Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up.

Methods

This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.


Bone & Joint 360
Vol. 13, Issue 5 | Pages 28 - 30
1 Oct 2024

The October 2024 Foot & Ankle Roundup360 looks at: Hemiarthroplasty for hallux rigidus; Fixed or mobile-bearing ankle arthroplasty? A meta-analysis; Bone grafting for periprosthetic bone cysts following total ankle arthroplasty; Diabetic foot ulcer after first-ray amputation; Early motion after ankle surgery: the path to faster recovery?; Are podiatrists and orthopaedic surgeons approaching zone 2 fifth metatarsal fractures in the same way?.


Bone & Joint Open
Vol. 5, Issue 9 | Pages 749 - 757
12 Sep 2024
Hajialiloo Sami S Kargar Shooroki K Ammar W Nahvizadeh S Mohammadi M Dehghani R Toloue B

Aims

The ulna is an extremely rare location for primary bone tumours of the elbow in paediatrics. Although several reconstruction options are available, the optimal reconstruction method is still unknown due to the rarity of proximal ulna tumours. In this study, we report the outcomes of osteoarticular ulna allograft for the reconstruction of proximal ulna tumours.

Methods

Medical profiles of 13 patients, who between March 2004 and November 2021 underwent osteoarticular ulna allograft reconstruction after the resection of the proximal ulna tumour, were retrospectively reviewed. The outcomes were measured clinically by the assessment of elbow range of motion (ROM), stability, and function, and radiologically by the assessment of allograft-host junction union, recurrence, and joint degeneration. The elbow function was assessed objectively by the Musculoskeletal Tumor Society (MSTS) score and subjectively by the Toronto Extremity Salvage Score (TESS) and Mayo Elbow Performance Score (MEPS) questionnaire.


Aims

This study intended to investigate the effect of vericiguat (VIT) on titanium rod osseointegration in aged rats with iron overload, and also explore the role of VIT in osteoblast and osteoclast differentiation.

Methods

In this study, 60 rats were included in a titanium rod implantation model and underwent subsequent guanylate cyclase treatment. Imaging, histology, and biomechanics were used to evaluate the osseointegration of rats in each group. First, the impact of VIT on bone integration in aged rats with iron overload was investigated. Subsequently, VIT was employed to modulate the differentiation of MC3T3-E1 cells and RAW264.7 cells under conditions of iron overload.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 935 - 941
1 Sep 2024
Ailaney N Guirguis PG Ginnetti JG Balkissoon R Myers TG Ramirez G Thirukumaran CP Ricciardi BF

Aims

The purpose of this study was to determine the association between prior sleeve gastrectomy in patients undergoing primary total hip and knee arthroplasty, and 90-day complications, incidence of revision arthroplasty, and patient-reported outcome scores at final follow-up.

Methods

This is a retrospective, single-centre analysis. Patients undergoing primary hip or knee arthroplasty with a prior sleeve gastrectomy were eligible for inclusion (n = 80 patients). A morbidly obese control group was established from the same institutional registry using a 1:2 match, for cases:controls with arthroplasty based on propensity score using age, sex, pre-sleeve gastrectomy BMI, Current Procedural Terminology code to identify anatomical location, and presurgical haemoglobin A1C. Outcomes included 90-day complications, incidence of revision arthroplasty, and patient-reported outcome scores at final follow-up. Multivariable logistic regressions evaluated associations of underlying preoperative demographic and treatment characteristics with outcomes.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 80 - 80
19 Aug 2024
Malchau H
Full Access

The painful hip without obvious clinical or radiographic signs of complications is a well-known scenario for surgeons. The clinical tools we have access to currently lack a dynamic test for detecting early signs of motion between implant and bone. A new software, Sectra IMA, has a potential to facilitate diagnosis of early implant loosening by analysis of paired CT exams. In clinical practise the two scans are acquired by endpoint of a possible motion, “a provocation CT”, for example maximal external and internal rotation in a CT hip examination. 20 years of research by Olivecrona and Weidenhielm is the scientific background for the technique. Early results are presented by Sandberg et al 2022. To further validate and create clinical evidence more extensive clinical studies is needed. A proposal for a multicentric, global clinical study will be presented for further discussion. A new technique for diagnosis of acetabular cup loosening using computed tomography: preliminary experience in 10 patients. Henrik Olivecrona, Lotta Olivecrona, Lars Weidenhielm, Marilyn E Noz, Jocelyn Kardos, Gerald Q Maguire Jr, Michael P Zeleznik, Peter Aspelin. Acta Orthop. 2008 Jun;79(3):346-53. doi: 10.1080/17453670710015247. Inducible displacement CT increases the diagnostic accuracy of aseptic loosening in primary total hip arthroplasty. Sandberg O, Carlsson S, Harbom E, Cappelen V, Tholén S, Olivecrona H, Wretenberg P. Acta Orthop. 2022 Oct 31;93:831-836. doi: 10.2340/17453674.2022.5240.PMID: 36314542


Bone & Joint Open
Vol. 5, Issue 6 | Pages 489 - 498
12 Jun 2024
Kriechling P Bowley ALW Ross LA Moran M Scott CEH

Aims

The purpose of this study was to compare reoperation and revision rates of double plating (DP), single plating using a lateral locking plate (SP), or distal femoral arthroplasty (DFA) for the treatment of periprosthetic distal femur fractures (PDFFs).

Methods

All patients with PDFF primarily treated with DP, SP, or DFA between 2008 and 2022 at a university teaching hospital were included in this retrospective cohort study. The primary outcome was revision surgery for failure following DP, SP, or DFA. Secondary outcome measures included any reoperation, length of hospital stay, and mortality. All basic demographic and relevant implant and injury details were collected. Radiological analysis included fracture classification and evaluation of metaphyseal and medial comminution.


Bone & Joint Research
Vol. 13, Issue 6 | Pages 272 - 278
5 Jun 2024
Niki Y Huber G Behzadi K Morlock MM

Aims. Periprosthetic fracture and implant loosening are two of the major reasons for revision surgery of cementless implants. Optimal implant fixation with minimal bone damage is challenging in this procedure. This pilot study investigates whether vibratory implant insertion is gentler compared to consecutive single blows for acetabular component implantation in a surrogate polyurethane (PU) model. Methods. Acetabular components (cups) were implanted into 1 mm nominal under-sized cavities in PU foams (15 and 30 per cubic foot (PCF)) using a vibratory implant insertion device and an automated impaction device for single blows. The impaction force, remaining polar gap, and lever-out moment were measured and compared between the impaction methods. Results. Impaction force was reduced by 89% and 53% for vibratory insertion in 15 and 30 PCF foams, respectively. Both methods positioned the component with polar gaps under 2 mm in 15 PCF foam. However, in 30 PCF foam, the vibratory insertion resulted in a clinically undesirable polar gap of over 2 mm. A higher lever-out moment was achieved with the consecutive single blow insertion by 42% in 15 PCF and 2.7 times higher in 30 PCF foam. Conclusion. Vibratory implant insertion may lower periprosthetic fracture risk by reducing impaction forces, particularly in low-quality bone. Achieving implant seating using vibratory insertion requires adjustment of the nominal press-fit, especially in denser bone. Further preclinical testing on real bone tissue is necessary to assess whether its viscoelasticity in combination with an adjusted press-fit can compensate for the reduced primary stability after vibratory insertion observed in this study. Cite this article: Bone Joint Res 2024;13(6):272–278


Bone & Joint 360
Vol. 13, Issue 3 | Pages 35 - 36
3 Jun 2024

The June 2024 Spine Roundup360 looks at: Intraoperative navigation increases the projected lifetime cancer risk in patients undergoing surgery for adolescent idiopathic scoliosis; Intrawound vancomycin powder reduces delayed deep surgical site infections following posterior spinal fusion for adolescent idiopathic scoliosis; Characterizing negative online reviews of spine surgeons; Proximal junctional failure after surgical instrumentation in adult spinal deformity: biomechanical assessment of proximal instrumentation stiffness; Nutritional supplementation and wound healing: a randomized controlled trial.


Bone & Joint Open
Vol. 5, Issue 6 | Pages 464 - 478
3 Jun 2024
Boon A Barnett E Culliford L Evans R Frost J Hansen-Kaku Z Hollingworth W Johnson E Judge A Marques EMR Metcalfe A Navvuga P Petrie MJ Pike K Wylde V Whitehouse MR Blom AW Matharu GS

Aims

During total knee replacement (TKR), surgeons can choose whether or not to resurface the patella, with advantages and disadvantages of each approach. Recently, the National Institute for Health and Care Excellence (NICE) recommended always resurfacing the patella, rather than never doing so. NICE found insufficient evidence on selective resurfacing (surgeon’s decision based on intraoperative findings and symptoms) to make recommendations. If effective, selective resurfacing could result in optimal individualized patient care. This protocol describes a randomized controlled trial to evaluate the clinical and cost-effectiveness of primary TKR with always patellar resurfacing compared to selective patellar resurfacing.

Methods

The PAtellar Resurfacing Trial (PART) is a patient- and assessor-blinded multicentre, pragmatic parallel two-arm randomized superiority trial of adults undergoing elective primary TKR for primary osteoarthritis at NHS hospitals in England, with an embedded internal pilot phase (ISRCTN 33276681). Participants will be randomly allocated intraoperatively on a 1:1 basis (stratified by centre and implant type (cruciate-retaining vs cruciate-sacrificing)) to always resurface or selectively resurface the patella, once the surgeon has confirmed sufficient patellar thickness for resurfacing and that constrained implants are not required. The primary analysis will compare the Oxford Knee Score (OKS) one year after surgery. Secondary outcomes include patient-reported outcome measures at three months, six months, and one year (Knee injury and Osteoarthritis Outcome Score, OKS, EuroQol five-dimension five-level questionnaire, patient satisfaction, postoperative complications, need for further surgery, resource use, and costs). Cost-effectiveness will be measured for the lifetime of the patient. Overall, 530 patients will be recruited to obtain 90% power to detect a four-point difference in OKS between the groups one year after surgery, assuming up to 40% resurfacing in the selective group.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_7 | Pages 4 - 4
8 May 2024
Nurm T Ramaskandhan J Nicolas A Siddique M
Full Access

Introduction. Total ankle arthroplasty (TAA) is an increasingly popular treatment option for patients with end-stage ankle arthritis. However, for most implant systems, failure rates of 10–20% have been reported within the first 10 years after primary TAA. Pain is the primary symptom that indicates failure of TAA but cause of it can be difficult to establish. Methods. All patients who underwent a primary TAA at our center were included in the study. The clinical outcomes were studied for patients requiring a further revision procedure following primary TAA. The reasons for revision surgery and outcomes of surgery were analyzed using appropriate inferential statistical tests. Results. Between 2007 and 2018, 42 primary TAA required revisions in 40 patients. There were 25 men (59.5%) and 15 women (35.7%) with mean age of 57.5 years the time of primary TAA. All patients had undergone primary procedure at a mean duration of 3.5 years previously (range: 3 months to 10 years). Of the total revision procedures, 12/40 (30%) of revisions were carried out due to malalignment, 10/42 (23.8%) due to loosening of the implants or bone subsidence, 5/42 procedures (11.9%) following infection, 4/42 (9.5%) due to polyethylene migration, 1/42 (2.3%) due to fracture and 1/42 (2.3%) due to Charcot arthropathy. In 9/42 (21.4%) cases, imaging showed no objective reason for pain. 50% of patients who underwent revision TAA reported 78.5% satisfaction with results of surgery at 2 years follow up post-operatively. Conclusion. Major reasons for revising primary TAR at our centre are mal-alignment, implant loosening / bone subsidence and suspicion of infection and pain. In-spite of undergoing a complex revision surgery, patients report 78.5% satisfaction from outcomes of surgery


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 118 - 124
1 May 2024
Macheras GA Argyrou C Tzefronis D Milaras C Tsivelekas K Tsiamtsouris KG Kateros K Papadakis SA

Aims

Accurate diagnosis of chronic periprosthetic joint infection (PJI) presents a significant challenge for hip surgeons. Preoperative diagnosis is not always easy to establish, making the intraoperative decision-making process crucial in deciding between one- and two-stage revision total hip arthroplasty (THA). Calprotectin is a promising point-of-care novel biomarker that has displayed high accuracy in detecting PJI. We aimed to evaluate the utility of intraoperative calprotectin lateral flow immunoassay (LFI) in THA patients with suspected chronic PJI.

Methods

The study included 48 THAs in 48 patients with a clinical suspicion of PJI, but who did not meet European Bone and Joint Infection Society (EBJIS) PJI criteria preoperatively, out of 105 patients undergoing revision THA at our institution for possible PJI between November 2020 and December 2022. Intraoperatively, synovial fluid calprotectin was measured with LFI. Cases with calprotectin levels ≥ 50 mg/l were considered infected and treated with two-stage revision THA; in negative cases, one-stage revision was performed. At least five tissue cultures were obtained; the implants removed were sent for sonication.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 475 - 481
1 May 2024
Lee M Lee G Lee K

Aims. The purpose of this study was to assess the success rate and functional outcomes of bone grafting for periprosthetic bone cysts following total ankle arthroplasty (TAA). Additionally, we evaluated the rate of graft incorporation and identified associated predisposing factors using CT scan. Methods. We reviewed a total of 37 ankles (34 patients) that had undergone bone grafting for periprosthetic bone cysts. A CT scan was performed one year after bone grafting to check the status of graft incorporation. For accurate analysis of cyst volumes and their postoperative changes, 3D-reconstructed CT scan processed with 3D software was used. For functional outcomes, variables such as the Ankle Osteoarthritis Scale score and the visual analogue scale for pain were measured. Results. Out of 37 ankles, graft incorporation was successful in 30 cases. Among the remaining seven cases, four (10.8%) exhibited cyst re-progression, so secondary bone grafting was needed. After secondary bone grafting, no further progression has been noted, resulting in an overall 91.9% success rate (34 of 37) at a mean follow-up period of 47.5 months (24 to 120). The remaining three cases (8.1%) showed implant loosening, so tibiotalocalcaneal arthrodesis was performed. Functional outcomes were also improved after bone grafting in all variables at the latest follow-up (p < 0.05). The mean incorporation rate of the grafts according to the location of the cysts was 84.8% (55.2% to 96.1%) at the medial malleolus, 65.1% (27.6% to 97.1%) at the tibia, and 81.2% (42.8% to 98.7%) at the talus. Smoking was identified as a significant predisposing factor adversely affecting graft incorporation (p = 0.001). Conclusion. Bone grafting for periprosthetic bone cysts following primary TAA is a reliable procedure with a satisfactory success rate and functional outcomes. Regular follow-up, including CT scan, is important for the detection of cyst re-progression to prevent implant loosening after bone grafting. Cite this article: Bone Joint J 2024;106-B(5):475–481


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 59 - 65
1 May 2024
Liu WKT Cheung A Fu H Chan PK Chiu KY

Aims

Isolated acetabular liner exchange with a highly crosslinked polyethylene (HXLPE) component is an option to address polyethylene wear and osteolysis following total hip arthroplasty (THA) in the presence of a well-fixed acetabular shell. The liner can be fixed either with the original locking mechanism or by being cemented within the acetabular component. Whether the method used for fixation of the HXLPE liner has any bearing on the long-term outcomes is still unclear.

Methods

Data were retrieved for all patients who underwent isolated acetabular component liner exchange surgery with a HXLPE component in our institute between August 2000 and January 2015. Patients were classified according to the fixation method used (original locking mechanism (n = 36) or cemented (n = 50)). Survival and revision rates were compared. A total of 86 revisions were performed and the mean duration of follow-up was 13 years.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 82 - 88
1 May 2024
Villa JM Rajschmir K Hosseinzadeh S Manrique-Succar J Grieco P Higuera-Rueda CA Riesgo AM

Aims

Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation.

Methods

A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707).


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 468 - 474
1 May 2024
d'Amato M Flevas DA Salari P Bornes TD Brenneis M Boettner F Sculco PK Baldini A

Aims

Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis).

Methods

Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 435 - 441
1 May 2024
Angelomenos V Mohaddes M Kärrholm J Malchau H Shareghi B Itayem R

Aims

Refobacin Bone Cement R and Palacos R + G bone cement were introduced to replace the original cement Refobacin Palacos R in 2005. Both cements were assumed to behave in a biomechanically similar fashion to the original cement. The primary aim of this study was to compare the migration of a polished triple-tapered femoral stem fixed with either Refobacin Bone Cement R or Palacos R + G bone cement. Repeated radiostereometric analysis was used to measure migration of the femoral head centre. The secondary aims were evaluation of cement mantle, stem positioning, and patient-reported outcome measures.

Methods

Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 115 - 120
1 Mar 2024
Ricotti RG Flevas DA Sokrab R Vigdorchik JM Mayman DJ Jerabek SA Sculco TP Sculco PK

Aims

Periprosthetic femoral fracture (PPF) is a major complication following total hip arthroplasty (THA). Uncemented femoral components are widely preferred in primary THA, but are associated with higher PPF risk than cemented components. Collared components have reduced PPF rates following uncemented primary THA compared to collarless components, while maintaining similar prosthetic designs. The purpose of this study was to analyze PPF rate between collarless and collared component designs in a consecutive cohort of posterior approach THAs performed by two high-volume surgeons.

Methods

This retrospective series included 1,888 uncemented primary THAs using the posterior approach performed by two surgeons (PKS, JMV) from January 2016 to December 2022. Both surgeons switched from collarless to collared components in mid-2020, which was the only change in surgical practice. Data related to component design, PPF rate, and requirement for revision surgery were collected. A total of 1,123 patients (59.5%) received a collarless femoral component and 765 (40.5%) received a collared component. PPFs were identified using medical records and radiological imaging. Fracture rates between collared and collarless components were analyzed. Power analysis confirmed 80% power of the sample to detect a significant difference in PPF rates, and a Fisher’s exact test was performed to determine an association between collared and collarless component use on PPF rates.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 130 - 136
1 Mar 2024
Morlock M Perka C Melsheimer O Kirschbaum SM

Aims

Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients.

Methods

A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using uncemented acetabular components without reconstruction shells, from the German arthroplasty registry were included in the study. The revision rates three years postoperatively for four types of stem (uncemented, uncemented with collar, uncemented short, and cemented) were compared within four age groups: < 60 years (Young), between 61 and 70 years (Mid-I), between 71 and 80 years (Mid-II), and aged > 80 years (Old). A noninferiority analysis was performed on the most frequently used designs of stem.