Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 44 - 44
19 Aug 2024
Park C Lim S Park Y
Full Access

Periprosthetic femoral fractures (PFFs) remain a major concern following cementless total hip arthroplasty (THA). This study aimed to evaluate the association between different types of cementless tapered stems and the risk of postoperative PFF. A retrospective review of primary THAs performed at a single center from January 2011 to December 2018 included 3,315 hips (2,326 patients). Cementless stems were classified according to their design geometry using the system proposed by Radaelli et al. The incidence of PFF was compared between flat taper porous-coated stems (type A), rectangular taper grit-blasted stems (type B1), and quadrangular taper hydroxyapatite-coated stems (type B2). Multivariate regression analyses were performed to identify independent factors related to PFF. The mean follow-up duration was 61 months (range, 12‒139 months). Overall, 45 (1.4%) postoperative PFFs occurred. The incidence of PFF was significantly higher in type B1 stems than in type A and type B2 stems (1.8 vs. 0.7 vs. 0.7%; P=0.022). Additionally, more surgical treatments (1.7 vs. 0.5 vs. 0.7%; P=0.013) and femoral revisions (1.2 vs. 0.2 vs. 0%; P=0.004) were required for PFF in type B1 stems. After controlling for confounding variables, older age (P<0.001), diagnosis of hip fracture (P<0.001), and use of type B1 stems (P=0.001) were significant factors associated with PFF. Type B1 rectangular taper stems were found to have higher risks for postoperative PFF and PFF requiring surgical management than type A and type B2 stems in THA. Femoral stem geometry should be considered when planning for cementless THA in elderly patients with compromised bone quality


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 548 - 554
1 Jun 2024
Ohyama Y Minoda Y Masuda S Sugama R Ohta Y Nakamura H

Aims

The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group).

Methods

The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.


Bone & Joint 360
Vol. 13, Issue 1 | Pages 13 - 16
1 Feb 2024

The February 2024 Hip & Pelvis Roundup. 360. looks at: Trial of vancomycin and cefazolin as surgical prophylaxis in arthroplasty; Is preoperative posterior femoral neck tilt a risk factor for fixation failure? Cemented versus uncemented hemiarthroplasty for displaced intracapsular fractures of the hip; Periprosthetic fractures in larger hydroxyapatite-coated stems: are collared stems a better alternative for total hip arthroplasty?; Postoperative periprosthetic fracture following hip arthroplasty with a polished taper slip versus composite beam stem; Is oral tranexamic acid as good as intravenous?; Stem design and the risk of early periprosthetic femur fractures following THA in elderly patients; Does powered femoral broaching compromise patient safety in total hip arthroplasty?


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1113 - 1121
14 Sep 2020
Nantha Kumar N Kunutsor SK Fernandez MA Dominguez E Parsons N Costa ML Whitehouse MR

Aims

We conducted a systematic review and meta-analysis to compare the mortality, morbidity, and functional outcomes of cemented versus uncemented hemiarthroplasty in the treatment of intracapsular hip fractures, analyzing contemporary and non-contemporary implants separately.

Methods

PubMed, Medline, EMBASE, CINAHL, and Cochrane Library were searched to 2 February 2020 for randomized controlled trials (RCTs) comparing the primary outcome, mortality, and secondary outcomes of function, quality of life, reoperation, postoperative complications, perioperative outcomes, pain, and length of hospital stay. Relative risks (RRs) and mean differences (with 95% confidence intervals (CIs)) were used as summary association measures.


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 699 - 708
1 Jun 2020
Nieuwenhuijse MJ Vehmeijer SBW Mathijsen NMC Keizer SB

Aims

Short, bone-conserving femoral components are increasingly used in total hip arthroplasty (THA). They are expected to allow tissue-conserving implantation and to render future revision surgery more straightforward but the long-term data on such components is limited. One such component is the global tissue-sparing (GTS) stem. Following the model for stepwise introduction of new orthopaedic implants, we evaluated early implant fixation and clinical outcome of this novel short-stem THA and compared it to that of a component with established good long-term clinical outcome.

Methods

In total, 50 consecutive patients ≤ 70 years old with end-stage symptomatic osteo-arthritis were randomized to receive THA with the GTS stem or the conventional Taperloc stem using the anterior supine intermuscular approach by two experienced hip surgeons in two hospitals in the Netherlands. Primary outcome was implant migration. Patients were followed using routine clinical examination, patient reported outcome using Harris Hip Score (HHS), Hip Disability And Osteoarthritis Outcome Score (HOOS), EuroQol five-dimension questionnaire (EQ5D), and Roentgen Stereophotogrammetric Analysis (RSA) at three, six, 12, and 24 months. This study evaluated the two-year follow-up results.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1192 - 1198
1 Oct 2019
Sköldenberg OG Rysinska AD Chammout G Salemyr M Mukka SS Bodén H Eisler T

Aims

Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results.

Patients and Methods

We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (sd 5), in a double-blinded, noninferiority, randomized controlled trial. The subjects were randomized in a 1:1 ratio to receive a reverse hybrid THA with a cemented component of either argon-gas gamma-sterilized PE component (controls) or VEPE, with identical geometry. The primary endpoint was proximal implant migration of the component at two years postoperatively measured with RSA. Secondary endpoints included total migration of the component, penetration of the femoral head into the component, and patient-reported outcome measurements.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 8 - 8
1 Apr 2019
Wilson C Critchely O Callary S Campbell D
Full Access

Introduction & aims. The magnitude and pattern of acceptable migration in clinically successful cementless stems is not well understood. Radiostereometric analysis (RSA) is a well-recognised method of assessing implant migration. Previous studies have reported long term RSA results for cemented stems. The aim of this study was to assess the long-term migration characteristics of the Corail hydroxyapatite-coated cementless stem at 10 years using RSA. Method. A prospective cohort of 30 patients undergoing primary total hip arthroplasty for primary hip osteoarthritis were enrolled into a study to characterise the migration behaviour of a cementless stem. Tantalum markers were attached to the stem and placed in the bone intraoperatively, allowing for RSA measurements to be taken in vivo. Previous 5-year results have been presented. A total of 14 patients (total 15 hips, one bilateral) with mean age 82 years (range, 69–92 years) underwent repeat long-term RSA radiographs at minimum 10 years post op. The mean time to follow up was 13.9 years (range, 13.3–14.4 years). The RSA radiographs were analysed to assess for implant rotation and translation. Results. None of the 14 patients (15 hips) followed up have been revised. The migration (rotation and translation) at 6 months, 1 year, 2 years, and 6 years has been described previously. No stems had additional subsidence of more than 0.25 mm between 6 months and 6 years. The resultant mean subsidence between 2 years and 6 years was 0.03 mm, which is below the limit measurable by RSA. The long-term, 10-year results, on implant rotation and translation will be presented. Conclusions. The Corail stem exhibited variable subsidence within the first 6 months after which there is persistent stabilisation through to 6 years. 10-year results on long-term migration show no further migration. These results can be used as a gold standard for other uncemented stems


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 34 - 34
1 Aug 2018
García-Rey E García-Cimbrelo E Cordero-Ampuero J
Full Access

We assessed the clinical and radiological outcome of a single uncemented total hip replacement (THR) after twenty years, analysing polyethylene wear and osteolysis. 82 hips implanted between 1992 and 1995 were prospectively evaluated. Mean follow-up was 20.6 years (18–23). A hemispherical porous-coated cup matched to a proximally hydroxyapatite-coated anatomic stem. A 28mm PE liner, sterilised by gamma irradiation in air, was used in all hips. Radiological position, eventual loosening and osteolysis were recorded over time. Penetration of the head into the liner was measured by the Roentgen Monographic Analysis (ROMAN) Tool at 6 weeks, 6 months, one year and yearly thereafter. Six cups were revised due to wear and four cups because of late dislocation. All stems showed osseointegration and all cups appeared radiographically well-fixed. Six unrevised hips showed osteolysis on the acetabular side and two hips on the proximal femur. Creep at one year was 0.30±0.23 mm. Mean total femoral head penetration was 1.23mm at ten years, 1.52mm at 15 years and 1.92mm at 23 years. Overall mean wear was 0.12±0.1 mm/year and 0.09±0.06 mm/year after the creep period. Mean wear was 0.08±0.06 mm/year in hips without osteolysis and 0.14±0.03 mm/year in hips revised or hips with osteolysis (p<0.001). Although continued durable fixation can be observed with porous-coated cups and proximally hydroxyapatite-coated anatomic stems, true wear continues to increase at a constant level over time. PE wear remains as the main reason for osteolysis and revision surgery in uncemented THR after twenty years


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1087 - 1093
1 Aug 2018
Barenius B Inngul C Alagic Z Enocson A

Aims

The aim of this study was to compare the functional and radiological outcomes in patients with a displaced fracture of the hip who were treated with a cemented or a cementless femoral stem.

Patients and Methods

A four-year follow-up of a randomized controlled study included 141 patients who underwent surgery for a displaced femoral neck fracture. Patients were randomized to receive either a cemented (n = 67) or a cementless (n = 74) stem at hemiarthroplasty (HA; n = 83) or total hip arthroplasty (THA; n = 58).


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 60 - 60
1 Apr 2018
Garcia-Rey E Cimbrelo EG
Full Access

Introduction. Durable bone fixation of uncemented porous-coated acetabular cups can be observed at a long-term, however, polyethylene (PE) wear and osteolysis may affect survivorship. Accurate wear measurements correlated with clinical data may offer unique research information of clinical interest about this highly debated issue. Objetive. We assessed the clinical and radiological outcome of a single uncemented total hip replacement (THR) after twenty years analysing polyethylene wear and the appearance of osteolysis. Materials and Methods. 82 hips implanted between 1992 and 1995 were prospectively evaluated with a mean follow-up of 20.6 years (range, 18 to 23). A hemispherical porous-coated acetabular cup matched to a proximally hydroxyapatite-coated anatomic stem and a 28 mm standard PE liner, sterilised by gamma irradiation in air, was used in all hips. Radiological position and the possible appearance of loosening and osteolysis were recorded over time. Penetration of the prosthetic head into the liner was measured by the Roentgen Monographic Analysis (ROMAN) Tool at 6 weeks, 6 months, one year and yearly thereafter. Results. Six cups were revised due to wear and four due to late dislocation. All cups were radiographically well-fixed and all stems showed radiographic ingrowth. Six un-revised hips showed osteolysis on the acetabular side and two on the proximal femoral side. Creep at one year was 0.30 (±0.23) mm. Mean total femoral head penetration was 1.23 mm at 10 years, 1.52 mm at 15 years and 1.92 mm at 23 years. Overall mean wear was 0.12 (± 0.1) mm/year and 0.09 (±0.06) mm/year after the creep period. Mean wear was 0.08 (± 0.06) mm/year in hips without osteolysis and 0.14 (±0.03) mm/year in revised hips or with osteolysis (p<0.001). Conclusions. Although continued durable fixation can be observed with a porous-coated cups and a proximally hydroxyapatite-coated anatomic stem, true wear continues to increase at a constant level over time. PE wear remains as the main reason for revision surgery and osteolysis in uncemented THR after twenty years


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 9 - 9
1 Apr 2018
Garcia-Rey E Carbonell R Cordero J Gomez-Barrena E
Full Access

Introduction. Durable bone fixation of uncemented porous-coated acetabular cups can be observed at a long-term, however, polyethylene (PE) wear and osteolysis may affect survivorship. Accurate wear measurements correlated with clinical data may offer unique research information of clinical interest about this highly debated issue. Objetive. We assessed the clinical and radiological outcome of a single uncemented total hip replacement (THR) system after twenty years analysing polyethylene wear and the appearance of osteolysis. Materials and Methods. 82 hips implanted between 1992 and 1995 were prospectively evaluated. The mean follow-up was 20.6 years (range, 18 to 23). A hemispherical porous-coated acetabular cup matched to a proximally hydroxyapatite-coated anatomic stem and a 28 mm standard PE liner, sterilised by gamma irradiation in air, was used in all hips. Radiological position and the possible appearance of loosening and osteolysis were recorded over time. Penetration of the prosthetic head into the liner was measured by the Roentgen Monographic Analysis (ROMAN) Tool at 6 weeks, 6 months, one year and yearly thereafter. Results. Six cups were revised due to wear and four due to late dislocation. All cups were radiographically well-fixed and all stems showed radiographic ingrowth. Six un-revised hips showed osteolysis on the acetabular side and two on the proximal femoral side. Creep at one year was 0.30 (±0.23) mm. Mean total femoral head penetration was 1.23 mm at 10 years, 1.52 mm at 15 years and 1.92 mm at 23 years. Overall mean wear was 0.12 (± 0.1) mm/year and 0.09 (±0.06) mm/year after the creep period. Mean wear was 0.08 (± 0.06) mm/year in hips without osteolysis and 0.14 (±0.03) mm/year in revised hips or with osteolysis (p<0.001). Conclusions. Although continued durable fixation can be observed with a porous-coated cups and a proximally hydroxyapatite-coated anatomic stem, true wear continues to increase at a constant rate over time. PE wear remains as the main reason for revision surgery and osteolysis in uncemented THR and does not stop after twenty years


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 741 - 748
1 Jun 2017
Lee YK Ha YC Yoo J Jo WL Kim K Koo KH

Aims

We conducted a prospective study of a delta ceramic total hip arthroplasty (THA) to determine the rate of ceramic fracture, to characterise post-operative noise, and to evaluate the mid-term results and survivorship.

Patients and Methods

Between March 2009 and March 2011, 274 patients (310 hips) underwent cementless THA using a delta ceramic femoral head and liner. At each follow-up, clinical and radiological outcomes were recorded. A Kaplan-Meier analysis was undertaken to estimate survival.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 421 - 431
1 Apr 2017
Veldman HD Heyligers IC Grimm B Boymans TAEJ

Aims

Our aim was to prepare a systematic review and meta-analysis to compare the outcomes of cemented and cementless hemiarthroplasty of the hip, in elderly patients with a fracture of the femoral neck, to investigate the mortality, complications, length of stay in hospital, blood loss, operating time and functional results.

Materials and Methods

A systematic review and meta-analysis was conducted following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines on randomised controlled trials (RCTs), studying current generation designs of stem only. The synthesis of results was done of pooled data, with a fixed effects or random effects model, based on heterogeneity.


Bone & Joint 360
Vol. 6, Issue 2 | Pages 10 - 13
1 Apr 2017


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1475 - 1480
1 Nov 2015
Inngul C Blomfeldt R Ponzer S Enocson A

The aim of this randomised controlled study was to compare functional and radiological outcomes between modern cemented and uncemented hydroxyapatite coated stems after one year in patients treated surgically for a fracture of the femoral neck. A total of 141 patients aged > 65 years were included. Patients were randomised to be treated with a cemented Exeter stem or an uncemented Bimetric stem. The patients were reviewed at four and 12 months.

The cemented group performed better than the uncemented group for the Harris hip score (78 vs 70.7, p = 0.004) at four months and for the Short Musculoskeletal Function Assesment Questionnaire dysfunction score at four (29.8 vs 39.2, p = 0.007) and 12 months (22.3 vs 34.9, p = 0.001). The mean EQ-5D index score was better in the cemented group at four (0.68 vs 0.53, p = 0.001) and 12 months (0.75 vs 0.58, p = < 0.001) follow-up. There were nine intra-operative fractures in the uncemented group and none in the cemented group.

In conclusion, our data do not support the use of an uncemented hydroxyapatite coated stem for the treatment of displaced fractures of the femoral neck in the elderly.

Cite this article: Bone Joint J 2015;97-B:1475–80.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (sd 1.71) vs 0.32 mm (sd 0.45), p = 0.66; rotation 0.96° (sd 1.49) vs 1.41° (sd 2.95), p = 0.88; and total migration 1.09 mm (sd 1.74) vs 0.73 mm (sd 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem.

Cite this article: Bone Joint J 2015; 97-B:595–602.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 541 - 541
1 Nov 2011
Rollier J Philippot R Vidalain J
Full Access

Purpose of the study: The pertinence of locking for major femoral revision remains a controversial issue. We conducted a retrospective multicentric study to assess the benefit and potential clinical and radiographic complications after using a long locked stem entirely coated with hydroxyapatite.

Material and methods: Our series included 77 patients (42 women, 35 men), mean age 71 years (range 34–90) reviewed at minimum one year. A modular implant was used; the long curved stem allowed total integration. Screws guaranteed distal locking. There were 71 revision THA on trochanteric-shaft fractures, three shaft nonunions, 34 aseptic loosenings, 9 septic loosening, 21 fractures on prosthesis, 6 implant failures and one instability. Seventy percent of patients had stage 3 and 4 bone lesions. Mean follow-up was 60 months.

Results: At last follow-up, 90% of patients were satisfied or very satisfied. The mean Harris function score improved from 32 to 83 and the PMA from 7 to 15. Sixty-four patients were pain free and 22% had pain solely under stress. There were 15 bone complications: dislocation (n=3), early infection (n=3 including 2 recurrences), implant failure (n=2), secondary femur fracture (n=5). Stem survival was 94% at nine years. Radiographically, total stem stability was noted in 73 patients (95%); instability was noted in four cases of fracture. Undeniable metaphyseal improvement was noted in 20% of cases, more modest improvement in 32%; the situation was considered unchanged in 31%. Locking was successful in all cases; cortical bone in contact with the locking screw was unchanged in 59 cases, slightly thickened in 16, including one case with a context of septic recurrence. The locking was dismantled in two cases (one empirically and one for pain), but the stem remained stable.

Discussion: Major damage to the femur can compromise stabilisation of the proximal or diaphyseal implants. Distal locking contributes to the initial mechanical stability, indispensable for secondary osteointegration of the implant. However, with total hydroxyapatite coating, rapid fixation in healthy zones is also crucial. Reconstruction of metaphyseal bone is not easy to quantify, but the absence of bone absorption is noteworthy.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1189 - 1193
1 Sep 2011
Zhao X Zhu Z Xie Y Yu B Yu D

When performing total hip replacement (THR) in high dislocated hips, the presence of soft-tissue contractures means that most surgeons prefer to use a femoral shortening osteotomy in order to avoid the risk of neurovascular damage. However, this technique will sacrifice femoral length and reduce the extent of any leg-length equalisation. We report our experience of 74 THRs performed between 2000 and 2008 in 65 patients with a high dislocated hip without a femoral shortening osteotomy. The mean age of the patients was 55 years (46 to 72) and the mean follow-up was 42 months (12 to 78). All implants were cementless except for one resurfacing hip implant. We attempted to place the acetabular component in the anatomical position in each hip. The mean Harris hip score improved from 53 points (34 to 74) pre-operatively to 86 points (78 to 95) at final follow-up. The mean radiologically determined leg lengthening was 42 mm (30 to 66), and the mean leg-length discrepancy decreased from 36 mm (5 to 56) pre-operatively to 8.5 mm (0 to 18) postoperatively. Although there were four (5%) post-operative femoral nerve palsies, three had fully resolved by six months after the operation. No loosening of the implant was observed, and no dislocations or infections were encountered.

Total hip replacement without a femoral shortening osteotomy proved to be a safe and effective surgical treatment for high dislocated hips.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 88 - 88
1 May 2011
Mäkelä K Eskelinen A Pulkkinen P Paavolainen P Remes V
Full Access

Background: According to the mid-term results obtained from the previous registry-based studies, survival of cementless stems for aseptic loosening in younger patients with primary osteoarthritis has been better than the survival of cemented stems. However, it has not been clear if the endurance against aseptic loosening of cementless cups is comparable to that of cemented cups. The aim of the present study was to analyze population-based long-term survival rates of the cemented and cementless total hip replacements in patients under the age of fifty-five years with primary osteoarthritis in Finland. Patients and Methods: Between 1980 and 2006, a total of 7310 primary total hip replacements performed for primary osteoarthritis in patients under the age of fifty-five years were entered in the Finnish Arthroplasty Registry. 4,032 of them fulfilled our inclusion criteria and were subjected to analysis. The implants included were classified in one of the three following groups: implants with a cementless, straight, proximally circumferentially porous-coated stem and a porous-coated press-fit cup (cementless group #1); implants with a cementless, anatomic, proximally circumferentially porous-coated and/ or hydroxyapatite-coated stem with a porous-coated and/or hydroxyapatite-coated press-fit cup (cementless group #2); and a cemented stem combined with a cemented all-polyethylene cup (the cemented group). Results: Cementless total hip replacements, as well as cementless stems and cups analyzed separately, had a significantly reduced risk of revision for aseptic loosening compared with cemented hip replacements. The 15-year survivorship of cementless stem groups for aseptic loosening was higher than that of cemented stems (89% and 90% vs. 72%). The 15-year survivorship of cementless press-fit porous-coated cups for aseptic loosening was higher than that of cemented cups (80% vs. 71%). When revision for any reason was the end point in survival analyses, however, there were no significant differences among the groups. Conclusions: Both cementless stems and cementless cups have better resistance to aseptic loosening than cemented implants in long term follow-up in younger patients. Even if liner-exchange revisions are taken into account, the long-term survival of cementless total hip replacements is comparable to that of cemented implants


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 271 - 271
1 May 2010
Bidar R Asencio G
Full Access

Purpose of the study: Cementless hydroxyapatite-coated prostheses have proven their reliability, but how about their longevity? The intimate contact between the receiver bone and the prosthetic implant, due to the bone ongrowth, allows ‘union’ between the inert and living material. The purpose of this restrospective study was to evaluate the behavior of an anatomic femoral hydroxyapatite-coated stem at long-term follow-up. Materials and Methods: One hundred and eleven total hip arthroplasties were performed with a ABG 1 HA-Coated hip prosthesis (ABG, Stryker Newbury, England) in 101 patients, by one senior operator. A posterolateral approach was done every time. Average patient age at surgery was 61, 3 years [33–83]. Results: Average follow-up was 13 years [10–17,5]. Seventy five THA were reviewed, 21 were dead and 15 were lost. Clinical outcome have been good. The average preoperative Harris hip score and Postel Merle d’Aubigné score were 56,7 and 11,8, which rose to 87,3 and 16,3 at the longest follow-up. Most patients were satisfied with the outcome, no thigh pain was deplored. There were sixteen revisions: 1 bipolar revision for loosening of both component, 14 cup revisions for osteolysis (11) and loosening (3) and 1 stem revision for periprosthetic fracture. Radiographic changes were consistent with bone remodeling. Osteointegration was achieved in 98,7%. Proximal stress-shielding was noted in 36,4%, osteolysis was developed in 66,7% focused essentially in gruen zones 1a and 7a and never below HA coating. Cancellous densifications were found to be mainly in zones 2, 6, 9 and 13 (92,7%, 94,2%, 91,3% and 85,5%), extending distally in zones 3, 5, 10 and 12 (33,3%, 31,9%, 34,8% and 39,1%). A femoral shaft hypertrophy was noted in 27,5%. Discussion: In this serie, the global survivorship of ABG 1 THA at 13 years of follow-up was 78,7%. Acetabular cup survivorship was 80%. Failure of acetabular cup was well explained by osteolysis and high wear rate. Stem survivorship was 97,3% and the failure rate of femoral stem due to aseptic loosening was only 1,3%. Excellent results of femoral stem were due to the good osteointegration, cause of the anatomic design of the stem and the hydroxyapatite coating. Conclusion: The results of our series have proven that anatomic cementless stem with HA coating provide a long term reliability and longevity