Advertisement for orthosearch.org.uk
Results 1 - 20 of 26
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 10 | Pages 1067 - 1073
1 Oct 2024
Lodge CJ Adlan A Nandra RS Kaur J Jeys L Stevenson JD

Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected knee prosthesis between September 2007 and January 2020 at a single institution. Radiographs and clinical records were used to assess and classify the incidence and causes of static spacer failure. Of the 94 cases, there were 19 primary total knee arthroplasties (TKAs), ten revision TKAs (varus-valgus constraint), 20 hinged TKAs, one arthrodesis (nail), one failed spacer (performed elsewhere), 21 distal femoral endoprosthetic arthroplasties, and 22 proximal tibial arthroplasties. Results. A total of 35/94 patients (37.2%) had spacer-related complications, of which 26/35 complications (74.3%) were because of mechanical failure of the spacer construct, while 9/35 (25.7%) were due to recurrence of infection. Risk factors for internal failure were a construct where the total intramedullary spacer length was less than twice the length of the central osseous defect (p = 0.009), where proximal or distal intraosseous spacer contact was < 10%, and after tibial tubercle osteotomy (p = 0.005). The incidence of spacer complications significantly increased the time to second stage: mean 157 days (42 to 458) in those without complications versus 227 days (11 to 528) with complications (p = 0.014). Conclusion. The failure rate of static antibiotic-loaded cement spacers is much higher than anticipated. Complications of the spacer significantly increased the time to second-stage revision. The risk of mechanical failure is significantly increased if the spacer is less than double the size of the segmental defect, or if inadequate reinforcement is inserted into the residual bone. These findings serve as a guide for surgeons to avoid mechanical complications with static spacers. Cite this article: Bone Joint J 2024;106-B(10):1067–1073


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 1000 - 1007
1 Sep 2024
Gong T Lu M Sheng H Li Z Zhou Y Luo Y Min L Tu C

Aims

Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes.

Methods

This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 593 - 601
1 Jun 2023
Scott CEH Yapp LZ Howard T Patton JT Moran M

Periprosthetic femoral fractures are increasing in incidence, and typically occur in frail elderly patients. They are similar to pathological fractures in many ways. The aims of treatment are the same, including 'getting it right first time' with a single operation, which allows immediate unrestricted weightbearing, with a low risk of complications, and one that avoids the creation of stress risers locally that may predispose to further peri-implant fracture. The surgical approach to these fractures, the associated soft-tissue handling, and exposure of the fracture are key elements in minimizing the high rate of complications. This annotation describes the approaches to the femur that can be used to facilitate the surgical management of peri- and interprosthetic fractures of the femur at all levels using either modern methods of fixation or revision arthroplasty.

Cite this article: Bone Joint J 2023;105-B(6):593–601.


Bone & Joint 360
Vol. 11, Issue 1 | Pages 41 - 43
1 Feb 2022


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1633 - 1640
1 Oct 2021
Lex JR Evans S Parry MC Jeys L Stevenson JD

Aims. Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. Methods. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%). Results. At a mean 74.4 months follow-up, the overall revision rate was 15.0%. Primary malignancy (p < 0.001) and age < 50 years (p < 0.001) were risk factors for revision. The risks of death and implant failure were similar in patients with primary disease (p = 0.872), but the risk of death was significantly greater for patients who had metastatic bone disease (p < 0.001). Acetabular-related implant failures comprised 74.3% of revisions; however, no difference between hemiarthroplasty or arthroplasty groups (p = 0.209), or between monopolar or bipolar hemiarthroplasties (p = 0.307), was observed. There was greater radiological wear in patients with longer follow-up and primary bone malignancy. Re-revision rates following a revision PFEPR was 34.3%, with dual-mobility bearings having the lowest rate of instability and re-revision (15.4%). Conclusion. Hemiarthroplasty and arthroplasty PFEPRs carry the same risk of revision in the medium term, and is primarily due to acetabular complications. There is no difference in revision rates or erosion between monopolar and bipolar hemiarthroplasties. The main causes of failure were acetabular wear in the hemiarthroplasty group and instability in the arthroplasty group. These risks should be balanced and patient prognosis considered when contemplating the bearing choice. Dual-mobility, constrained bearings, or large diameter heads (> 32 mm) are recommended in all revision PFEPRs. Cite this article: Bone Joint J 2021;103-B(10):1633–1640


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 309 - 320
1 Feb 2021
Powell-Bowns MFR Oag E Ng N Pandit H Moran M Patton JT Clement ND Scott CEH

Aims

The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter).

Methods

This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 285 - 292
1 Mar 2020
Tanaka A Katagiri H Murata H Wasa J Miyagi M Honda Y Takahashi M

Aims

The aim of this study is to evaluate the clinical results of operative intervention for femoral metastases which were selected based on expected survival and to discuss appropriate surgical strategies.

Methods

From 2002 to 2017, 148 consecutive patients undergoing surgery for femoral metastasis were included in this study. Prognostic risk assessments were performed according to the Katagiri and revised Katagiri scoring system. In general, the low-risk group underwent resection and reconstruction with endoprosthetic replacement (EPR), while the high-risk group underwent internal fixation (IF) and radiation therapy. For the intermediate-risk group, the operative choice depended on the patient’s condition, degree of bone destruction, and radio-sensitivity. Overall survival, local failure, walking ability, and systemic complications were evaluated.


Bone & Joint 360
Vol. 8, Issue 5 | Pages 4 - 10
1 Oct 2019
Tsoi K Samuel A Jeys LM Ashford RU Gregory JJ


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 522 - 528
1 May 2019
Medellin MR Fujiwara T Clark R Stevenson JD Parry M Jeys L

Aims. The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR). Patients and Methods. In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors. Results. The revision-free survival of the implant was 71% at five years and 63.3% at ten years. Three prostheses reached 15 years without revision. The mean Musculoskeletal Tumor Society score in the group was 26 (23 to 28). The mechanisms of failure were infection in 18%, structural failures in 6%, tumour progression in 5%, aseptic loosening in 2%, and soft-tissue failures in 1%. Prostheses used for primary reconstruction after oncological resections had lower infection rates than revision implants (8% vs 25%; p = 0.001). The rates of infection in silver-coated and non-silver-coated prosthesis were similar (17.4% vs 19.%; p = 0.869). The incidence of hip dislocation was 10%. Rotating hinge prosthesis had a lower failure rate than fixed hinge prosthesis (5.3% vs 11%). After Cox regression, the independent factors associated with failures were the history of previous operations (hazard ratio (HR) 3.7; p = 0.041), and the associated arthroplasty of the proximal tibia (HR 3.8; p = 0.034). At last follow-up, 11 patients (13%) required amputation. Conclusion. TFEPR offers a reliable reconstruction option for massive bone loss of the femur, with a good survival when the prosthesis is used as a primary implant. The use of a rotating hinge at the knee and dual mobility bearing at the hip may be adequate to reduce the risk of mechanical and soft-tissue failures. Infection remains the main concern and there is insufficient evidence to support the routine use of silver-coated endoprosthesis. Cite this article: Bone Joint J 2019;101-B:522–528


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1689 - 1695
1 Dec 2017
Stevenson JD Wigley C Burton H Ghezelayagh S Morris G Evans S Parry M Jeys L

Aims

Following the resection of an extensive amount of bone in the treatment of a tumour, the residual segment may be insufficient to accept a standard length intramedullary cemented stem. Short-stemmed endoprostheses conceivably have an increased risk of aseptic loosening. Extra-cortical plates have been added to minimise this risk by supplementing fixation. The aim of this study was to investigate the survivorship of short-stemmed endoprostheses and extra-cortical plates.

Patients and Methods

The study involved 37 patients who underwent limb salvage surgery for a primary neoplasm of bone between 1998 and 2013. Endoprosthetic replacement involved the proximal humerus in nine, the proximal femur in nine, the distal femur in 13 and the proximal tibia in six patients. There were 12 primary (32%) and 25 revision procedures (68%). Implant survivorship was compared with matched controls. The amount of bone that was resected was > 70% of its length and statistically greater than the standard control group at each anatomical site.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 263 - 269
1 Feb 2014
Batta V Coathup MJ Parratt MT Pollock RC Aston WJ Cannon SR Skinner JA Briggs TW Blunn GW

We reviewed the outcome of 69 uncemented, custom-made, distal femoral endoprosthetic replacements performed in 69 patients between 1994 and 2006. There were 31 women and 38 men with a mean age at implantation of 16.5 years (5 to 37). All procedures were performed for primary malignant bone tumours of the distal femur. At a mean follow-up of 124.2 months (4 to 212), 53 patients were alive, with one patient lost to follow-up. All nine implants (13.0%) were revised due to aseptic loosening at a mean of 52 months (8 to 91); three implants (4.3%) were revised due to fracture of the shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic fracture. Bone remodelling associated with periosteal cortical thinning adjacent to the uncemented intramedullary stem was seen in 24 patients but this did not predispose to failure. All aseptically loose implants in this series were diagnosed to be loose within the first five years. The results from this study suggest that custom-made uncemented distal femur replacements have a higher rate of aseptic loosening compared to published results for this design when used with cemented fixation. Loosening of uncemented replacements occurs early indicating that initial fixation of the implant is crucial. Cite this article: Bone Joint J 2014;96-B:263–9


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 314 - 314
1 Jul 2011
Ahmed I Gray A Aderinto J Howie C Patton J
Full Access

Background: Chronic prosthetic joint infection is a cause of patient morbidity and can be challenging to treat. Surgeons performing revision arthroplasty of the hip and knee are confronted with a growing number of patients with extensive loss of bone stock. The use of a modular endoprosthesis is a possible method of treatment in such patients. Aim: The purpose of this study was to assess the functional outcomes and the success of a single and two stage revision procedure in eradicating chronic prosthetic joint infection using a femoral endoprosthesis. Methods: A prospective database was reviewed of 20 patients who underwent a proximal, total or distal femoral endoprosthetic replacement after chronic prosthetic infection. Radiographs performed at the time of latest follow up were evaluated for signs of loosening, osteomyelitis and implant failure. The functional status was assessed using the Short Form (SF)-36 health survey score, Toronto extremity salvage score (TESS) and the Enneking score. Results: Thirteen patients underwent a single stage revision procedure and seven had been treated with a staged revision. At the latest follow up none of the 13 patients treated with a single stage procedure had evidence of ongoing infection. Of the seven patients who had a staged revision, 3 patients had evidence of ongoing infection. The mean pre operative Enneking score for the entire group was 17.1 points and this improved to 47.5 points post operatively (p< 0.0002). The mean pre operative TESS score for the entire group was 42% and this improved to 59% post operatively (p< 0.005). There was also a statistically significant improvement in all of the components of the SF-36 score. Conclusion: We believe that the use of a modular endo-prosthesis in the treatment of chronic prosthetic joint infection is a successful and viable option in eradicating infection, preserving the limb and providing a good functional result


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 352 - 352
1 Jul 2011
Ioannou M Papanastassiou I Farfalli G Carol M Morris C Healey J
Full Access

The objective of this study was to compare the results of two consecutive series of patients with either intra-medullary uncemented stems (UCS) distal femoral endoprosthetic replacement or the Compress. ®. (CMP) distal femoral implant. Patients were divided into two groups: those who received UCS prosthesis (Group-1: 54 patients) and those who received CMP prosthesis (Group-2: 42 patients).The most frequent diagnosis was osteosarcoma. Age and gender were similar both groups. In Group-1, at a mean follow-up of 144 months, 37 prostheses were still in place. The overall Kaplan-Meier prosthetic survival rates were 79% at five and 62 % at ten years. Most of failures were long term complications. Aseptic loosening was the primary cause of late prosthetic failure. On Cox regression analysis, prosthetic stem diameter under 13mm was a significant negative prognostic factor for prosthetic survival (p=0.016). In Group-2, at a mean follow-up of 84 months, 36 prostheses were still in place. The overall rate of CMP prosthesis survival was 86% at 5 years. All complications were during the first postoperative year, being femoral fracture the main revision cause. The patients who retained the prosthesis had mainly good or excellent MSTS functional results in both groups


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 266 - 266
1 Jul 2011
Jones KB Riad S Griffin A Deheshi B Bell RS Ferguson P Wunder JS
Full Access

Purpose: Few functional outcomes of total femoral endoprosthetic replacement (TFEPR) using contemporary modular systems are available. We compared functional results between TFEPR patients receiving fixed- and rotating-hinge knee componentry following oncologic resections. Method: Eighteen TFEPR patients were identified from a prospectively gathered sarcoma database. Six were secondary procedures and 12 primary. Four patients had metastatic carcinoma, 8 osteosarcoma, 4 non-osteogenic spindle cell sarcomas of bone, 1 Ewing’s sarcoma, and 1 femur-invading soft-tissue sarcoma. All reconstructions used modular implants from a single company. Proximally, all were bipolar hip hemiarthoplasties, 12 including abductor reattachment. Distally, 8 had fixed- and 10 had rotating-hinge knee componentry. Toronto Extremity Salvage Score (TESS), and both Musculoskeletal Tumor Society Scores (MSTS) were compared between fixed- and rotating-hinge groups using the Mann-Whitney test. Results: Complications included 1 hip dislocation, 1 femoral malrotation, and wound problems requiring 3 debridements and 1 amputation. One metastatic carcinoma patient developed local relapse. Follow-up averaged 4 years (range 1 month to 14 years). At latest follow-up, 10 patients had died of disease. Eight remained alive, 6 disease-free, 2 with distant disease. Among patients surviving 6 months, 6 used no assistive devices, 5 used a single cane, and 4 were wheelchair bound, each at least partly due to distant disease progression. TESS averaged 74.5±17.4, MSTS1987 25.2±4.4; and MSTS1993 58.6±22.9 among the 12 patients for whom functional results were available from latest follow-up. No statistically significant differences or even trends were detected between fixed-hinge and rotating-hinge patients (lowest p = 0.755), but both instability problems were in the rotating-hinge group. Conclusion: While both rotating- and fixed-hinge TFEPR reconstructions may function well, consideration should be given to fixed-hinge knee reconstruction when massive myectomies or poorer conditioning make hip and knee stability a primary concern in the short-term


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 708 - 710
1 May 2011
Gaston CL Tillman RM Grimer RJ

We report a case of spontaneous physeal growth arrest of the distal femur in a nine-year-old child with Ewing’s sarcoma of the proximal femur treated with chemotherapy and endoprosthetic replacement. Owing to the extent of disuse osteoporosis at the time of surgery, the entire intramedullary canal up to the distal femoral physis was filled with cement. Three years later, the femur remained at its pre-operative length of 19 cm. Pre-operative calculations of further growth failed to account for the growth arrest, and the initial expandable growing prosthesis inserted has been revised to a longer one in order to address the leg-length discrepancy.

To our knowledge, this is the only reported case of distal femoral physeal growth arrest following cemented endoprosthetic replacement of the proximal femur.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 867 - 874
1 Jun 2010
Hanna SA Sewell MD Aston WJS Pollock RC Skinner JA Cannon SR Briggs TWR

Segmental resection of malignant bone disease in the femoral diaphysis with subsequent limb reconstruction is a major undertaking. This is a retrospective review of 23 patients who had undergone limb salvage by endoprosthetic replacement of the femoral diaphysis for a primary bone tumour between 1989 and 2005. There were 16 males and seven females, with a mean age of 41.3 years (10 to 68). The mean overall follow-up was for 97 months (3 to 240), and 120 months (42 to 240) for the living patients. The cumulative patient survival was 77% (95% confidence interval 63% to 95%) at ten years. Survival of the implant, with failure of the endoprosthesis as an endpoint, was 85% at five years and 68% (95% confidence interval 42% to 92%) at ten years. The revision rate was 22% and the overall rate of re-operation was 26%. Complications included deep infection (4%), breakage of the prosthesis (8%), periprosthetic fracture (4%), aseptic loosening (4%), local recurrence (4%) and metastases (17%). The 16 patients who retained their diaphyseal endoprosthesis had a mean Musculoskeletal Tumour Society score of 87% (67% to 93%). They were all able to comfortably perform most activities of daily living. Femoral diaphyseal endoprosthetic replacement is a viable option for reconstruction following segmental resection of malignant bone disease. It allows immediate weight-bearing, is associated with a good long-term functional outcome, has an acceptable complication and revision rate and, most importantly, does not appear to compromise patient survival


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 64 - 64
1 Mar 2010
Sewell M Spiegelberg B Hanna S Aston W Cannon S Briggs T
Full Access

Introduction: Total femoral endoprosthetic replacement can be used as an alternative to hip disarticulation following extensive tumour excision or in cases of more severe bone loss. To date there have been no long term studies on the oncological and functional outcome of patients who have had this procedure for malignant bone disease. We report our experience of over 25 years employing this procedure using a custom-made and modular total femoral endoprosthetic replacement. Methods: This is a retrospective, single centre study of 27 patients who underwent consecutive total femoral replacement as a primary procedure between 1978 and 2005. Information was collected from the bone tumour database, medical records and clinic review. Outcome was assessed using the Musculoskeletal Tumour Society (MSTS) rating score, the Harris Hip Score (HHS) and the Oxford Knee Score (OKS). Results: There were 15 males and 12 females, with a mean age of 30 years (5 to 65). The overall mean follow-up was 4.3 years (1 to 16.4) for all patients, 9.1 years (1 to 16.4) for the 7 patients who were alive at the time of this review, and 2.6 years (1 to 13) for the 20 patients who had died. 24 patients had primary malignant bone tumours of the femur and 3 had metastatic disease from a known primary elsewhere. 3 patients developed local recurrence, 1 patient developed new lung metastases. Mean MSTS score was 63% (19/30), mean HHS was 68/100 and mean OKS was 34/48. Discussion: Total femoral replacement can be an effective alternative to hip disarticulation in patients with malignant bone disease. Good functional outcomes can be achieved without compromising survival


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1513 - 1520
1 Nov 2009
Sewell MD Spiegelberg BGI Hanna SA Aston WJS Bartlett W Blunn GW David LA Cannon SR Briggs TWR

We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 401 - 404
1 Mar 2009
Chandrasekar CR Grimer RJ Carter SR Tillman RM Abudu A Jeys LM

We undertook a cemental unipolar proximal femoral endoprosthetic replacement in 131 patients with a mean age of 50 years (2 to 84). Primary malignant tumours were present in 54 patients and 67 had metastatic disease. In addition, eight patients had either lymphoma or myeloma and two had non-oncological disorders. The mean follow-up was 27 months (0 to 180). An acetabular revision was required later in 14 patients, 12 of whom had been under the age of 21 years at the time of insertion of their original prosthesis. The risk of acetabular revision in patients over 21 years of age was 8% at five years compared with 36% in those aged under 21 years. All the unipolar hips in this younger age group required revision within 11 years of the initial operation. We conclude that unipolar replacement should not be used in younger patients and should be avoided in patients with a life expectancy of more than five years


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 108 - 112
1 Jan 2009
Chandrasekar CR Grimer RJ Carter SR Tillman RM Abudu A Buckley L

Endoprosthetic replacement of the proximal femur may be required to treat primary bone tumours or destructive metastases either with impending or established pathological fracture. Modular prostheses are available off the shelf and can be adapted to most reconstructive situations for this purpose. We have assessed the clinical and functional outcome of using the METS (Stanmore Implants Worldwide) modular tumour prosthesis to reconstruct the proximal femur in 100 consecutive patients between 2001 and 2006. We compared the results with the published series for patients managed with modular and custom-made endoprosthetic replacements for the same conditions.

There were 52 males and 48 females with a mean age of 56.3 years (16 to 84) and a mean follow-up of 24.6 months (0 to 60). In 65 patients the procedure was undertaken for metastases, in 25 for a primary bone tumour, and in ten for other malignant conditions. A total of 46 patients presented with a pathological fracture, and 19 presented with failed fixation of a previous pathological fracture. The overall patient survival was 63.6% at one year and 23.1% at five years, and was significantly better for patients with a primary bone tumour than for those with metastatic tumour (82.3% vs 53.3%, respectively at one year (p = 0.003)). There were six early dislocations of which five could be treated by closed reduction. No patient needed revision surgery for dislocation. Revision surgery was required by six (6%) patients, five for pain caused by acetabular wear and one for tumour progression. Amputation was needed in four patients for local recurrence or infection.

The estimated five-year implant survival with revision as the endpoint was 90.7%. The mean Toronto Extremity Salvage score was 61% (51% to 95%). The implant survival and complications resulting from the use of the modular system were comparable to the published series of both custom-made and other modular proximal femoral implants.

We conclude that at intermediate follow-up the modular tumour prosthesis for proximal femur replacement provides versatility, a low incidence of implant-related complications and acceptable function for patients with metastatic tumours, pathological fractures and failed fixation of the proximal femur. It also functions as well as a custom-made endoprosthetic replacement.