The aim of this study was to evaluate the suitability of the tapered cone stem in total hip arthroplasty (THA) in patients with excessive femoral anteversion and after femoral osteotomy. We included patients who underwent THA using Wagner Cone due to proximal femur anatomical abnormalities between August 2014 and January 2019 at a single institution. We investigated implant survival time using the endpoint of dislocation and revision, and compared the prevalence of prosthetic impingements between the Wagner Cone, a tapered cone stem, and the Taperloc, a tapered wedge stem, through simulation. We also collected Oxford Hip Score (OHS), visual analogue scale (VAS) satisfaction, and VAS pain by postal survey in August 2023 and explored variables associated with those scores.Aims
Methods
For over a decade, modular titanium fluted tapered (TFT) stems have demonstrated excellent clinical success for femoral revision total hip arthroplasty (THA) surgery. The aim of this study was to report the short-term outcomes of a novel modern monoblock TFT stem used for revision and complex primary THA with a minimum of 2 years follow-up. We identified 126 patients who received a single monoblock TFT stem - 26 patients for complex THA (failed fracture fixation) and 100 patients for revision THA. The reasons for revision THA included 40 for previous prosthetic joint infection (PJI), 42 for aseptic loosening, 9 for trunnionosis, 9 for periprosthetic fractures. The Paprosky grading for femoral bone loss at the time of surgery and the measured subsidence of femoral stems at 3 months follow-up were determined. We evaluated the number and indications for re-operations. The mean time from surgery was 3.9 years (range 2.0 to 6.9 years). A paired t-test analysis showed significant improvement from pre-operative versus post-operative clinical outcome scores (p<0.001) for HHS (38.76 +/- 15.24vs. 83.42 +/- 15.38), WOMAC (45.6 ± 19.0 vs. 69.9 ± 21.3) and SF-12 Physical component (31.7 ± 8.1 vs. 37.8 ± 11.3) and SF-12 Mental component (48.2 ± 12.2 vs. 51.6 ± 12.5). The Paprosky grading for femoral bone loss was Grade 1 (3.9%), Grade 2 (35.7%), Grade 3A (47.6%), Grade 3B (11.1%) and Grade 4 (1.6%) cases. There were 18 re-operations (14.7%) with 13 for PJI (7 treated with implant retention, 6 treated with a two-staged revision), 4 for instability and one for acetabular aseptic loosening. There were no aseptic failures of the stem. This novel modern monoblock TFT stem provided reliable femoral fixation and has increasingly supplanted the use of modular TFT stems for complex primary and revision surgery in our institution.
We had previously reported on early outcomes on a new fluted, titanium, monobloc stem with a three degree taper that has been designed for challenging femoral reconstruction in the setting of extensive bone loss. The aim of this study was to report its mid-term clinical and radiographic outcomes. This is a retrospective review of prospectively collected data carried out at a single institution between Jan 2017 and Dec 2019. 85 femoral revisions were performed using a new tapered, fluted, titanium, monobloc (TFTM) revision stem. Complications, clinical and radiographic data were obtained from medical records and a locally maintained database. Clinical outcomes were assessed using the Oxford Hip Score (OHS) and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC). All post-operative radiographs were analysed for subsidence, osteolysis and femoral cortical bone remodelling. Mean follow-up was 60 months (range 28–84 months). Subsidence of 1.2 mm was noted in one patient. No cases of clinically significant subsidence (10 mm) were observed. At final follow-up, a statistically significant improvement was noted in functional outcome scores. The mean OHS preoperatively and at final follow-up were 24 (SD 13) and 42 (SD15). p = 0.04 mean difference 18 (95% CI 15–22). The mean WOMAC scores preoperatively and at final follow-up were 62 (SD23) and 88 (SD7) respectively (p < 0.001, mean difference 26; 95% CI 21–34). No stem fractures were noted within the follow-up period. Two patients had revision of the stem's one for infection and another for persistent pain. Positive mid-term clinical and radiological outcomes have been observed with this tapered, fluted, titanium, monobloc stem. Based on these results, this implant may be considered as a viable option in the majority of uncemented femoral revisions.
Aim of this study was to identify reoperation rates in patients with short oblique and transverse fractures around a well fixed cemented polished taper slip stem and to determine any associations with treatment failure. Retrospective cohort study of 31 patients with AO transverse or short oblique Vancouver B1 PFFs around THA (total hip arthroplasty) cemented taper slip stems: 12 male (39%); mean age 74±11.9 (range 44–91); mean BMI 28.5±1.4 (range 16–48); and median ASA 3. Patient journeys were assessed, re-interventions reviewed. The primary outcome measure was reoperation. Time from primary THA to fracture was 11.3±7.8yrs (0.5–26yrs). Primary surgical management was fixation in 27/31 and rTHA (revision total hip arthroplasty) in 4/31. 10 of 31 (32%) patients required reoperation, 9 within 2 years of fracture: 1 following rTHA and 8 following ORIF. The commonest mode of failure was non-union (n=6). No significant associations with reoperation requirement were identified. Kaplan-Meier free from reoperation was 67.4% (49.8–85.0 95% CI) at 2 years and this was unaffected by initial management with ORIF or rTHA (Log rank 0.898). Of those reoperated, 6/10 required multiple reoperations to obtain either bony union or a stable revision construct and 13% ultimately required proximal femoral endoprostheses. The relative risk of 1 year mortality was 1.6 (0.25 to 10.1 95%CI) among patients who required reoperation compared to those who did not. These are difficult fractures to manage, should not be underestimated and patients should be counselled that there is a 30% risk of reoperation and 20% of requiring multiple reoperations.
There is a paucity of long-term data on modular fluted tapered (MFT) stems for two-stage reimplantation following periprosthetic joint infection (PJI). The purpose of this study was to evaluate implant survivorship, radiographic results, and clinical outcomes in a large cohort of reimplantation THAs using MFT stems. We identified 236 reimplantation THAs from a single tertiary care academic institution from 2000 to 2020. Two designs of MFT stems were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years, mean BMI was 32 kg/m2, and 46% were female. Median stem diameter was 19 mm, and median stem length was 195 mm. Mean follow-up was 7 years. A competing risk model accounting for death was utilized. The 15-year cumulative incidence of any revision was 24%. There were 48 revisions, with the most common reasons being dislocation (n=25) and PJI (n=16). The 15-year cumulative incidence of any reoperation was 28%. Only 13 revisions involved the fluted tapered portion of the component (FTC), for a 15-year cumulative incidence of any FTC revision of 8%. Only 2 FTCs were revised for aseptic loosening, resulting in a 15-year cumulative incidence of FTC revision for aseptic loosening of 1%. Stem subsidence >5 mm occurred in 2% of unrevised cases, and all stems were radiographically stable at most recent follow-up. Mean HHS was 77 at most recent follow-up. This series demonstrated that MFT stems were durable and reliable even in the setting of two-stage reimplantation for infection. While the incidence of aseptic loosening was very low, the incidence of any revision was 24% at 15 years, primarily caused by dislocation and recurrent PJI. Level of Evidence: IV
Previous studies have reported excellent results with tapered, titanium alloy, porous plasma-sprayed components in patients undergoing uncemented primary total hip arthroplasty (THA). The purpose of this study was to examine survival and clinical results at minimum 25-year follow-up. We reviewed all patients who underwent primary THA at our center through 1995 with a specific femoral component, the Mallory-Head Porous (MHP; Zimmer Biomet, Warsaw, IN). This device, marketed in the U.S. until December 2021, was essentially unchanged since its 1984 introduction, except the porous coating was continued circumferentially along the lateral aspect in 1987, a hydroxyapatite-coated option was offered in 1988, and an offset option was added in 1999 after the study period. Three hundred thirty-two patients (396 THA) had a minimum of 25-year follow-up. Mean age at surgery was 47.6 years (range, 21–70 years). Mean follow-up in non-failed patients was 28.7 years (range, 25 to 37 years). There were 31 femoral revisions (7.8%): 9 infection, 3 failure of ingrowth, 5 aseptic loosening, 8 osteolysis revised well-fixed, 2 periprosthetic fracture, 2 polyethylene wear with trochanteric avulsion, 1 component breakage, and 1 malalignment well-fixed. Kaplan-Meier survival with endpoint of stem revision for all causes was 94.8% (95% CI: ±0.9%) at 36.7 years, and survival with endpoint of aseptic loosening/failure of ingrowth was 98.7% (95% CI: ±0.5) at 36.7 years. Harris hip scores improved significantly from 43 preoperatively to 76 most recently. This tapered, titanium, porous plasma spray-coated femoral component continues to demonstrate high long-term survival with a low rate of femoral component revision for any reason or aseptic loosening/failure of ingrowth.
Subsidence remains a concern when utilizing modern tapered fluted titanium (TFT) femoral stems and may lead to leg length discrepancy, impingement, instability and failure to obtain stem osseointegration. This study aims to compare stem subsidence across three modern TFT stems. Our secondary aim was to investigate the influence of bicortical contact or ‘scratch fit’ on subsidence, as well as the role of intraoperative imaging in maximizing this bicortical contact and preventing stem subsidence. A retrospective review of 271 hip arthroplasties utilizing modern TFT stems in a single institution was performed. Three stem designs were included in the analysis: one monoblock TFT stem (n=91) and two modular TFT stems (Modular A [n=90]; Modular B [n=90]). Patient demographics, Paprosky femoral bone loss classification, bi-cortical contact, utilization of intra-operative imaging and stem subsidence (comparison of initial post-operative radiograph to the latest follow up radiograph - minimum three months) were recorded. There was no statistically significant difference in the amount of subsidence between the three stems (Monoblock: 2.33mm, Modular A: 3.43mm, Modular B: 3.02mm; p=0.191). There was no statistical difference in subsidence >5mm between stems (Monoblock: 9.9%, Modular A: 22.2%, Modular B: 16.7%). Subgroup analysis based on femoral bone loss grading showed no difference in subsidence between stems. Increased bicortical contact was strongly associated with reduced subsidence (p=0.004). Intra-operative imaging was used in 46.5% (126/271) of cases; this was not correlated with bicortical contact (p=0.673) or subsidence (p=0.521). Across all groups, only two stems were revised for subsidence (0.7%). All three modern TFT stems were highly successful and associated with low rates of subsidence, regardless of modular or monoblock design. Surgeons should select the stem that they feel is most clinically appropriate.
Periprosthetic femoral fractures are increasingly seen in recent years, adding considerable burden to the National Health Service. These require complex revision or fixation and prolonged post-operative care, with significant morbidity with associated costs. The purpose of this study was to assess whether the size of femoral cement mantle is associated with periprosthetic femoral fractures (PPF). This retrospective study was carried out on a cohort of 49 patients (Fracture Group - FG) who previously had a revision procedure following a proximal PPF between 2010 and 2021. Inclusion criteria – all primary cemented total hip replacements (THR). Exclusion criteria – complex primary THR, any implant malposition that required early revision surgery or any pre-fracture stem loosening. The antero-posterior (AP) radiographs from this cohort of patients were assessed and compared to an age, sex, time since THR-matched control group of 49 patients without PPF (Control Group - CG). Distal cement mantle area (DCMA) was calculated on an AP radiograph of hip; the position of the femoral stem tip prior to fracture was also recorded: valgus, varus or central. Limitations: AP radiographs only. Statistical analyses were performed using Microsoft® Excel. Chi-square test demonstrated statistically significant difference in DCMA between FG and CG. DCMA of 700 to 900 mm² appeared to be protective when compared to DCMA of 0 to 300 mm². Also, a valgus position observed in 23% in FG Vs 4 % in CG increased the risk, with a smaller area of DCMA. This study demonstrates and recommends that a size of 700 – 900 mm² of the DCMA is protective against periprosthetic fractures, which are further influenced by the positioning of the distal stem tip. This could be due to the gradual decrease in the stiffness gradient from proximal to distal around the stem tip than steep changes, thereby decreasing possibility of a stress riser just distal to the cement mantle or restrictor. Further biomechanical research specific to this finding may be helpful to validate the observation, progressing to suggest a safe standardised surgical technique.
The aim of the study was to report the survival of open reduction and internal fixation (ORIF) of Vancouver B fractures associated with the Exeter Stem (ES) at a minimum of 5 years. This retrospective cohort study assessed 129 consecutive patients with Vancouver B type fractures treated with ORIF from 2008-2016 at a minimum of 5 years. Patient records were examined, and the following recorded: details of primary prosthesis, details of injury, Vancouver classification, details of operative management, complications, and requirement for reoperation. Data was analysed using SPSS. Survival analysis was undertaken using the endpoint ‘reoperation for any reason’. Mean age at fracture was 78.2 (SD10.6, 46-96) and 54 (43%) were female. Vancouver subclassifications were: 24% B1, 70.5% B2 and 5.5% B3. For all Vancouver B fractures, Kaplan Meier analysis demonstrated a 5 year survival free from reoperation of 88.8% (82.0-94.7 95%CI). Fourteen patients required reoperation, most commonly within the first year for non-union and plate fracture (5.4%). Five-year survival for any reoperation differed significantly according to fracture type (p=0.016) and was worst in B1s: B1 76.6% (61.3-91.9); B2 92.6% 986.9-98.3); and 100% of B3. Univariate analysis identified B1 type (p=0.008) and a transverse fracture pattern (p=0.003) to be significantly associated with the need for reoperation. Adopting a strategy of fixation of all Vancouver B fractures involving the ES where the fracture was anatomically reducible and the bone cement interface was well-fixed was associated with a 5 year survival, free from reoperation of 88.8%.
Cementless stem designs in total hip arthroplasty differ in relation to geometry and area of fixation. We utilised radiostereometric analysis (RSA) to evaluate the 2-year migration of a novel, short, proximally coated femoral stem. 30 participants undergoing primary total hip replacement for any cause (rheumatoid or inflammatory arthritis, osteoarthritis) were prospectively recruited in this study. Osteoporotic patients and cases of suspected infection were excluded. All patients received a short blade stem, proximally coated with a reduced lateral shoulder and narrow triple taper geometry to minimise bone removal. RSA radiographs were performed post-operatively and at 6 weeks, 6 months, 1- and 2 years. The Harris Hip Score (HHS), Oxford Hip Score (OHS) and EQ-5D were collected at baseline and at 2 years post-operatively. The stability of implants and complications were captured during each follow-up visit. A total of 14 female and 16 male patients were recruited with a mean age of 64.8 (range 47 to 75). At two years the mean subsidence of the stem was 0.34 mm (SD 0.62) and the total migration 0.74 mm (SD 0.60). The mean medial translation at two years was 0.059 (0.24) and the mean anterior translation 0.12 (0.59) respectively. Baseline PROM scores improved significantly at 2-years from pre-operatively (median and interquartile range): HHS from 33 (18.25) to 92 (19), EQ5D from 0.5 (0.35) to 0.94 (0.17), OHS from 21 (18.25) to 42 (4.25). P-value for all comparisons was <0.001. 2-year follow up data revealed no complications. There were no stem revisions in study participants and no heterotopic ossifications were identified on radiographs. 2-year migration results of a cementless, short blade, proximally coated tapered femoral stem using RSA, showed the stem exhibits a predictable migration pattern and achieves initial stability. This is highly likely to translate to mid and long-term stability, which needs to be corroborated by long-term outcome studies. Furthermore, participants demonstrated excellent clinical, patient reported and radiological outcomes after 2 years of follow up to support expansion in the use of this prosthesis.
Cement-in-cement revision of the femoral component represents a widely practiced technique for a variety of indications. In this study we compare the clinical and radiological outcomes of two polished tapered stems. From our prospectively collated database we identified all patients undergoing cement-in-cement revision from January 2005 – 2013 who had a minimum of two years follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-stem AMT high offset No 1. Patients were followed-up annually with clinical and radiological assessment. Ninety-seven patients matched the inclusion criteria. There were 50 Exeter and 47 C-stem AMT components. There were no significant differences between the patient demographics in either group. Mean follow-up was 9.7 years. A significant improvement in OHS, WOMAC and SF-12 scores was observed in both cohorts. Leg lengths were significantly shorter in the Exeter group, with a mean of -4mm in this cohort compared to 0mm in the C-stem AMT group. One patient in the Exeter group had early evidence of radiological loosening. In total, 16 patients (15%) underwent further revision of the femoral component (seven in the C-stem AMT group and nine in the Exeter group). No femoral components were revised for aseptic loosening. There were two cases of femoral component fracture in the Exeter group. Our series shows promising long-term outcomes for the cement-in-cement revision technique using either the Exeter or C-stem AMT components. These results demonstrate that cement-in-cement revision using a double or triple taper-slip design is a safe and reliable technique when used for the correct indications.
The aim of this study is to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). This is a retrospective cohort study of 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems. 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Radiographs were assessed and classified by 3 observers. The primary outcome measure was revision of ≥1 component. Kaplan Meier survival analysis was performed. Logistic regression was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay and mortality. Fractures (B1 n=74 (49%); B2 n=50 (33%); and B3 n=28 (18%)) occurred at mean 6.7±10.4 years after primary THA (n=143) or hemiarthroplasty (n=15). Mean follow up was 6.5 ±2.6 years (3.2 to 12.1). Rates of revision and reoperation were significantly higher following revision arthroplasty compared to ORIF for B2 (p=0.001) fractures and B3 fractures (p=0.05). Five-year survival was significantly better following ORIF: 92% (86.4 to 97.4 95%CI) Vs 63% (41.7 to 83.3), p<0.001. No independent predictors of revision following ORIF were identified: fixation of B2 or B3 fractures was not associated with an increased risk of revision. Dislocation was the commonest mode of failure after revision arthroplasty. ORIF was associated with reduced blood transfusion requirement and reoperations, but there were no differences in medical complications, hospital stay or mortality between surgical groups. When the bone-cement interface was intact and the fracture was anatomically reducible, Vancouver B2 fractures around Exeter stems can be treated with fixation as opposed to revision arthroplasty. Fixation of Vancouver B3 fractures can be performed in frail elderly patients without increasing revision risk.
Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.Aims
Methods
Stemmed tibial components are frequently used in revision total knee arthroplasty (TKA). The purpose of this study was to evaluate patient satisfaction, overall pain, and diaphyseal tibial pain in patients who underwent revision TKA with cemented or uncemented stemmed tibial components. This is a retrospective cohort study involving 110 patients with revision TKA with cemented versus uncemented stemmed tibial components. Patients who underwent revision TKA with stemmed tibial components over a 15-year period at a single institution with at least two-year follow-up were assessed. Pain was evaluated through postal surveys. There were 63 patients with cemented tibial stems and 47 with uncemented stems. Radiographs and Knee Society Scores were used to evaluate for objective findings associated with pain or patient dissatisfaction. Postal surveys were analyzed using Fisher’s exact test and the independent-samples Aims
Methods
We present the clinical and radiological results at a minimum follow-up of 20 years using a second-generation uncemented total hip arthroplasty (THA). These results are compared to our previously published results using a first-generation hip arthroplasty followed for 20 years. A total of 62 uncemented THAs in 60 patients were performed between 1993 and 1994. The titanium femoral component used in all cases was a Taperloc with a reduced distal stem. The acetabular component was a fully porous coated threaded hemispheric titanium shell (T-Tap ST). The outcome of every femoral and acetabular component with regard to retention or revision was determined for all 62 THAs. Complete clinical follow-up at a minimum of 20 years was obtained on every living patient. Radiological follow-up was obtained on all but one.Aims
Methods
Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.Aims
Methods
Revision hip arthroplasty presents a challenge in the setting of femoral bone loss.
To investigate predictors of periprosthetic fracture level (around stem (Vancouver B) or distal to stem (Vancouver C/D) in cemented polished tapered femoral stems. Retrospective cohort study of 188 patients (mean age 79 (range 30–91); 99 (53%) male) with unilateral periprosthetic femoral fractures associated with CPT stems. Medical notes were reviewed and the following recorded: patient demographics, past medical history, drug history, date of prosthesis insertion, and date of injury. Radiographs analysis included Vancouver classification, cement restrictor type, cement mantle to implant tip distance, cortical thickness, femoral diameter and DORR classification. Univariate, multivariate and ROC curve analysis was performed. Fractures occurred at mean 7.5 years following primary procedure: 152 (83%) were B fractures; and 36 (19%) C/D. On univariate analysis female gender, lower BMI, osteoporosis, NSAID use, Bisphosphonate therapy, cortical thickness, distal cement mantle length and distal cement mantle length:femoral diameter ratio were significantly associated with C level fractures (p<0.05). Distal cement mantle lengths of >19.6mm (AUC 0.688, p<0.001) were associate with C level fractures. Multivariate analysis demonstrated female gender and distal cement mantle length:femoral diameter ratio to be independent predictors of C level periprosthetic fractures. Though female sex is the largest independent predictor of periprosthetic fractures distal to a CPT femoral stem, the relationship between cortical thickness and distal cement mantle length appears significant. As fractures distal to the stem are invariably managed by ORIF, whereas fractures around the stem frequently require revision arthroplasty, this has relevance at primary surgery in osteoporotic females to reduce the need for complex revisions.
Recently, short shaped stem becomes popular in total hip arthroplasty (THA). Advantages of the short stem are preserving femoral bone stock, thought to be less thigh pain, suitable for minimally invasive THA. However, bony reaction around the short stem has not been well known. The purpose of this study was to compare the two years difference of radiographic change around the standard tapered round stem with the shorter tapered round stem. Evaluation was performed in 96 patients (100 joints) who underwent primary THA. Standard tapered round stem (Bicontact D stem) was used in 44 patients from January 2011 to May 2013. Shorter stem (Bicontact E stem) was used in 56 patients from May 2015 to March 2016. The proximal shapes of these two stems are almost the same curvature. The mean age at surgery was 64 years. The mean BMI at surgery was 24.0 kg/m2. Eighty-six patients had osteoarthrosis and 10 patients had osteonecrosis. Evaluation was performed 2 years after surgery with standard AP radiographs. The OrthoPilot imageless navigation system was used during surgery. Evaluation of the stem fixation, stress shielding, and cortical hypertrophy were carried out.INTRODUCTION
MATERIALS AND METHODS