Advertisement for orthosearch.org.uk
Results 1 - 5 of 5
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1183 - 1191
1 Sep 2015
Briggs TWR Hanna SA Kayani B Tai S Pollock RC Cannon SR Blunn GW Carrington RWJ

The long term biological effects of wear products following total hip arthroplasty (THA) are unclear. However, the indications for THA are expanding, with increasingly younger patients undergoing the procedure.

This prospective, randomised study compared two groups of patients undergoing THA after being randomised to receive one of two different bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal (MoM) n = 23. We investigated the relationship between three variables: bearing surface (MoP vs MoM), whole blood levels of chromium (Cr) and cobalt (Co) and chromosomal aberrations in peripheral lymphocyte pre-operatively and at one, two and five years post-surgery.

Our results demonstrated significantly higher mean cobalt and chromium (Co and Cr) blood levels in the MoM group at all follow-up points following surgery (p < 0.01), but there were no significant differences in the chromosomal aberration indices between MoM and MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p = 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34) for translocations, breaks, loss and gain of chromosomes respectively. Regression analysis showed a strong linear relationship between Cr levels and the total chromosomal aberration indices in the MoM group (R2 = 0.90016), but this was not as strong for Co (R2 = 0.68991). In the MoP group, the analysis revealed a poor relationship between Cr levels and the total chromosomal aberration indices (R2 = 0.23908) but a slightly stronger relationship for Co (R2 = 0.64292). Across both groups, Spearman’s correlation detected no overall association between Co and Cr levels and each of the studied chromosomal aberrations. There remains no clear indication which THA bearing couple is the most biocompatible, especially in young active patients. While THA continues to be very successful at alleviating pain and restoring function, the long-term biological implications of the procedure still require further scrutiny.

Cite this article: Bone Joint J 2015;97-B:1183–91.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1040 - 1044
1 Aug 2013
Kazi HA Perera JR Gillott E Carroll FA Briggs TWR

We prospectively assessed the efficacy of a ceramic-on-metal (CoM) hip bearing with uncemented acetabular and femoral components in which cobalt­–chrome acetabular liners and alumina ceramic heads were used.

The cohort comprised 94 total hip replacements (THRs) in 83 patients (38 women and 45 men) with a mean age of 58 years (42 to 70). Minimum follow-up was two years. All patients had pre- and post-operative assessment using the Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Oxford hip score and Short-Form 12 scores. All showed a statistically significant improvement from three months post-operatively onwards (all p < 0.001).

After two years whole blood metal ion levels were measured and chromosomal analysis was performed. The levels of all metal ions were elevated except vanadium. Levels of chromium, cobalt, molybdenum and titanium were significantly higher in patients who underwent bilateral THR compared with those undergoing unilateral THR (p < 0.001). Chromosomal analysis demonstrated both structural and aneuploidy mutations. There were significantly more breaks and losses than in the normal population (p < 0.001). There was no significant difference in chromosomal aberration between those undergoing unilateral and bilateral procedures (all analyses p ≥ 0.62).

The use of a CoM THR is effective clinically in the short-term, with no concerns, but the significance of high metal ion levels and chromosomal aberrations in the long-term remains unclear.

Cite this article: Bone Joint J 2013;95-B:1040–44.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_III | Pages 62 - 62
1 Feb 2012
Dunstan E Ladon D Whittingham-Jones P Cannon S Briggs T Case P
Full Access

Purpose. Metal-on-Metal (MoM) hip bearings are being implanted in ever-increasing numbers and into ever-younger patients. The consequence of chronic exposure to metal ions is a cause for concern. Therefore, using cytogenetic biomarkers, we investigated a group of patients who have had MoM bearings in situ for in excess of 30 years. Method. Whole blood specimens were obtained from an historical group of patients who have had MoM bearings in situ for in excess of 30 years. Blood was also obtained from an age and sex matched control group and from patients with Metal-on-Polyethylene (MoP) components of the same era. The whole blood was cultured with Pb-Max karyotyping medium and harvested for cytogenetics after 72 hrs. The 24 colour FISH (Fluorescent In Situ Hybridisation) chromosome painting technique was performed on the freshly prepared slides, allowing chromosomal mapping. Each slide was evaluated for chromosomal aberrations (deletions, fragments and translocations) against the normal 46 (22 pairs and two sex) chromosomes. At least 20 metaphases per sample were scored and the number of aberrations per cell calculated. Results. Chromosomal aberrations, including deletions, fragments and translocations were only detected in the peripheral blood lymphocytes isolated from the group that had MoM bearings. These changes were not present in the age and sex matched control group. The chromosomal aberrations were also detected in the patients previously exposed to MoM bearings who have been revised to a MoP articulation. Conclusion. We have detected dramatic chromosomal aberrations in peripheral blood lymphocytes in a group of patients chronically exposed (over 30 years) to elevated metal ions. It is not known whether these aberrations have clinical consequences or whether they are reproduced in other cells in the body. The results emphasise the need for further investigations into the effect of chronic exposure to elevated metal ions produced by orthopaedic implants


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 363 - 364
1 Jul 2008
Dunstan E Ladon D Whittingham-Jones P Cannon S Case P Briggs T
Full Access

Background: Metal-on-Metal (MoM) hip bearings are being implanted in ever increasing numbers and into ever-younger patients. The consequence of chronic exposure to metal ions is a cause for concern. Therefore, by using cytogenetic biomarkers, we investigated a group of patients who have had MoM bearings in-situ for in excess of 30 years. Method: Whole blood specimens were obtained from an historical group of patients who have had MoM bearings in-situ for in excess of 30 years. Blood was also obtained from an age and sex matched control group and from patients with Metal-on-Polyethylene (MoP) components of the same era. The whole blood was cultured with Pb-Max karyotyping medium and harvested for cytogenetics after 72 h. The 24 colour FISH (Fluorescent In Situ Hybridisation) chromosome painting technique was performed on the freshly prepared slides allowing chromosomal mapping. Each slide was evaluated for chromosomal aberrations (deletions, fragments and translocations) against the normal 46 (22 pairs and two sex) chromosomes. At least 20 metaphases per sample were scored and the number of Aberrations per cell calculated. Results: Chromosomal aberrations, including deletions, fragments and translocations were only detected in the peripheral blood lymphocytes isolated from the group that had MoM bearings. These changes were not present in the age and sex matched control group. The chromosomal aberrations were also detected in the patients previously exposed to MoM bearings who have been revised to a MoP articulation. Conclusion: We have detected dramatic chromosomal aberrations in peripheral blood lymphocytes in a group of patients chronically exposed (over 30 years) to elevated metal ions. It is not known whether these aberrations have clinical consequences or whether they are reproduced in other cells in the body. The results emphasise the need for further investigations into the effect of chronic exposure to elevated metal ions produced by Orthopaedic implants


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 169 - 173
1 Feb 2007
Daniel J Ziaee H Pradhan C Pynsent PB McMinn DJW

This is a longitudinal study of the daily urinary output and the concentrations in whole blood of cobalt and chromium in patients with metal-on-metal resurfacings over a period of four years.

Twelve-hour urine collections and whole blood specimens were collected before and periodically after a Birmingham hip resurfacing in 26 patients. All ion analyses were carried out using a high-resolution inductively-coupled plasma mass spectrometer. Clinical and radiological assessment, hip function scoring and activity level assessment revealed excellent hip function.

There was a significant early increase in urinary metal output, reaching a peak at six months for cobalt and one year for chromium post-operatively. There was thereafter a steady decrease in the median urinary output of cobalt over the following three years, although the differences are not statistically significant. The mean whole blood levels of cobalt and chromium also showed a significant increase between the pre-operative and one-year post-operative periods. The blood levels then decreased to a lower level at four years, compared with the one-year levels. This late reduction was statistically significant for chromium but not for cobalt.

The effects of systemic metal ion exposure in patients with metal-on-metal resurfacing arthroplasties continue to be a matter of concern. The levels in this study provide a baseline against which the in vivo wear performance of newer bearings can be compared.