This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.Aims
Methods
This study aimed to develop and validate a fully automated system that quantifies proximal femoral bone mineral density (BMD) from CT images. The study analyzed 978 pairs of hip CT and dual-energy X-ray absorptiometry (DXA) measurements of the proximal femur (DXA-BMD) collected from three institutions. From the CT images, the femur and a calibration phantom were automatically segmented using previously trained deep-learning models. The Hounsfield units of each voxel were converted into density (mg/cm3). Then, a deep-learning model trained by manual landmark selection of 315 cases was developed to select the landmarks at the proximal femur to rotate the CT volume to the neutral position. Finally, the CT volume of the femur was projected onto the coronal plane, and the areal BMD of the proximal femur (CT-aBMD) was quantified. CT-aBMD correlated to DXA-BMD, and a receiver operating characteristic (ROC) analysis quantified the accuracy in diagnosing osteoporosis.Aims
Methods
Aims. Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane. Methods. Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation. Results. Overall, 26/40 hips had a centre of rotation displaced posteriorly compared to the contralateral hip, increasing to 33/40 once corrected for sagittal tilt, with a mean posterior displacement of 7 mm. Linear regression analysis indicated that stem anteversion needed to be increased by 10.8° to recreate the head centre in the AP plane. Merely matching the native version would result in a 12 mm posterior displacement. Conclusion. This study demonstrates the significant incidence of posterior displacement of the head centre in
Aims. There are concerns regarding initial stability and early periprosthetic fractures in
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
Currently, periprosthetic fractures are excluded from the American Society for Bone and Mineral Research (ASBMR) definition of atypical femoral fracture (AFFs). This study aims to report on a series of periprosthetic femoral fractures (PFFs) that otherwise meet the criteria for AFFs. Secondary aims were to identify predictors of periprosthetic atypical femoral fractures (PAFFs) and quantify the complications of treatment. This was a retrospective case control study of consecutive patients with periprosthetic femoral fractures between 2007 and 2017. Two observers identified 16 PAFF cases (mean age 73.9 years (44 to 88), 14 female patients) and 17 typical periprosthetic fractures in patients on bisphosphonate therapy as controls (mean age 80.7 years (60 to 86, 13 female patients). Univariate and multivariate analysis was performed to identify predictors of PAFF. Management and complications were recorded.Aims
Patients and Methods
The objective of this study was to evaluate the short term clinical and radiological results of a new short stem hip implant. In 29 consecutive patients suffering from osteoarthritis with 33 affected hip joints, the clinical and radiological results of 33
Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (Aims
Patients and Methods
Ceramic-on-ceramic bearings provide a solution to the osteolysis seen with traditional metal-on-polyethylene bearings. Sporadic reports of ceramic breakage and squeaking concern some surgeons and this bearing combination can show in vivo signs of edge loading wear which was not predicted from in vitro studies. Taper damage or debris in the taper between the ceramic and metal may lead to breakage of either a ceramic head or insert. Fastidious surgical technique may help to minimise the risk of ceramic breakage. Squeaking is usually a benign complication, most frequently occurring when the hip is fully flexed. Rarely, it can occur with each step of walking when it can be sufficiently troublesome to require revision surgery. The etiology of squeaking is multifactorial in origin. Taller, heavier and younger patients with higher activity levels are more prone to hips that squeak. Cup version and inclination are also relevant factors. Osteolysis following metal-on-UHMW polyethylene Total Hip Arthroplasty (THA) is well reported. Earlier generation ceramic-on-ceramic bearings did produce some osteolysis, but in flawed implants. As third and now fourth generation ceramic THAs come into mid- and long-term service, the orthopaedic community has begun to see reports of high survival rates and very low incidence of osteolysis in these bearings. The technique used by radiologists for identifying the nature of lesions on Computed Tomography (CT) scan is the Hounsfield score which will identify the density of the tissue within the lucent area. Commonly the radiologist will have no access to previous imaging, especially pre-operative imaging if a long time has elapsed. With such a low incidence of osteolysis in this patient group, what, then, should a surgeon do on receiving a CT report on a ceramic-on-ceramic THA, which states there is osteolysis? This retrospective review aims to determine the accuracy of CT in identifying true osteolysis in a cohort of long-term third generation ceramic-on-ceramic
Undersizing of an uncemented femoral stem is a post-operative description of a radiograph with unclear clinical importance. It may not always translate into a suboptimal clinical outcome. To describe a novel classification system of undersizing in a fully hydroxyapatite coated femoral implant, assess if it could be reliably reported between observers, and describe a simple way to assess the degree of undersizing on an AP radiograph.Introduction
Aim
There is an ongoing discussion on what bearing surfaces to use in different age groups of total hip replacement patients. We report results from uncemented total hip arthroplasty using ceramic on ceramic bearings reported to the Norwegian Arthroplasty Register in the period 1997–2014. Kaplan Meier and Cox regression analysis adjusting for age, gender and diagnosis was used to assess prosthesis survival at 10 and 15 years after primary operation and to report relative risk of revision. Endpoint was any revision. Comparison with the cemented Charnley prosthesis (n=17180), with metal-on-polyethylene articulations from the same time period was done. Results in age groups <55, 55–64, 65–74, >=75 were investigated. The following femur/cup combinations were used; Filler/Igloo (n=2590), Corail/Pinnacle (n=783), Corail/Duraloc (n=467), SCP Unique/Trilogy (n=363), Polar/R3 (n=253), ABGII/ABGII (n=131), other combinations (n=339). Alumina bearing was used in 3807 hips and Alumina-Zirconium composite bearings in 1119 hips. The overall ten and fifteen years unadjusted Kaplan Meier survivals were 94.8 % and 92.0 % respectively with no statistically significant difference between the brands. The cemented Charnley hip arthroplasty had 93.6 % survival at 15 years, and was not statistically significant different from the uncemented ceramic/ceramic group. The 10 years survival in the age groups <55, 55–64, 65–74, >=75 for the uncemented ceramic/ceramic group was 93.7%, 95.3%, 96.0% and 95.4% respectively with no difference between the uncemented brands and the cemented Charnley prosthesis. We found less revisions in patients >=75 in the Charnley group. In the ceramic on ceramic group 11 head fractures and 3 liner fractures were reported. The 15 years result of
Ceramic-on-ceramic bearings provide a solution to the osteolysis seen with traditional metal-on-polyethylene bearings. Sporadic reports of ceramic breakage and squeaking concern some surgeons and this bearing combination can show in vivo signs of edge loading wear which was not predicted from in vitro studies. Taper damage or debris in the taper between the ceramic and metal may lead to breakage of either a ceramic head or insert. Fastidious surgical technique may help to minimise the risk of ceramic breakage. Squeaking is usually a benign complication, most frequently occurring when the hip is fully flexed. Rarely, it can occur with each step of walking when it can be sufficiently troublesome to require revision surgery. The etiology of squeaking is multifactorial origin. Taller, heavier and younger patients with higher activity levels are more prone to hips that squeak. Cup version and inclination are also relevant factors. Fifty-five ceramic bearings revised at our center were collected over 12 years. Median time to revision was 2.7 years. Forty-six (84%) cases had edge loading wear. The median femoral head wear volume overall was 0.2mm. 3. /yr, for anterosuperior edge loading was 2.0mm. 3. /yr, and the median volumetric wear rate for posterior edge loading was 0.15mm. 3. /yr (p=0.005). Osteolysis following metal-on-polyethylene total hip arthroplasty (THA) is well reported. Earlier generation ceramic-on-ceramic bearings did produce some osteolysis, but in flawed implants. As 3rd and now 4th generation ceramic THAs come into mid- and long-term service, the orthopaedic community has begun to see reports of high survival rates and very low incidence of osteolysis in these bearings. The technique used by radiologists for identifying the nature of lesions on Computed Tomography (CT) scan is the Hounsfield score which will identify the density of the tissue within the lucent area. Commonly the radiologist will have no access to previous imaging, especially pre-operative imaging if a long time has elapsed. With such a low incidence of osteolysis in this patient group, what, then, should a surgeon do on receiving a CT report on a ceramic-on-ceramic THA, which states there is osteolysis? This retrospective review aims to determine the accuracy of CT in identifying true osteolysis in a cohort of long-term 3rd generation ceramic-on-ceramic
Introduction. Osteloysis following metal-on-UHMW polyethylene Total Hip Arthroplasty (THA) is well reported, as is lack of osteolysis following Ceramic-on-Ceramic (CoC)THA. Early ceramic failures did report some osteolysis, but in flawed implants. As 3rd and now 4th generation ceramic THAs come into mid- and long-term use, the orthopaedic community has begun to see reports of high survival rates and very low incidence of osteolysis in these bearings. Osteolysis reported after 3rd generation CoC THA often included metallosis due to neck rim impingement. In our department we have revised only 2 hips in over 6000 CoC THAs for osteolysis. Both had evidence of metallosis as well as ceramic wear. The technique used by Radiologists for identifying the nature of lesions on CT is the Hounsfield score which will identify the density of the tissue within the lucent area. It is common for radiologist to have no access to previous imaging, especially pre-operative imaging if a long time has elapsed. With such a low incidence of osteolysis in this patient group, what, then, should a surgeon do on receiving a CT report on a ceramic-on-ceramic THA which states there is osteolysis? Revision of such implants in elderly patients has a high risk of morbidity and mortality. Objectives. This retrospective review aims to determine the accuracy of CT in identifying true osteolysis in a cohort of long-term third generation ceramic-on-ceramic
The aim of this study was to evaluate the ten-year
clinical and functional outcome of hip resurfacing and to compare it
with that of
The October 2014 Hip &
Pelvis Roundup360 looks at: functional acetabular orientation; predicting re-admission following THR; metal ions and resurfacing; lipped liners increase stability; all anaesthetics equal in hip fracture surgery; revision hip surgery in very young patients; and uncemented hips.
Background:. Historically, performing a successful hip joint replacement in patients aged fewer than 30 years has been an orthopedic challenge. The newer generation of prostheses and surgical techniques has the potential to increase the longevity of implants. The purpose of this study was to evaluate the outcomes of