Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 45 - 45
19 Aug 2024
Perez SFG Zhao G Tsukamoto I Labott JR Restrepo DJ Hooke AW Zhao C Sierra RJ
Full Access

Previous studies have highlighted differences in the risk of periprosthetic fracture between tapered slip (TS) and composite beam (CB) stems. This biomechanical study explored periprosthetic fractures around these stems and the effect of adding a 16-gauge calcar or diaphyseal wire to TS stems on their resistance to torque.

A power analysis determined a sample size of 7 specimens per group, assuming a standard deviation of 14.8 Nm in peak torque, to provide 90% power to detect a difference of at least 30 Nm between groups. Twenty-one TS stems (eight alone, six with calcar wiring, seven with diaphyseal wiring placed 2 cm distal to the lesser trochanter) and seven CB stems were cemented into standard Sawbones. A servo-hydraulic test machine applied a 1000 N load with a 1-degree rotation per second until failure. The peak torque at failure was measured, and the fracture location recorded. Comparisons were performed using two-sample t-tests.

CB stems exhibited a significantly higher peak torque at failure (205.3 Nm) than TS stems (159.5 Nm, p=0.020). Calcar-wire-TS (148.2 Nm, p=0.036) and diaphyseal-wire-TS (164.9 Nm, p=0.036) were both weaker than CB stems. Wired-TS stems showed no significant difference from non-wired-TS stems. Additionally, the study could not conclude that calcar wiring is stronger than diaphyseal wiring. All TS fractures occurred at the mid-stem, simulating a B-type fracture, while the addition of the diaphyseal wire shifted the fracture location more distally in four of seven stems (p=0.0699).

This biomechanical study supports the clinical evidence that CB stems have stronger resistance to torque than TS stems and may explain lower risk of periprosthetic fracture. The addition of calcar or diaphyseal wires to TS stems resulted in no significant changes in peak torque to fracture. In patients at high risk of periprosthetic fracture, CB cemented stems should be considered.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 97 - 97
10 Feb 2023
Gibbons J Bodian C Powell A Sharr J Lash N
Full Access

PFFs are an increasing burden presenting to the acute trauma services. The purpose of this study is to show that cemented revision for Vancouver B2/B3 PFFs is a safe option in the geriatric population, allows early pain-free weight bearing and comparable to a control-group of uncemented stems with regard to return to theatre and revision surgery.

A retrospective review was conducted of all PFFs treated in a Level 1 trauma centre from 2015-2020. Follow up x-rays and clinical course through electronic chart was reviewed for 78 cemented revisions and 49 uncemented revisions for PFF. Primary endpoints were all cause revision and return to theatre for any reason. Secondary endpoints recorded mobility status and all-cause mortality.

In the cemented group there were 73 Vancouver B2, 5 Vancouver B3 PFF; the mean age was 79.7 years and mean radiological follow-up of 11.9 months. In the cementless group there were 32 Vancouver B2 and 17 Vancouver B3 PFFs; with all 49 patients undergoing distally bearing uncemented revision, the mean age was 72.7 years and mean radiological follow-up of 21.3 months.

Patients treated with a cemented prosthesis had significantly higher ASA score (2.94 -v- 2.43, p<0.001). The primary endpoints showed that there was no significant difference in all cause revision 3/78 and 5/49 p=0.077, or return to theatre 13/78 -v- 12/49 p=0.142.

Secondary endpoints revealed no significant difference in in-hospital mortality. The cementless group were more likely to be mobilising without any aid at latest follow-up 35/49 -v- 24/78 p<0.001.

The use of cemented revision femoral component in the setting of PFFs is one option in the algorithm for management of unstable PFFs according to the Vancouver classification. Evidence from this case-control study, shows that the all-cause revision and return to theatre for any cause was comparable in both groups.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 124 - 134
1 Feb 2023
Jain S Farook MZ Aslam-Pervez N Amer M Martin DH Unnithan A Middleton R Dunlop DG Scott CEH West R Pandit H

Aims

The aim of this study was to compare open reduction and internal fixation (ORIF) with revision surgery for the surgical management of Unified Classification System (UCS) type B periprosthetic femoral fractures around cemented polished taper-slip femoral components following primary total hip arthroplasty (THA).

Methods

Data were collected for patients admitted to five UK centres. The primary outcome measure was the two-year reoperation rate. Secondary outcomes were time to surgery, transfusion requirements, critical care requirements, length of stay, two-year local complication rates, six-month systemic complication rates, and mortality rates. Comparisons were made by the form of treatment (ORIF vs revision) and UCS type (B1 vs B2/B3). Kaplan-Meier survival analysis was performed with two-year reoperation for any reason as the endpoint.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 331 - 340
1 Mar 2022
Strahl A Kazim MA Kattwinkel N Hauskeller W Moritz S Arlt S Niemeier A

Aims

The aim of this study was to determine whether total hip arthroplasty (THA) for chronic hip pain due to unilateral primary osteoarthritis (OA) has a beneficial effect on cognitive performance.

Methods

A prospective cohort study was conducted with 101 patients with end-stage hip OA scheduled for THA (mean age 67.4 years (SD 9.5), 51.5% female (n = 52)). Patients were assessed at baseline as well as after three and months. Primary outcome was cognitive performance measured by d2 Test of Attention at six months, Trail Making Test (TMT), FAS-test, Rivermead Behavioural Memory Test (RBMT; story recall subtest), and Rey-Osterrieth Complex Figure Test (ROCF). The improvement of cognitive performance was analyzed using repeated measures analysis of variance.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1405 - 1413
1 Aug 2021
Ogura K Fujiwara T Morris CD Boland PJ Healey JH

Aims

Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model.

Methods

We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 458 - 462
1 Apr 2020
Limberg AK Tibbo ME Pagnano MW Perry KI Hanssen AD Abdel MP

Aims

Varus-valgus constrained (VVC) implants are often used during revision total knee arthroplasty (TKA) to gain coronal plane stability. However, the increased mechanical torque applied to the bone-cement interface theoretically increases the risk of aseptic loosening. We assessed mid-term survivorship, complications, and clinical outcomes of a fixed-bearing VVC device in revision TKAs.

Methods

A total of 416 consecutive revision TKAs (398 patients) were performed at our institution using a single fixed-bearing VVC TKA from 2007 to 2015. Mean age was 64 years (33 to 88) with 50% male (199). Index revision TKA diagnoses were: instability (n = 122, 29%), aseptic loosening (n = 105, 25%), and prosthetic joint infection (PJI) (n = 97, 23%). All devices were cemented on the epiphyseal surfaces. Femoral stems were used in 97% (n = 402) of cases, tibial stems in 95% (n = 394) of cases; all were cemented. In total, 93% (n = 389) of cases required a stemmed femoral and tibial component. Femoral cones were used in 29%, and tibial cones in 40%. Survivorship was assessed via competing risk analysis; clinical outcomes were determined using Knee Society Scores (KSSs) and range of movement (ROM). Mean follow-up was four years (2 to 10).


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 951 - 959
1 Aug 2019
Preston N McHugh GA Hensor EMA Grainger AJ O’Connor PJ Conaghan PG Stone MH Kingsbury SR

Aims

This study aimed to develop a virtual clinic for the purpose of reducing face-to-face orthopaedic consultations.

Patients and Methods

Anonymized experts (hip and knee arthroplasty patients, surgeons, physiotherapists, radiologists, and arthroplasty practitioners) gave feedback via a Delphi Consensus Technique. This consisted of an iterative sequence of online surveys, during which virtual documents, made up of a patient-reported questionnaire, standardized radiology report, and decision-guiding algorithm, were modified until consensus was achieved. We tested the patient-reported questionnaire on seven patients in orthopaedic clinics using a ‘think-aloud’ process to capture difficulties with its completion.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 18 - 18
1 Jun 2018
Taunton M
Full Access

Pathologic fractures about the hip are an uncommon, but increasingly prevalent, clinical scenario encountered by orthopaedic surgeons. These fractures about the hip usually necessitate operative management. Life expectancy must be taken into account in management, but if survivorship is greater than 1 month, operative intervention is indicated. Determination must be made prior to operative management if the lesion is a solitary or metastatic lesion. Imaging of the entire femur is necessary to determine if there are other lesions present. Bone lesions that have a large size, permeative appearance, soft tissue mass, and rapid growth are all characteristics that suggest an aggressive lesion. Biopsy of the lesion in coordination with the operative surgeon should be conducted if the primary tumor is unknown. Metastatic disease is much more common than primary tumors in the adult population. Many metastatic fractures in the intertrochanteric region, and all fractures in the femoral neck and head are an indication for hemiarthroplasty or total hip arthroplasty. Cemented femoral implants are generally indicated. This allows immediate weight bearing in a bone with compromised bone stock, thus reducing the risk of peri-operative fractures. Additionally, patients are often treated with radiation and/or chemotherapy, which may prevent proper osseointegration of an ingrowth femoral component. Highly porous ingrowth shells have been shown to provide reliable and durable fixation even in these situations. Management of a periacetabular pathologic fracture, particularly resulting in a pelvic discontinuity is a particularly challenging situation. Use of a highly porous acetabular component combined with an acetabular cage, a custom acetabular component, a cemented Harrington technique, or a primary acetabular reconstruction cage may be utilised. Patients with neoplastic disease are often at risk for infection and thromboembolic disease both from the disease and treatment. Pre-operative evaluation of nutrition status by measuring albumin and pre-albumin will give the surgeon insight. Additionally, dehydration is commonly seen in cancer patients, and adequate pre-operative optimization of fluids and electrolytes may reduce peri-operative complications from other organ systems


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 192 - 198
1 Feb 2017
Schmitz MWJL Bronsema E de Kam DCJ Gardeniers JWM Veth RPH Schreurs BW

Aims

We carried out a further study of the long-term results of the cemented Exeter femoral component in patients under the age of 40 with a mean follow-up of 13.6 years (10 to 20).

Patients and Methods

We reviewed our original cohort of 104 cemented Exeter stems in 78 consecutive patients with a mean age of 31 years (16 to 39). Only one patient was lost to radiological follow-up.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 31 - 31
1 Jun 2016
Westerman R Whitehouse S Howell J Hubble M Timperley A Wilson M
Full Access

Introduction

The Exeter Hip femoral component remains largely unchanged from the original design, introduced in 1970. It is a highly polished, modular, double tapered stem and has undergone various minor modifications to surface, modularity and most recently the taper; changed to the current V40TM design in 2000.

The effect of any design modification cannot easily be foreseen and greater emphasis is now placed on ensuring appropriate monitoring for such implants.

Methods

We present the results of the first 540 V40TM Exeter THAs performed in our Centre between December 2000 and May 2002. All patients were reviewed prospectively at 1, 5 and 10 years following surgery.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 51 - 51
1 Nov 2015
Highcock A Siney P Wroblewski B Jones HW Raut V
Full Access

Introduction. Severe osteolysis of the femur secondary to aseptic loosening in hip arthroplasty, remains a difficult revision scenario. Multiple techniques have been developed to aid the surgeon, including restoration of bone stock with impaction bone grafting or strut allografts, various distal fixation prostheses and mega-prostheses. Cemented femoral components, with integration of the cement into the cavitations, has largely fallen out of favour. We examined the long-term outcomes with this technique. Patients/Materials & Methods. Between 1977 and 1990, 109 patients had a cemented stem revision (without bone grafting) for severe femoral osteolysis in the absence of infection. Severe osteolysis was defined as cavitation in a minimum of 4 Gruen zones. Follow-up included functional scoring, radiological assessment and any complications. Further revision and survivorship analysis for stem failure or aseptic loosening of the femoral component were recorded. Results. 109 patients (117 hips) were reviewed with an average age of 65.4years (range 31–82). 39 patients had five zone involvement, 25 had six zone involvement and in 13 patients all seven zones were affected. 17 patients had an isolated stem revision, 100 patients had both components revised. Post-operative complications during the entire follow-up period were 7 dislocations, 10 trochanteric non-unions, 2 periprosthetic fractures and no infections. Only 9 patients had further revision of the femoral stem, 8 for aseptic loosening and 1 for stem fracture. In addition, re-revision procedures were performed for aseptic loosening of the cup (8 procedures) and dislocation (2 procedures). Analysis for stem revision demonstrated 91.17% survival at 12 years (minimum 40 hips at risk). Conclusion. Cemented revision arthroplasty for severe osteolysis of the femur has results comparable, if not superior, to alternative treatment modalities. Where restoration of bone stock is not a principle goal in the revision setting, cemented revision arthroplasty is a viable option in severe femoral osteolysis


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 36 - 36
1 Dec 2013
Cook R Shearwood-Porter N Nicolae C Bolland B Latham J Wood R
Full Access

Introduction:. Cemented femoral components have been used in hip replacement surgery since its inception. For many patients this works well, but recent retrieval studies. 1–4. and more fundamental studies. 5, 6. have highlighted the issues of damage and material loss from the both matt and polished cemented stems. Materials and methods:. This study will focus on a cohort of retrievals from the Southampton Orthopaedics Centre for Arthroplasty Retrieval Surgery (SOCARS). The cohort consisted of a number of hybrid modular total hip replacements with cemented femoral components, both from mixed and matched manufacturer stem and head combinations. Femoral stems were polished, collarless, tapered designs; head sizes ranged from 28–54 mm. For each femoral stem, samples of Palacos R + G cement (Heraeus Medical GmbH, Hanau, Germany) were retrieved from the proximal region of the cement mantle (Gruen zones 1 and 7), corresponding to both macroscopically damaged and undamaged surfaces of the stem. The areas of damage were determined using calibrated digital photography; damaged surfaces were then imaged in detail using an Alicona InfiniteFocus microscope (Alicona Imaging GmbH, Graz, Austria). The technique uses optical microscopy and focus variation technology to extract 3D morphology and depth information from the surface with a resolution of 10 nm. A series of measurements were made and two different analysis routes were used to provide volumetric material loss measurements from the stem surface. High-resolution microscopy and elemental analysis of the cement and stem surfaces was conducted via SEM and EDX to identify the mechanisms leading to material loss at the cement-stem interface. Results:. The results demonstrate that material loss from polished femoral stems results from a progressive tribocorrosion process; the major damage mechanism is thought to be the micro-motion between the femoral stem surface and zirconium dioxide radiopacifier agglomerates originating from the cement. No significant link was found between the extent of damage to the femoral stem and either the head size or the amount of wear occurring at the head-cup bearing surface. The scale of stem damage varied between implants but often exceeded the volumetric material loss measured at the bearing surfaces. Conclusions:. Tribo-corrosive damage to the femoral stems of cemented total hip prostheses is a major potential source of material loss in vivo; in severely affected arthroplasties, measurements of volumetric wear of the stem at the cement-stem interface were greater than at either the head-cup bearing surface or the taper junction. The mechanism of material loss in this study was identified as a wear-dominated tribocorrosion interaction between the cement and stem, with zirconium dioxide radiopacifier agglomerates within the cement providing the hard particles which damaged the surface of cobalt-chrome femoral stems


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 324 - 324
1 Dec 2013
Ginsel B Crawford R Whitehouse S Taher A
Full Access

Aim:

This prospective cohort study investigated whether the use of preoperative anticoagulants is an independent risk factor for the outcomes of surgical treatment of patients with a neck of femur fracture.

Methods:

Data was obtained from a prospectively collected database. All patients admitted for a neck of femur fracture between Nov 2010 and Oct 2011 were included. This resulted in three hundred twenty-eight patients with 330 neck of femur fractures. Four groups were defined; patients preoperatively (i) on aspirin (n = 105); (ii) on clopidogrel (n = 28); (iii) on warfarin (n = 30); and (iv) without any anticoagulation history (n = 167, the control group). The non-warfarin group included the aspirin group, clopidogrel group and the control group. Primary outcome was the in-hospital mortality. Secondary outcomes were the postoperative complications, return to theatre and length of stay.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 58 - 58
1 May 2012
Hubble M Williams D Crawford R Timperley J Gie G
Full Access

Favourable long-term results have been reported with the standard Exeter cemented stem. We report our experience with a version for use in smaller femora, the Exeter 35.5 mm stem. Although, also a collarless polished taper, the stem is slimmer and 25 mm shorter than a standard stem.

Between August 1988 and August 2003, 192 primary hip arthroplasties were performed in 165 patients using the Exeter 35.5 mm stem. Clinical and operative data were collected prospectively. Clinical evaluation was by the Charnley, Harris and Oxford hip scores and radiographs were analysed post-operatively and at latest follow up.

The mean age at time of operation was 53 years (18 to 86), with 73 patients under the age of 50 years. The diagnosis was osteoarthritis in 91, hip dysplasia in 77, inflammatory arthritis in 18, septic arthritis of the hip in three, secondary to Perthes disease in two and avascular necrosis of the hip in one patient. The fate of every implant is known.

At a median follow-up of 8 years (5 to 19), survivorship with revision of the femoral stem for aseptic loosening as the endpoint was 100%. Fifteen cases (7.8%) underwent further surgery 11 for acetabular revision, one for stem fracture and three others.

Although, smaller than a standard Exeter Universal polished tapered cemented stem—with a shorter, slimmer taper—the performance of the Exeter 35.5 mm stem was equally good even in this young, diverse group of patients.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_III | Pages 59 - 59
1 Feb 2012
Carrington N Sierra R Hubble M Gie G Ling R Howell J
Full Access

Purpose

We describe an update of our experience with the implantation of the first 325 Exeter Universal hips. The fate of every implant is known.

Methods and results

The first 325 Exeter Universal stems (309 patients) were inserted between March 1988 and February 1990. The procedures were undertaken by surgeons of widely differing experience. Clinical and radiological review was performed at a mean of 15.7 years. At last review 185 patients had died (191 hips). 103 hips remain in situ. Survivorship at 17 years with revision for femoral component aseptic loosening was 100% (95% CI 97 to 100), with revision for acetabular component aseptic loosening was 90.4% (95% CI 83.1 to 94.7) and with any re-operation as the endpoint was 81.1% (95% CI 72.5 to 89.7). 12 patients (12 hips) were not able to attend for review due to infirmity or emigration, and scores were obtained by phone (x-rays were obtained in 4 patients). Mean D'Aubigné and Postel scores (Charnley modification) at review were 5.4 for pain and 4.8 for function. The mean Oxford score was 21.6 +/− 9.8 and the mean Harris score 71.7 +/− 19.7. On radiological review there were no femoral component failures. Three sockets (2.9%) were loose as demonstrated by migration or change in orientation (two patients were asymptomatic) and 5 sockets (4.9%) had radiolucent lines in all 3 zones but no migration. There are two patients awaiting socket revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 665 - 677
1 May 2011
Sköldenberg OG Salemyr MO Bodén HS Lundberg A Ahl TE Adolphson PY

Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck.

We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years.

Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D.

Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years.

The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 45 - 45
1 Jan 2011
Williams D Howell J Hubble M Timperley A Gie G
Full Access

Survivorship of the standard Exeter Universal cemented stem with revision of the femoral component for aseptic loosening as the endpoint has been reported as 100% at 12 years. A version for use in smaller femora, the Exeter 35.5 mm stem, was introduced in 1988. Although also a collarless polished taper, the stem is slimmer and 25 mm shorter than a standard stem.

Between August 1988 and August 2003 192 primary hip arthroplasties were performed in 165 patients using the Exeter 35.5 mm stem. Clinical and operative data were collected prospectively. Clinical evaluation was by the Charnley, Harris and Oxford hip scores and radiographs were analysed post-operatively and at latest follow up.

The mean age at time of operation was 53 (18 to 86) years with 73 patients under the age of 50 years. The diagnosis was osteoarthritis 91, hip dysplasia in 77, inflammatory arthritis in 18, septic arthritis of the hip in 3, secondary to Perthes disease in 2 and avascular necrosis of the hip in 1 patient. The fate of every implant is known.

At a median follow-up of 8 (5 to 19) years survivorship with revision of the femoral stem for aseptic loosening as the endpoint was 100%. 15 cases (7.8%) underwent further surgery – 11 for acetabular revision, 1 for stem fracture and 3 others.

Although smaller than a standard Exeter Universal polished tapered cemented stem, with a shorter, slimmer taper, the performance of the Exeter 35.5 mm stem was equally good even in this young, diverse group of patients.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 300 - 300
1 May 2010
Hamadouche M Baqué F Lefevre N Kerboull L Kerboull M Courpied J
Full Access

Introduction: The purpose of this study was to report on the minimal 10-year followup results of a prospective randomized and a historical series of low friction cemented hip arthroplasties according to the surface finish of the femoral implant.

Patients and Methods: The prospective randomized series included 284 patients (310 hips) with a mean age of 64.1 years. Among these 310 hips, the femoral component had a highly polished surface (Ra = 0.04 micron, MKIII, Stryker) in 165 hips, and a matte surface finish (Ra = 1.7 microns, CMK3, Vector Orthopedique) in the remaining 145 hips. The historical series that was operated by the sames surgeons according to the same surgical technique included 111 patients (123 hips) with a satin finish femoral component (Ra = 0.9 micron, CMK2, Sanortho). Clinical results were rated according to the Merle d’Aubigne hip score. Radiologic analysis was performed according to the criteria of Barrack et al. for the definition of loosening. Moreover, a survival analysis according to the actuarial method was conducted.

Results: At the minimum 10-year follow-up evaluation, 43 patients (48 hips) were lost to follow-up (0.3 to 8.7 years), 80 patients (83 hips) were deceased (0.1 to 13.6 years), 26 patients had revision of either or both components (0.9 to 15.9 years), and 246 patients (276 hips) were alive and had not been revised after a mean 12.3 ± 1.9 years (10 to 16 years). Radiologic loosening of the femoral component, including revision, occured in one, four, and 15 hips for the polished, stain, and matte groups, respectively. The survival rate at 13 years of the femoral component, using loosening as the end point, was 97.3 ± 2.6% (95% CI, 92.2 to 100%), 97.1 ± 2.1% (95% CI, 93 to 100%), and 78.9 ± 5.8% (95% CI, 67.6 to 90.3%) for polished, satin, and matte stems, respectively.

Discussion: This study demonstrated that cement fixation of a femoral component was more reliable in the long term with a polished or satin surface finish. Based upon our results and the review of the literature, we recommend abondoning the use of cemented stems with a surface roughness greater than 1 micron.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 98 - 99
1 Mar 2010
Sochart D Chaudhary N
Full Access

Cemented acetabular components continue to be used in more than half of the total hip replacements performed in the United Kingdom. The implants are relatively inexpensive but the results rely heavily on precise surgical technique, with restoration of the centre of rotation of the hip and the creation of an equal cement mantle with good pressurisation and penetration of the cement into the bone. The Opera all polyethylene acetabular component was designed with a malleable flange, which could be independently pressurised, a long posterior wall to promote stability and instrumentation which ensured that pressurisation could be maintained throughout the curing process. We present for the first time, the medium-term results of 409 consecutive cemented flanged Opera acetabular components performed in 374 patients. 247 operations were performed via a trochanteric osteotomy and 162 via a posterior approach, using multiple key holes, cement pressurisation and Palacos-R cement. Autograft was used in 32 cases. Cemented femoral components were used in all cases. There were 241 Charnley stems with 22mm heads and 168 polished triple tapered C-stems (146 with 22mm heads and 22 with 26mm heads). The average age at the time of surgery was 68.2 years (32–87) and the average duration of follow-up was 89 months (60–130). 54 patients (56 hips) died during the follow-up period. The acetabulum was assessed using the zones of DeLee and Charnley, and the Hodgkinson classification. There was one temporary femoral nerve palsy, two dislocations and 3 non-fatal pulmonary emboli. Both components were revised in two hips for deep sepsis, and in two hips only the femoral implant was revised, one for a fractured stem and one for aseptic loosening, with the original acetabular components remaining in situ. Of the remaining original acetabular components 6% have shown progressive radiological demarcation, none have migrated, but in two hips there is evidence of rapid wear and the development of osteolysis


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1422 - 1427
1 Nov 2008
Utting MR Raghuvanshi M Amirfeyz R Blom AW Learmonth ID Bannister GC

We have reviewed 70 Harris-Galante uncemented acetabular components implanted as hybrid hip replacements with cemented stems between 1991 and 1995 in 53 patients whose mean age was 40 years (19 to 49). The mean follow-up was for 13.6 years (12 to 16) with no loss to follow-up. We assessed the patients both clinically and radiologically.

The mean Oxford hip score was 20 (12 to 46) and the mean Harris hip score 81 (37 to 100) at the final review. Radiologically, 27 hips (39%) had femoral osteolysis, 13 (19%) acetabular osteolysis, and 31 (44%) radiolucent lines around the acetabular component. Kaplan-Meier survival curves were constructed for the outcomes of revision of the acetabular component, revision of the component and polyethylene liner, and impending revision for progressive osteolysis.

The cumulative survival for revision of the acetabular component was 94% (95% confidence interval 88.4 to 99.7), for the component and liner 84% (95% confidence interval 74.5 to 93.5) and for impending revision 55.3% (95% confidence interval 40.6 to 70) at 16 years.

Uncemented acetabular components with polyethylene liners undergo silent lysis and merit regular long-term radiological review.