Source of the study: University of Auckland, Auckland, New Zealand and University of Otago, Christchurch, New Zealand. Outcomes following knee arthroplasty are typically defined as implant survivorship at defined timepoints, or revision incidence over time. These estimates are difficult to conceptualise, and lack context for younger patients with more remaining years of life. We therefore aimed to determine a ‘lifetime’ risk of revision as a more useful metric for total (TKA) and unicompartmental knee arthroplasty (UKA). The New Zealand Joint Registry was used to identify 96,497 primary TKAs and 13,481 primary UKAs performed between 1999 and 2019. Patient mortality and revision incidence were also extracted. Estimates of lifetime risk were calculated using an
Unicompartmental knee arthroplasty (UKA) has a higher risk of revision than total knee arthroplasty (TKA), particularly for younger patients. The outcome of knee arthroplasty is typically defined as implant survival or revision incidence after a defined number of years. This can be difficult for patients to conceptualize. We aimed to calculate the ‘lifetime risk’ of revision for UKA as a more meaningful estimate of risk projection over a patient’s remaining lifetime, and to compare this to TKA. Incidence of revision and mortality for all primary UKAs performed from 1999 to 2019 (n = 13,481) was obtained from the New Zealand Joint Registry (NZJR). Lifetime risk of revision was calculated for patients and stratified by age, sex, and American Society of Anesthesiologists (ASA) grade.Aims
Methods
The success of total knee arthroplasty (TKA) is usually measured using functional outcome scores and revision-free survivorship. However, reporting the lifetime risk of revision may be more meaningful to patients when gauging risks, especially in younger patients. We aimed to assess the lifetime risk of revision for patients in different age categories at the time of undergoing primary TKA. The New Zealand Joint Registry database was used to obtain revision rates, mortality, and the indications for revision for all primary TKAs performed during an 18-year period between January 1999 and December 2016. Patients were stratified into age groups at the time of the initial TKA, and the lifetime risk of revision was calculated according to age, sex, and the American Society of Anesthesiologists (ASA) grade. The most common indications for revision were also analyzed for each age group.Aims
Methods
Joint registries typically use revision of an implant as an endpoint and report survival rates after a defined number of years. However, reporting lifetime risk of revision may be more meaningful, especially in younger patients. We aimed to assess lifetime risk of revision for patients in defined age groups at the time of primary surgery. The New Zealand Joint Registry (NZJR) was used to obtain rates and causes of revision for all primary total hip arthroplasties (THAs) performed between January 1999 and December 2016. The NZJR is linked to the New Zealand Registry of Births, Deaths and Marriages to obtain complete and accurate data. Patients were stratified by age at primary surgery, and lifetime risk of revision calculated according to age, sex, and American Society of Anesthesiologists (ASA) classification. The most common causes for revision were also analyzed for each age group.Aims
Methods
The purpose of this retrospective study was to evaluate the minimum
five-year outcome of revision total hip arthroplasty (THA) using
the Kerboull acetabular reinforcement device (KARD) in patients
with Paprosky type III acetabular defects and destruction of the
inferior margin of the acetabulum. We identified 36 patients (37 hips) who underwent revision THA
under these circumstances using the KARD, fresh frozen allograft
femoral heads, and reconstruction of the inferior margin of the
acetabulum. The Merle d’Aubigné system was used for clinical assessment.
Serial anteroposterior pelvic radiographs were used to assess migration
of the acetabular component.Aims
Patients and Methods
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
Introduction: The purpose of this study was to report on the minimal 10-year followup results of a prospective randomized and a historical series of low friction cemented hip arthroplasties according to the surface finish of the femoral implant. Patients and Methods: The prospective randomized series included 284 patients (310 hips) with a mean age of 64.1 years. Among these 310 hips, the femoral component had a highly polished surface (Ra = 0.04 micron, MKIII, Stryker) in 165 hips, and a matte surface finish (Ra = 1.7 microns, CMK3, Vector Orthopedique) in the remaining 145 hips. The historical series that was operated by the sames surgeons according to the same surgical technique included 111 patients (123 hips) with a satin finish femoral component (Ra = 0.9 micron, CMK2, Sanortho). Clinical results were rated according to the Merle d’Aubigne hip score. Radiologic analysis was performed according to the criteria of Barrack et al. for the definition of loosening. Moreover, a survival analysis according to the
We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component. At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.
We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30. There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.
The topic of this study was to research the survival rate of ATLAS hip prosthesis (acetabular cup) performed by one operating surgeon only and with a minimum of 10 years follow up. Material and method: The ATLAS cup was hemispheric, in titanium alloy TA6V4, with a thin layer (2.5mm). It had a large central orifice as well as a radial slot allowing a certain elasticity enabling a pressfit impaction. Between January 1989 and December 1995, 297 ATLAS were implanted in the Clinique Saint André of Reims, 171 ATLAS II non hydroxyapatite coated and 126 ATLAS III hydroxyapatite coated. There were 176 women (59,2%) and 121 men (40,8%). The average age was 66 (20 to 94). Each patient was contacted by phone to find out if the prosthesis was still in place or whether a new operation had been performed. For the deceased patients, the family or the usual doctor were contacted by phone to answer the question with a maximum of details. The non-parametric survival rates were performed using the
We analysed at a mean follow-up of 7.25 years the clinical and radiological outcome of 117 patients (125 knees) who had undergone a primary, cemented, modular Freeman-Samuelson total knee replacement. While the tibial and femoral components were cemented, the patellar component was uncemented. A surface-cementing technique was used to secure the tibial components. A total of 82 knees was available for radiological assessment. Radiolucent lines were seen in 41 knees (50%) and osteolytic lesions were seen in 13 knees (16%). Asymptomatic, rotational loosening of the patellar implant was seen in four patients and osteolysis was more common in patients with a patellar resurfacing. Functional outcome scores were available for 41 patients (41 knees, 35%) and the mean Western Ontario McMasters Universities score was 77.5 (
Purpose: The purpose of this retrospective study was to evaluate the clinical and radiological results of Charnley-Keroboull total hip arthroplasty performed in patients aged less than 50 years. We searched for factors which might affect wear and sustained fixation. Material and methods: Among the 2,804 arthroplasties performed in patients aged less than 50 years between 1975 and 1995, we selected randomly 287 (10% of the annual operations). These prostheses were implanted in 222 patients (144 women and 78 men), mean age 40.1±8 years (15–50). All of the arthroplasties were inserted via a transtrochanteric approach. Charnley-Kerboull implants were cemented in all patients using a metal polyethylene bearing. Functional outcome was assessed with the Postel-Merle-d’Aubigné score. Cup wear was measured with the Chevrot technique. The
Purpose: Among 115 cemented patellofemoral Bechtol-Blazina prostheses implanted between 1974 and 1991, we reviewed 59 knees in 41 women and nine men (nine bilateral implants). Material and methods: Mean patient age was 60 years and mean follow-up was ten years (6–16). The main indication was primary degenerative disease (82%) followed by chondrocalcinosis (6%), chondromalacia (7%) and secondary degeneration (5%). A previous procedure had been performed on 18 knees (14 patellofemoral and four femorotibial). Associated procedures were tubercle translocation (n=15), tibial osteotomy (n=2), and tension of the medial wing (n=1). The International Knee Society score was noted. Results: According to the IKS, outcome was good or excellent in 91% of the knees and insufficient in 9% (knee score and function score). Mean ten-year survival determined with the
A prospective nation-wide study of knee arthroplasty has been under way in Sweden since October 1975. By the end of 1983, 4505 arthroplasties for osteoarthritis and 3495 for rheumatoid arthritis had been recorded and reviewed one, three and six years after the operation. Using
Seventy-one Geomedic total knee replacements were reviewed retrospectively at a maximum of eight and a half years. The prosthesis was found to be successful in its primary aim of relieving pain. However, there was an overall failure rate of 18.3%. Analysis of the results by an
The risk factors associated with 34 deep infections from a consecutive series of 803 total hip replacements have been analysed using standard
The survivorship of total hip replacements in patients operated on at the Royal National Orthopaedic Hospital has been analysed using standard