Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 72 - 72
23 Jun 2023
Greenwald AS
Full Access

Advances in total hip and knee replacement technologies have heretofore been largely driven by corporate marketing hype with each seeming advancement accompanied by a cost increase often out in front of peer-reviewed reports documenting their efficacy or not.

As example, consider the growing use of ceramic femoral heads in primary total hip arthroplasty (THA). The question to consider is “Can an upcharge of $350 for a ceramic femoral head in primary THA be justified?” The answer to this question lies in an appreciation of whether the technology modifies the potential for costly revision arthroplasty procedures.

Peer-Reviewed Laboratory & Clinical Review - According to the 2022 Australian National Joint Replacement Registry, the four leading causes of primary THA failure requiring revision are: 1.) infection, 2.) dislocation/instability, 3.) periprosthetic fracture and 4.) loosening, which constitute 87.5% of the reported reasons for revision.

Focusing on these failure modes, hip simulator findings report that ceramic femoral heads dramatically reduce wear debris generation, decreasing the potential for osteolytic response leading to loosening. Further, ceramic materials enable the utilization of larger head sizes, avoiding the potential for dislocation. The overall mid- to long-term survival rate reported in the peer-reviewed, clinical literature for these bearings has exceeded 95% with virtually no osteolysis.

Also, could bearing surface choice influence periprosthetic joint infection (PJI)? A study on a total of more than 10,500 primary THA procedures reported a confirmed PJI incidence of 2.4% for cobalt-chrome and 1.6% for ceramic femoral heads, suggesting that the employ of a ceramic bearing surface may also play a role in decreasing the potential for infection.

Review of the clinical data available for ceramic bearings justifies that it is better to “pay me now than to pay orders of magnitude later”, if in fact a revision THA can be avoided, significantly reducing the overall financial burden to the healthcare system.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 56 - 56
23 Jun 2023
Sugano N Maeda Y Fuji H Tamura K Nakamura N Takashima K Uemura K Hamada H
Full Access

The purposes of this study were to report the accuracy of stem anteversion for Exeter cemented stems with the Mako hip enhanced mode and to compare it to Accolade cementless stems.

We reviewed the data of 25 hips in 20 patients who underwent THA through the posterior approach with Exeter stems and 25 hips in 19 patients with Accolade stems were matched for age, gender, height, weight, disease, and approaches. There was no difference in the target stem anteversion (20°–30°) between the groups. Two weeks after surgery, CT images were taken to measure stem anteversion.

The difference in stem anteversion between the plan and the postoperative CT measurements was 1.2° ± 3.8° (SD) on average with cemented stems and 4.2° ± 4.2° with cementless stems, respectively (P <0.05). The difference in stem anteversion between the intraoperative measurements and the postoperative CT measurements was 0.75° ± 1.8° with Exeter stems and 2.2° ± 2.3° with Accolade stems, respectively (P <0.05).

This study demonstrated a high precision of anteversion for Exeter cemented stems with the Mako enhanced mode and its clinical accuracy was better with the cemented stems than that with the cementless stems. Although intraoperative stem anteversion measurements with the Mako system were more accurate with the cemented stems than that with the cementless stem, the difference was about 1° and the accuracy of intra-operative anteversion measurements was quite high even with the cementless stems. The smaller difference in stem anteversion between the plan and postoperative measurements with the cemented stems suggested that stem anteversion control was easier with cemented stems under the Mako enhanced mode than that with cementless stems.

Intraoperative stem anteversion measurement with Mako total hip enhanced mode was accurate and it was useful in controlling cemented stem anteversion to the target angle.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 55 - 55
23 Jun 2023
Alqazzaz A Bush A Zhuang T Nelson CL
Full Access

Surgical management of acetabular fractures in older patients remains controversial. The purpose of this study is to compare outcomes of primary THA with outcomes after THA for acute acetabular fractures (aTHA) as well as outcomes following delayed THA (dTHA) following prior acetabular fracture.

We analyzed data from a large, national administrative claims database containing diagnostic, procedural, and demographic records from over 157 million patients. We identified all patients undergoing primary total hip arthroplasty THA continuously enrolled in the database at least 2 years prior and after the index procedure. Patients with an initial diagnostic code for acetabular fracture occurring the same day as the THA were classified as acute acetabular fractures. Patients with an initial acetabular fracture diagnostic code occurring at least 6 months before THA were classified as chronic acetabular fractures. The comparator group was patients undergoing THA with no history of acetabular fracture. There were 426,734 patients undergoing primary THA, 235 patients undergoing aTHA and 1,255 patients undergoing dTHA.

Patients with an aTHA had higher complication rates, including revisions (9.8% vs 5.6%,), dislocations (8.9% vs 6.4%), and periprosthetic fracture (5.1% vs 2.3%) compared to dTHA. After adjusting for age, sex, region, and comorbidities, receiving an aTHA increased the odds of revision (OR = 3.65 [95% CI: 2.30–5.49]), dislocation (OR = 4.09 [95% CI: 2.53–6.27]), and periprosthetic fracture (OR = 4.29 [95% CI: 2.26–7.36]) compared to primary THA. Receiving a dTHA significantly increased the odds of revision (adjusted OR = 1.80 [95% CI: 1.40–2.27]), dislocation (adjusted OR = 2.50 [95% CI: 1.97–3.13]), and periprosthetic fracture (adjusted OR = 1.99 [95% CI: 1.34–2.83]) compared to primary THA.

Patients undergoing aTHA in the treatment of an acetabular fracture have significantly increased rates of revision, periprosthetic fracture, and dislocation compared to dTHA and primary THA.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 18 - 18
23 Jun 2023
Morlock M Melsheimer O
Full Access

The early revision rate in elective Total Hip Arthroplasty (THA) three years after surgery in elderly patients over 80 years is significantly lower for cemented stems in the German Arthroplasty Register (EPRD): cemented 3,1% (3.0 – 3.2) vs. uncemented 4.2% (4.1 – 4.3; p < 0.001). However, the mortality rate in elderly patients is elevated for cemented fixation. This study presents a detailed analysis of the influence of stem type and fixation on revision and mortality rate in this patient cohort.

Elective primary THA cases for primary Coxarthrosis using uncemented cups from the EPRD data base were analysed (n0= 37,183). Four stem type groups were compared: cementless, cementless with collar, cementless short, and cemented. Stems with at least 300 cases at risk three years after surgery were analysed individually. The reference stem was determined as the stem with the lowest revision rate and at least 1000 cases under surveillance 3 years after surgery (n3 = 28,637).

The revision rate for cemented stems (2.5% [2.2–1.81] was lower than for uncemented (4.5% [4.2–4.9]; p<0.001) and uncemented short stems (4.2% [3.1–5.7]; p=0.002). The revision rate of uncemented collared stems (2.3% [1.5–3.6]) was similar to cemented stems (p=0.89) and lower than for uncemented stems (p=0.02). One year mortality showed no sig. differences between the groups (p>0.17): cemented 3.2% [2.9–3.6], uncemented 3.4% [3.1–3.7], uncemented short 3.5% [2.5–4.9], uncemented collar 2.0% [1.2–3.2].

“Cementless” and “cementless short” stems should not be used in patients over 80 years due to the higher revision risk. If cementing should be avoided, “cementless collared” stems seem to be a good alternative combined with a tendency for a lower one year mortality rate.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 25 - 25
23 Jun 2023
Ricard M Pacheco L Koorosh K Poitras S Carsen S Grammatopoulos G Wilkin G Beaulé PE
Full Access

Our understanding of pre-arthritic hip disease has evolved tremendously but challenges remain in categorizing diagnosis, which ultimately impacts choice of treatments and clinical outcomes. This study aims to report patient reported outcome measures (PROMs) comparing four different condition groups within hip preservation surgery by a group of fellowship-trained surgeons.

From 2018 to 2021, 380 patients underwent hip preservation surgery at our center and were classified into five condition groups: dysplasia: 82 (21.6%), femoro-acetabular impingement (FAI): 173 (45.4%), isolated labral tear: 103 (27.1%), failed hip preservation: 20 (5.3%) and history of childhood disease/other: 2 (0.5%). International hip outcomes Tool 12 (IHOT-12), numeric pain score and patient-reported outcomes measurement information system (PROMIS) were collected pre-operatively and at 3 months and 1 year post-operatively, with 94% and 82% follow-up rate respectively.

Arthroscopy (75.5%) was the most common procedure followed by peri-acetabular osteotomy (PAO) (22.4%) and surgical dislocation (2.1%). Re-operation rate were respectively 18.3% (15), 5.8% (10), 4.9% (5), 30% (6) and 0%. There were 36 re-operations in the cohort, 14 (39%) for unintended consequences of initial surgery, 10 (28%) for mal-correction leading to a repeat operation, 8 (22%) progression of arthritis, and 4 (11%) for incorrect initial diagnosis/intervention. Most common re-operations were hardware removal 31% (7 PAO, 3 surgical hip dislocation and 1 femoral de-rotational osteotomy), arthroscopy 31% (11) and arthroplasty 28% (10). All groups had significant improvements in their IHOT-12 as well as PROMIS physical and numerical pain scales, except those with failed hip preservation. Dysplasia group showed a slower recovery.

Overall, this study demonstrated a clear relation between the condition groups, their respective intervention and the significant improvements in PROMs with isolated labral pathology being a valid diagnosis. Establishing tertiary referral centers for hip preservation and longer follow-up is needed to monitor the overall survivorship of these various procedures.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 2 - 2
23 Jun 2023
Hube R Zimmerer A Nonnenmacher L Najfeld M Fischer M Wassilew GI
Full Access

The use of trabecular metal (TM.) shells supported by two TM augments in the footing technique has been described as a potential option for the treatment of Paprosky 3B acetabular defects. The aim of this study was to assess the mid implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

We undertook a retrospective, double-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the footing technique using a TM shell supported by two TM augments, for severe acetabular bone loss between 2007 and 2020. The mean age at the time of surgery was 62,9 (28 to 86) years. The mean follow-up was 5,4 (1,5 to 15) years.

The cumulative mid survivorship of the implant with revision for any cause was 89%. 3 hips (7,6%) required further revision due to aseptic loosening, and 1 (2,8%) required revision for infection. The mean Harris Hip Score improved significantly from 48 (29 to 65) preoperatively to 79 points (62 to 98) at the latest follow-up (p < 0.001).

The reconstruction of Paprosky 3B acetabular defects with TM shells and two augments in footing-technique showed excellent mid-term results. This technique appears to be a viable option for treating these defects.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 40 - 40
23 Jun 2023
Millis MB Vakulenko-Lagun B Almakaris R Kim HJ
Full Access

LCPD can cause femoral head deformity and osteoarthritis requiring total hip replacement (THR). Currently, there is little data on how patients are functioning after a THR from patients’ perspective. The purpose of this study was to collect a large patient-reported outcome data set on adults with LCPD, including those who had a THR, using a Web-survey method and to compare their outcomes to a normative population.

An English REDCap-based survey was built and made available on a LCPD study group website. The survey included childhood and adult LCPD history, SF-36 Health Survey, and the Hip Disability and Osteoarthritis Outcome Score (HOOS). Statistical analysis included t-test and linear and proportional odds regressions.

Of the 1182 participants who completed the survey, 261 participants (89 M, 172 F) had a THR. The mean age at survey was 44.6±12.4 years (range 20–79). The mean duration since THR was 7.2±8 years (median 4, range 0–43). Gender and age matched analysis showed that THR participants had significantly lower HOOS Quality-of-Life and Sports scores (p<0.0001) for all age groups in comparison to a normative cohort. In women, the HOOS Symptoms, Daily Living, and Pain scores were also significantly lower in the <55 age groups (p<0.05). Similarly, SF-36 scores were significantly lower (p<0.05) in female <45 age groups in 5 out of 8 SF-36 scales. Overall, hip dysplasia and the number of years-from-THR were the main factors associated with worse SF-36 and HOOS scores. In comparison to the non-THR participants, THR participants had higher scores in some of the HOOS and SF-36 scales.

LCPD participants with THR had significantly worse HOOS and SF-36 scores in most of the scales studied than a normative cohort, especially in women. There is significant disability even after a THR, warranting continued efforts to improve treatment and outcome.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 44 - 44
23 Jun 2023
Scholz J Perka C Hipfl C
Full Access

Dual-mobility (DM) bearings are effective to mitigate dislocation in revision total hip arthroplasty (THA). However, data on its use for treating dislocation is scarce. Aim of this study was to compare DM bearings, standard bearings and constrained liner (CL) in revision THA for recurrent dislocation and to identify risk factors for re-dislocation.

We reviewed 100 consecutive revision THAs performed for dislocation from 2012 and 2019. 45 hips (45%) received a DM construct, while 44 hips (44%) and 11 hips (11%) had a standard bearing and CL, respectively. Rates of re-dislocation, re-revision for dislocation and overall re-revision were compared. Radiographs were assessed for cup positioning, restoration of centre of rotation, leg length and offset. Risk factors for re-dislocation were determined by cox regression analysis. Modified Harris hip scores (mHHS) were calculated. Mean follow-up was 53 months (1 to 103).

DM constructs were used more frequently in elderly patients (p=0.011) and hips with abductor deficiency (p< 0.001). The re-dislocation rate was 11.1% for DM bearings compared with 15.9% for standard bearings and 18.2% for CL (p=0.732). Revision-free survival for DM constructs was 83% (95% CI 0.77 – 0.90) compared to 75% (95% CI 0.68 – 0.82) for standard articulations and 71% (95% CI 0.56 – 0.85) for CL (p=0.455). Younger age (HR 0.91; p=0.020), lower comorbidity (HR 0.42; p=0.031), smaller heads (HR 0.80; p=0.041) and cup retention (HR 8.23; p=0.022) were associated with re-dislocation. Radiological analysis did not reveal a relationship between restoration of hip geometry and re-dislocation. mHHS significantly improved from 43.8 points to 65.7 points (p<0.001) with no differences among bearing types.

Our findings suggest that DM bearings do not sufficiently prevent dislocation in revision THA for recurrent dislocation. Reconstruction of the abductor complex may play a key role to reduce the burden in these high-risk patients.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 47 - 47
23 Jun 2023
Garbuz DS
Full Access

The purpose of this study was to determine long-term survival free from all-cause revision and stem-related failure, as well as radiographic and functional outcomes of the ZMR stem in revision THA.

We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR Revision Hip system from the year 2000 to 2007 (minimum two-year follow-up). Of the 112 ZMR hips (110 patients) identified, a total of 106 hips (104 patients) met inclusion criteria. The mean study follow-up 13.9 years (range 2–22). Indications for index ZMR revision included aseptic loosening (72.1%), periprosthetic joint infection (17.3%), periprosthetic fracture (9.6%), and stem fracture (1.0%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. Validated patient reported outcomes were collected and available radiographs were reviewed to determine implant stability.

A total of 17 hips (16.0%) underwent a re-revision of any component. The indications for re-revision were stem failure (35.3%), infection (29.4%), instability (29.4%), and aseptic loosening of the acetabular component (5.9%). The five- and 15-year all-cause survivorship was 89.5% (95% CI 86.5–92.5) and 83.0% (95% CI 79.2–86.8), respectively. There were six re-revisions for stem failure (5.7%); five for stem fracture and one for aseptic loosening. The average time to stem failure was 4.6 years (range 1.3–8.2). The five- and 15-year survivorship free from stem-related failure was 97.2% (95% CI 95.6–98.8) and 93.9% (95% CI 91.5–96.3), respectively. At final follow-up the mean Oxford hip score was 36.9 and all surviving ZMR stems were well fixed on radiographs.

Femoral revision with the ZMR stem offers satisfactory long-term survivorship and promising clinical outcomes. Although uncommon, stem fracture was the most common reason for stem-related failure.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 48 - 48
23 Jun 2023
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP
Full Access

There is a paucity of long-term data on modular fluted tapered (MFT) stems for two-stage reimplantation following periprosthetic joint infection (PJI). The purpose of this study was to evaluate implant survivorship, radiographic results, and clinical outcomes in a large cohort of reimplantation THAs using MFT stems.

We identified 236 reimplantation THAs from a single tertiary care academic institution from 2000 to 2020. Two designs of MFT stems were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years, mean BMI was 32 kg/m2, and 46% were female. Median stem diameter was 19 mm, and median stem length was 195 mm. Mean follow-up was 7 years. A competing risk model accounting for death was utilized.

The 15-year cumulative incidence of any revision was 24%. There were 48 revisions, with the most common reasons being dislocation (n=25) and PJI (n=16). The 15-year cumulative incidence of any reoperation was 28%. Only 13 revisions involved the fluted tapered portion of the component (FTC), for a 15-year cumulative incidence of any FTC revision of 8%. Only 2 FTCs were revised for aseptic loosening, resulting in a 15-year cumulative incidence of FTC revision for aseptic loosening of 1%. Stem subsidence >5 mm occurred in 2% of unrevised cases, and all stems were radiographically stable at most recent follow-up. Mean HHS was 77 at most recent follow-up.

This series demonstrated that MFT stems were durable and reliable even in the setting of two-stage reimplantation for infection. While the incidence of aseptic loosening was very low, the incidence of any revision was 24% at 15 years, primarily caused by dislocation and recurrent PJI.

Level of Evidence: IV


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 23 - 23
23 Jun 2023
Fehring TK Odum S Rosas S Buller LT Ihekweazu U Joseph H Gosthe RG Springer BD
Full Access

Periprosthetic joint infections (PJIs) centers are garnering the attention of different arthroplasty surgeons and practices alike. Nonetheless, their value has yet to be proven. Therefore, we evaluated weather PJI centers produce comparable outcomes to the national average of THA PJIs on a national cohort.

We performed a retrospective review of patient data available on PearlDiver from 2015 – 2021. PJI THA cases were identified through ICD-10 and CPT codes. Patients treated by 6 fellowship trained arthroplasty surgeons from a PJI center were matched based on age, gender, Charlson Comorbidity Index and Elixhauser comorbidity index at a 1:1 ratio to patients from the national cohort. Compared outcomes included LOS, ED visits, number of patients readmitted, total readmissions. Sample sized did not allow the evaluation of amputation, fusion or explantation. Normality was tested through the Kolmogorov-Smirnov test. And comparisons were made with Students t-tests and Chi Square testing.

A total of 33,001 THA PJIs and were identified. A total of 77 patients were identified as treated by the PJI center cohort and successfully matched. No differences were noted in regard to age, gender distribution, CCI or ECI (p=1, 1, 1 and 0.9958 respectively). Significant differences were noted in mean LOS (p<0.43), number of patients requiring readmissions (p=0.001) and total number of readmission events (p<0.001). No difference was noted on ED visits.

Our study demonstrates that a PJI for THA cases may be beneficial for the national growing trend of arthroplasty volume. Future data, that allows comparison of patient's specific data will allow for further validation of PJI centers and how these can play a role in helping the national PJI growing problem.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 28 - 28
23 Jun 2023
Massè A Giachino M Audisio A Donis A Secco D Turchetto L Limone B Via RG Aprato A
Full Access

Ganz's studies made it possible to address joint deformities on both femoral and acetabular side brought by the Legg-Calvè-Perthes disease (LCPD). Femoral head reduction osteotomy (FHRO) was developed to improve joint congruency along with periacetabular osteotomy (PAO). The purpose of this study is to show the clinical and morphologic outcomes of the technique, and an implemented planning approach.

From 2015 to 2023, 13 FHROs were performed on 11 patients for LCPD, in two centers. 11 of 13 hips had an associated PAO. A specific CT and MRI-based protocol for virtual simulation of the corrections was developed. Outcomes were assessed with radiographic parameters (sphericity index, extrusion index, integrity of the Shenton's line, LCE angle, Tonnis angle, CCD angle) and clinical parameters (ROM, VAS, Merle d'Aubigné-Postel score, modified-HHS, EQ5D-5L). Early and late complications were reported.

The mean follow-up was 40 months. The mean age at surgery was 11,4 years. No major complications were recorded. One patient required a total hip arthroplasty. Femoral Head Sphericity increased from 45% to 70% (p < 0,001); LCE angle from 18° to 42,8° (p < 0,001); extrusion Index from 36,6 to 8 (p < 0,001); Tonnis Angle from 14,4° to 6,2° (p = 0.1); CCD Angle from 131,7 to 136,5° (p < 0,023). The VAS score improved from 3,25 to 0,75,(p = 0.06); Merle d'Aubigné-Postel score from 14.75 to 16 (p = 0,1); Modified-HHS from 65,6 to 89,05 (p = 0,02). The EQ 5D 5L showed significant improvements. ROM increased especially in abduction and extra-rotation.

FHRO associated with periacetabular procedures is a safe technique that showed improved functional, clinical and morphologic outcomes in LCPD. The newly introduced simulation and planning algorithm may help to further refine the technique.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 33 - 33
23 Jun 2023
Engh CA Bhal V Hopper RH
Full Access

When the Birmingham Hip Resurfacing (BHR) metal-on-metal implant system was approved by the United States Food and Drug Administration in 2006, a multicenter, prospective, post-approval study (PAS) was required. This study uses data from the PAS to investigate metal level and glomerular filtration rate (GFR) trends over the first decade in vivo.

Between October 2006 and March 2011, 290 primary BHR procedures were performed among 262 patients at 5 sites. Whole blood samples were sent to a single specialized laboratory to determine GFR, cobalt (Co) and chromium (Cr) levels. The population for this study consists of 117 unrevised unilateral patients with a mean age at surgery of 51.3±6.5 years who had pre-operative, 1-year, 4-year, 5-year and 10-year laboratory data. The mean follow-up for these patients that included 36 females was 10.1±0.2 years.

Median metal levels at 1-year increased relative to pre-operative values for Co (by a factor of 9.7 from 0.13 to 1.26 ppb, p<0.001) and Cr (by a factor of 2.5 from 0. 60 to 1.50 ppb, p<0.001). Metal levels subsequently remained relatively constant over time with a median 10-year value of 1.12 ppb for Co and 1.29 ppb for Cr. Based on 585 blood samples from all 117 patients, there was no relationship between GFR and Co (=−0.06, p=0.14) or Cr (=0.05, p=0.27) levels. However, lower pre-operative GFR values were associated with larger increases in Co at 1-year relative to the pre-operative level (=−0.26, p=0.005). There was no relationship between pre-operative GFR values and changes in Cr at 1 year (=−0.13, p=0.15).

Through the first decade in vivo, elevated whole blood metal levels for unilateral BHR patients do not appear to adversely affect GFR. However, patients with lower pre-operative GFR values tend to have larger increases in their Co level at 1-year.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 34 - 34
23 Jun 2023
Bunyoz KI Tsikandylakis G Mortensen K Gromov K Mohaddes M Malchau H Troelsen A
Full Access

In metal-on-polyethylene (MoP) THA large femoral metal heads are designed to increase stability and to reduce dislocation risk. The increased head size could lead to increased taper corrosion with the release of metal ions and adverse reactions. Using blood ion measurements, we aimed to investigate the association between femoral head size and metal-ion release after MoP THA.

96 patients were enrolled at two centers and randomized to receive either a 32-mm metal head or a 36–44 mm metal head (the largest possible fitting the thinnest available polyethylene insert). Blood metal ions and PROMs (OHS, UCLA) were measured at two- and five-year follow-ups.

Both 2- and 5-year median chrome, cobalt, and titanium levels were below taper corrosion indicative ion levels. At 5 years, median chrome, cobalt, and titanium levels were 0.5 μg/L (0.50–0.62), 0.24 μg/L (0.18–0.30), and 1.16 μg/L (1.0–1.68) for the 32-mm group, and 0.5 μg/L (0.5–0.54), 0.23 μg/L (0.17–0.39), and 1.30 μg/L (1–2.05) for the 36–44 mm group, with no difference between groups (p=0.825, p=1.000, p=0.558). At 2 years, 7 (32-mm) versus 4 (36–44-mm) patients had elevated ions. At 5 years, 6 (32-mm) versus 7 (36–44-mm) patients had elevated ions. There was no difference in either OHS (p=0.665) or UCLA (p=0.831) between patients with or without elevated blood metal ions.

5 years after the insertion of MoP THAs, we found no differences in the blood metal ion levels between 32 mm heads and 36–44 mm heads and no corrosion-related revisions. As taper corrosion can debut after 5 years, there is still a need for long-term follow-up studies on the association between head size and corrosion in MoP THA.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 36 - 36
23 Jun 2023
Bizot P
Full Access

Genetic skeletal disorders constitute a rare and heterogeneous bone diseases often leading to poor quality of life. Several surgical options are available. The surgeon must deal with specific features (bone deformity, previous procedures, abnormal bone quality, stiffness or instability, muscle weakness). The questions concern the feasibility of the procedures and the surgical strategy.

55 patients (26 W, 29 M) were reviewed between 2016 and 2022. The mean age of the patients was 35 years (17–71). The diagnosis included 9 hereditary multiple exostoses, 8 osteogenesis imperfecta, 6 multiple epiphyseal dysplasia congenita, 6 achondroplasia, 4 osteopetrosis, 3 pycnodysostosis, 3 hypophosphatemic rickets, 3 fibrous dysplasia, 2 mucopolysaccharidosis, and 10 miscellaneous. 25 patients were referred for hip problems (40 hips). 4 patients (7 hips) requiring a THA have not been operated (4 planned). 4 patients (6 hips) had a proximal femoral fixation (2 osteotomies, 4 fracture fixations). 17 patients (27 hips) sustained a THA (25 primary, 2 revisions). All of them were operated by one operator, using a posterolateral approach and standard implants (including 7 dysplastic and 2 short stems). No customized implant has been used.

As regard the 27 THAs, the mean follow up was 4.2 years (1–12). The early complications included 2 femoral cracks and 1 femur fracture. There were 2 revisions (1 cup loosening at 2 years, 1 stem loosening at 4 years). No infection nor dislocation occurred. All the patients were satisfied with their treatment and regain some autonomy. 3 THA were considered as unfeasible.

Constitutional bone diseases need a multidisciplinary program of care. The indication for surgery is based on a mutual trust patient/surgeon, a careful evaluation of benefits/risks, and an accurate imaging to anticipate the difficulties. The expected results are a better function and quality of life, and a stability over time.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 7 - 7
23 Jun 2023
van Hellemondt GG Faraj S de Windt T van Hooff M Spruit M
Full Access

Pelvic discontinuity (PD) is a detrimental complication following total hip arthroplasty (THA). The aim of this study was to assess the clinical and radiological results of patients with PD who were revised using a custom-made triflange acetabular component (CTAC).

This is a single centre prospective study of patients with PD following THA who were treated with a CTAC. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Clinical and radiological complications, including reasons for revisions were registered. Trends over time are described and tested for significance and clinical relevance.

18 females with 22 CTACs were included with a mean age of 73.5 years (SD 7.7). There were significant improvements between baseline and final follow-up in HOOS (p<0.01), mOHS (p<0.01), EQ-5D-3L utility (p<0.01), EQ-5D-3L NRS (p<0.01), VAS pain rest (p<0.01), and VAS pain activity (p<0.01). A minimal clinically important improvement in mOHS and the HOOS was observed in 16 patients (73%) and 14 patients (64%), respectively. Definite healing of the PD was observed in 19 hips (86%). Complications included six cases with broken screws (27%), four cases (18%) with bony fractures, and one case (4.5%) with sciatic nerve paresthesia. One patient with concurrent bilateral PD had revision surgery due to recurrent dislocations. No revision surgery was performed for screw failure or implant breakage.

This is the first prospective assessment in clinical outcome of patients with PD who were treated with a CTAC. We have demonstrated that CTAC in patients with THA acetabular loosening and PD can result in stable constructs with no mechanical failures. Moreover, clinically relevant improvements in health-related quality of life at two years’ follow-up was observed.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 3 - 3
23 Jun 2023
Berdis GE Couch CG Larson DR Bedard NA Berry DJ Lewallen DG Abdel MP
Full Access

Cup-cage constructs are one of several methods commonly used to treat severe acetabular bone loss during contemporary revision total hip arthroplasty. The purpose of this study was to provide a long-term results of the technique with emphasis on implant survivorship, radiographic results, and clinical outcomes for both full and half cup-cage reconstructions.

We identified 57 patients treated with a cup-cage reconstruction for major acetabular bone loss between 2002–2012. All patients had Paprosky Type 2B through 3B bone loss, with 60% having an associated pelvic discontinuity. Thirty-one patients received a full cup-cage construct, and 26 a half cup-cage. Mean age at reconstruction was 66 years, 75% were female, and the mean BMI was 27 kg/m2. Mean follow-up was 10 years.

The 10-year cumulative incidences of any revision were 14% and 12% for the full and half cup-cage construct groups, respectively. Of the 9 revisions, 3 were for dislocation, 2 for aseptic loosening and construct failure (both were pelvic discontinuities), 1 for adverse local tissue reaction, and 1 for infection with persistent pelvic discontinuity. The 10-year cumulative incidences of revision for aseptic loosening were 4.5% and 5% for the full and half cup-cage constructs, respectively. Of the unrevised cases, incomplete and non-progressive zone 3 radiolucent lines were observed in 10% of patients in each group. Three patients experienced partial motor and sensory sciatic nerve palsies (2 in the full and 1 in the half cup-cage group). Both the full and half cup-cage cohorts demonstrated significantly improved Harris hip scores.

Full and half cup-cage reconstructions for major acetabular defects were successful at 10 years in regards to acetabular fixation without appreciable differences between the two techniques. However, zone 3 radiolucent lines were not uncommon in association with discontinuities, and dislocation continues to be a problem.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 8 - 8
23 Jun 2023
Baujard A Martinot P Demondion X Dartus J Girard J Migaud H
Full Access

Mechanical irritation or impingement of the iliopsoas tendon accounts for 2–6% of persistent postoperative pain cases after total hip arthroplasty (THA). The most common trigger is anterior cup overhang. CT-scan can be used to identify and measure this overhang; however, no threshold exists for symptomatic anterior iliopsoas impingement. We conducted a case–control study in which CT-scan was used to define a threshold that differentiates patients with iliopsoas impingement from asymptomatic patients after THA.

We analyzed the CT-scans of 622 patients (758 CT-scans) between 2011 and 2020. Out of this population we identified 136 patients with symptoms suggestive of iliopsoas impingement. Among them, 6 were subsequently excluded: three because the diagnosis was reestablished intra-operatively (one metallosis, two anterior instability related to posterior prosthetic impingement) and three because they had another obvious cause of impingement (one protruding screw, one protruding cement plug, one stem collar), leaving 130 patients in the study (impingement) group. They were matched to a control group of 138 patients who were asymptomatic after THA. The anterior cup overhang (anterior margin of cup not covered by anterior wall) was measured by an observer (without knowledge of the clinical status) on an axial CT slice based on anatomical landmarks (orthogonal to pelvic axis).

The impingement group had a median overhang of 8 mm [IQR: 5 to 11] versus 0 mm [IQR: 0 to 4] for the control group (p<.001). Using ROC curves, an overhang threshold of 4 mm was best correlated with a diagnosis of impingement (sensitivity 79%, specificity 85%, PPV = 75%, NPV = 85%).

Pain after THA related to iliopsoas impingement can be reasonably linked to acetabular overhang if it exceeds 4 mm on a CT scan. Below this threshold, it seems logical to look for another cause of iliopsoas irritation or another reason for the pain after THA before concluding impingement is present.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 58 - 58
23 Jun 2023
Fontalis A The CS Plastow R Mancino F Haddad FS
Full Access

In-hospital length of stay (LOS) and discharge disposition following arthroplasty could act as surrogate measures for improvement in patient pathways, and have major cost saving implications for healthcare providers. With the ever-growing adoption of robotic technology in arthroplasty, we wished to evaluate its impact on LOS. The objectives of this study were to compare LOS and discharge disposition following robotic-arm assisted (RO THA) versus conventional technique Total Hip Arthroplasty (CO THA).

This large-scale, single institution study included patients of any age undergoing primary THA (N = 1,732) for any cause between May 2019 and January 2023. Data extracted included patient demographics, LOS, need for Post Anaesthesia Care Unit (PACU) admission, anaesthesia type, readmission within 30 days and discharge dispositions. Univariate and multivariate logistic regression models were also employed to identify factors and patient characteristics related to delayed discharge.

The median LOS in the RO THA group was 54 hours (34, 78) versus 60 (51, 100) in the CO THA group, p<0.001. Discharge disposition was comparable between the two groups. In the multivariate model, age, need for PACU admission, ASA score > 2, female gender, general anaesthesia and utilisation of the conventional technique were significantly associated with LOS > 2 days.

Our study showed that robotic-arm assistance was associated with a shorter LOS in patients undergoing primary THA and no difference in discharge destination. Our results suggest that robotic-arm assistance could be advantageous in partly addressing the upsurge of hip arthroplasty procedures and the concomitant health care burden; however, this needs to be corroborated by long-term cost effectiveness analyses and data from randomised controlled studies.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 70 - 70
23 Jun 2023
Muratoglu OK Asik MD Nepple CM Wannomae KK Micheli BR Connolly RL Oral E
Full Access

Majority of ultra-high molecular weight polyethylene (UHMWPE) medical devices used in total joint arthroplasty are crosslinked using gamma radiation to improve wear resistance. Alternative methods of crosslinking are urgently needed to replace gamma radiation due to rapid decline in its supply. Peroxide crosslinking is a candidate method with widespread industrial applications. Oxidative stability and biocompatibility, which are critical requirements for medical device applications, can be achieved using vitamin-E as an additive and by removing peroxide by-products through high temperature melting, respectively. We investigated compression molded UHMWPE/vitamin-E/di-cumyl peroxide blends followed by high-temperature melting in inert gas as a material candidate for tibial knee inserts. Wear resistance increased and mechanical properties remained largely unchanged. Oxidation induction time was higher than most of the other clinically available formulations. The material passed the local-end point biocompatibility tests per ISO 10993. Compounds found in exhaustive extraction were of no concern with margin-of-safety values well above the accepted level, indicating a desirable toxicological risk profile. Peroxide crosslinked, vitamin-E stabilized, and high temperature melted UHMWPE has recently been cleared for clinical use in tibial knee inserts. With all the salient characteristics needed in a material that can provide superior long-term performance in total joint patients, peroxide crosslinking can replace gamma radiation crosslinking of UHMWPE for use in all total joint replacement implant including acetabular liners.