Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
Applied filters
The Bone & Joint Journal

Include Proceedings
Dates
Year From

Year To
The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1458 - 1462
1 Nov 2015
Langlois J Atlan F Scemama C Courpied JP Hamadouche M

Most published randomised controlled trials which compare the rates of wear of conventional and cross-linked (XL) polyethylene (PE) in total hip arthroplasty (THA) have described their use with a cementless acetabular component.

We conducted a prospective randomised study to assess the rates of penetration of two distinct types of PE in otherwise identical cemented all-PE acetabular components.

A total of 100 consecutive patients for THA were randomised to receive an acetabular component which had been either highly XL then remelted or moderately XL then annealed.

After a minimum of eight years follow-up, 38 hips in the XL group and 30 hips in the annealed group had complete data (mean follow-up of 9.1 years (7.6 to 10.7) and 8.7 years (7.2 to 10.2), respectively). In the XL group, the steady state rate of penetration from one year onwards was -0.0002 mm/year (sd 0.108): in the annealed group it was 0.1382 mm/year (sd 0.129) (Mann–Whitney U test, p < 0.001). No complication specific to either material was recorded.

These results show that the yearly linear rate of femoral head penetration can be significantly reduced by using a highly XLPE cemented acetabular component.

Cite this article: Bone Joint J 2015;97-B:1458–62.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 56 - 63
1 Jan 2015
Abane L Anract P Boisgard S Descamps S Courpied JP Hamadouche M

In this study we randomised 140 patients who were due to undergo primary total knee arthroplasty (TKA) to have the procedure performed using either patient-specific cutting guides (PSCG) or conventional instrumentation (CI).

The primary outcome measure was the mechanical axis, as measured at three months on a standing long-leg radiograph by the hip–knee–ankle (HKA) angle. This was undertaken by an independent observer who was blinded to the instrumentation. Secondary outcome measures were component positioning, operating time, Knee Society and Oxford knee scores, blood loss and length of hospital stay.

A total of 126 patients (67 in the CI group and 59 in the PSCG group) had complete clinical and radiological data. There were 88 females and 52 males with a mean age of 69.3 years (47 to 84) and a mean BMI of 28.6 kg/m2 (20.2 to 40.8). The mean HKA angle was 178.9° (172.5 to 183.4) in the CI group and 178.2° (172.4 to 183.4) in the PSCG group (p = 0.34). Outliers were identified in 22 of 67 knees (32.8%) in the CI group and 19 of 59 knees (32.2%) in the PSCG group (p = 0.99). There was no significant difference in the clinical results (p = 0.95 and 0.59, respectively). Operating time, blood loss and length of hospital stay were not significantly reduced (p = 0.09, 0.58 and 0.50, respectively) when using PSCG.

The use of PSCG in primary TKA did not reduce the proportion of outliers as measured by post-operative coronal alignment.

Cite this article: Bone Joint J 2015;97-B:56–63.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 342 - 348
1 Mar 2010
El Masri F Kerboull L Kerboull M Courpied JP Hamadouche M

We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component.

At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1439 - 1445
1 Nov 2007
Triclot P Grosjean G El Masri F Courpied JP Hamadouche M

We carried out a prospective randomised study designed to compare the penetration rate of acetabular polyethylene inserts of identical design but different levels of cross-linking at a minimum of four years follow-up. A total of 102 patients (102 hips) were randomised to receive either highly cross-linked Durasul, or contemporary Sulene polyethylene inserts at total hip replacement. A single blinded observer used the Martell system to assess penetration of the femoral head. At a mean follow-up of 4.9 years (4.2 to 6.1) the mean femoral head penetration rate was 0.025 mm/year (sd 0.128) in the Durasul group compared with 0.106 mm/year (sd 0.109) in the Sulene group (Mann-Whitney test, p = 0.0027). The mean volumetric penetration rate was 29.24 mm3/year (sd 44.08) in the Durasul group compared with 53.32 mm3/year (sd 48.68) in the Sulene group. The yearly volumetric penetration rate was 55% lower in the Durasul group (Mann-Whitney test, p = 0.0058). Longer term results are needed to investigate whether less osteolysis will occur.