Hips with metal-on-metal total hip arthroplasty (MoM THA) have a high rate of adverse local tissue reactions (ALTR), often associated with hypersensitivity reactions. Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) measures tissue perfusion with the parameter Ktrans (volume transfer constant of contrast agent). Our purpose was 1) to evaluate the feasibility of DCE-MRI in patients with THA and 2) to compare DCE-MRI in patients with MoM bearings with metal-on-polyethylene (MoP) bearings, hypothesising that the perfusion index Ktrans in hips with MoM THA is higher than in hips with MoP THA. In this pilot study, 16 patients with primary THA were recruited (eight MoM, eight MoP). DCE-MRI of the hip was performed at 1.5 Tesla (T). For each patient, Ktrans was computed voxel-by-voxel in all tissue lateral to the bladder. The mean Ktrans for all voxels was then calculated. These values were compared with respect to implant type and gender, and further correlated with clinical parameters.Objectives
Methods
The Oxford Hip and Knee Scores (OHS, OKS) have been demonstrated
to vary according to age and gender, making it difficult to compare
results in cohorts with different demographics. The aim of this
paper was to calculate reference values for different patient groups
and highlight the concept of normative reference data to contextualise an
individual’s outcome. We accessed prospectively collected OHS and OKS data for patients
undergoing lower limb joint arthroplasty at a single orthopaedic
teaching hospital during a five-year period.
T-scores were calculated based on the OHS and OKS distributions. Objectives
Methods