Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene
This is a multicentre, non-inventor, prospective observational study of 503 INFINITY fixed bearing total ankle arthroplasties (TAAs). We report our early experience, complications, and radiological and functional outcomes. Patients were recruited from 11 specialist centres between June 2016 and November 2019. Demographic, radiological, and functional outcome data (Ankle Osteoarthritis Scale, Manchester Oxford Questionnaire, and EuroQol five-dimension five-level score) were collected preoperatively, at six months, one year, and two years. The Canadian Orthopaedic Foot and Ankle Society (COFAS) grading system was used to stratify deformity. Early and late complications and reoperations were recorded as adverse events. Radiographs were assessed for lucencies, cysts, and/or subsidence.Aims
Methods
The interest in unicompartmental knee arthroplasty (UKA) for
medial osteoarthritis has increased rapidly but the long-term follow-up
of the Oxford UKAs has yet to be analysed in non-designer centres.
We have examined our ten- to 15-year clinical and radiological follow-up
data for the Oxford Phase III UKAs. Between January 1999 and January 2005 a total of 138 consecutive
Oxford Phase III arthroplasties were performed by a single surgeon
in 129 patients for medial compartment osteoarthritis (71 right
and 67 left knees, mean age 72.0 years (47 to 91), mean body mass
index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were
prospectively recorded and obtained at intervals. Of the 129 patients,
32 patients (32 knees) died, ten patients (12 knees) were not able
to take part in the final clinical and radiological assessment due
to physical and mental conditions, but via telephone interview it
was confirmed that none of these ten patients (12 knees) had a revision
of the knee arthroplasty. One patient (two knees) was lost to follow-up.Aims
Patients and Methods
Since redesign of the Oxford phase III mobile-bearing unicompartmental
knee arthroplasty (UKA) femoral component to a twin-peg design,
there has not been a direct comparison to total knee arthroplasty
(TKA). Thus, we explored differences between the two cohorts. A total of 168 patients (201 knees) underwent medial UKA with
the Oxford Partial Knee Twin-Peg. These patients were compared with
a randomly selected group of 177 patients (189 knees) with primary
Vanguard TKA. Patient demographics, Knee Society (KS) scores and
range of movement (ROM) were compared between the two cohorts. Additionally,
revision, re-operation and manipulation under anaesthesia rates
were analysed.Aims
Patients and Methods
There is a large amount of evidence available
about the relative merits of unicompartmental and total knee arthroplasty
(UKA and TKA). Based on the same evidence, different people draw
different conclusions and as a result, there is great variability
in the usage of UKA. The revision rate of UKA is much higher than TKA and so some
surgeons conclude that UKA should not be performed. Other surgeons
believe that the main reason for the high revision rate is that
UKA is easy to revise and, therefore, the threshold for revision
is low. They also believe that UKA has many advantages over TKA
such as a faster recovery, lower morbidity and mortality and better
function. They therefore conclude that UKA should be undertaken
whenever appropriate. The solution to this argument is to minimise the revision rate
of UKA, thereby addressing the main disadvantage of UKA. The evidence
suggests that this will be achieved if surgeons use UKA for at least
20% of their knee arthroplasties and use implants that are appropriate
for these broad indications. Cite this article:
The cementless Oxford unicompartmental knee replacement
has been demonstrated to have superior fixation on radiographs and
a similar early complication rate compared with the cemented version.
However, a small number of cases have come to our attention where,
after an apparently successful procedure, the tibial component subsides into
a valgus position with an increased posterior slope, before becoming
well-fixed. We present the clinical and radiological findings of
these six patients and describe their natural history and the likely
causes. Two underwent revision in the early post-operative period,
and in four the implant stabilised and became well-fixed radiologically with
a good functional outcome. This situation appears to be avoidable by minor modifications
to the operative technique, and it appears that it can be treated
conservatively in most patients. Cite this article: