Advertisement for orthosearch.org.uk
Results 1 - 5 of 5
Results per page:

Aims

This study aimed to compare the change in health-related quality of life of patients receiving a traditional cemented monoblock Thompson hemiarthroplasty compared with a modern cemented modular polished-taper stemmed hemiarthroplasty for displaced intracapsular hip fractures.

Patients and Methods

This was a pragmatic, multicentre, multisurgeon, two-arm, parallel group, randomized standard-of-care controlled trial. It was embedded within the WHiTE Comprehensive Cohort Study. The sample size was 964 patients. The setting was five National Health Service Trauma Hospitals in England. A total of 964 patients over 60 years of age who required hemiarthroplasty of the hip between February 2015 and March 2016 were included. A standardized measure of health outcome, the EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and at four months following the operation.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 310 - 316
1 Mar 2017
Hothi H Henckel J Shearing P Holme T Cerquiglini A Laura AD Atrey A Skinner J Hart A

Aims. The aim of this study was to compare the design of the generic OptiStem XTR femoral stem with the established Exeter femoral stem. Materials and Methods. We obtained five boxed, as manufactured, implants of both designs at random (ten in total). Two examiners were blinded to the implant design and independently measured the mass, volume, trunnion surface topography, trunnion roughness, trunnion cone angle, Caput-Collum-Diaphyseal (CCD) angle, femoral offset, stem length, neck length, and the width and roughness of the polished stem shaft using peer-reviewed methods. We then compared the stems using these parameters. Results. We found that the OptiStems were lighter (p < 0.001), had a rougher trunnion surface (p <  0.001) with a greater spacing and depth of the machined threads (p < 0.001), had greater trunnion cone angles (p = 0.007), and a smaller radius at the top of the trunnion (p = 0.007). There was no difference in stem volume (p = 0.643), CCD angle (p = 0.788), offset (p = 0.993), neck length (p = 0.344), stem length (p = 0.808), shaft width (p = 0.058 to 0.720) or roughness of the polished surface (p = 0.536). Conclusion. This preliminary investigation found that whilst there were similarities between the two designs, the generic OptiStem is different to the branded Exeter design. Cite this article: Bone Joint J 2017;99-B:310–16


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims

Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study.

Methods

In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 661 - 663
1 Jun 2020
Meek RMD Treacy R Manktelow A Timperley JA Haddad FS

In this review, we discuss the evidence for patients returning to sport after hip arthroplasty. This includes the choices regarding level of sporting activity and revision or complications, the type of implant, fixation and techniques of implantation, and how these choices relate to health economics. It is apparent that despite its success over six decades, hip arthroplasty has now evolved to accommodate and support ever-increasing patient demands and may therefore face new challenges.

Cite this article: Bone Joint J 2020;102-B(6):661–663.


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 941 - 950
1 Aug 2019
Scott CEH MacDonald DJ Howie CR

Aims

The EuroQol five-dimension (EQ-5D) questionnaire is a widely used multiattribute general health questionnaire where an EQ-5D < 0 defines a state ‘worse than death’ (WTD). The aim of this study was to determine the proportion of patients awaiting total hip arthroplasty (THA) or total knee arthroplasty (TKA) in a health state WTD and to identify associations with this state. Secondary aims were to examine the effect of WTD status on one-year outcomes.

Patients and Methods

A cross-sectional analysis of 2073 patients undergoing 2073 THAs (mean age 67.4 years (sd 11.6; 14 to 95); mean body mass index (BMI) 28.5 kg/m2 (sd 5.7; 15 to 72); 1253 female (60%)) and 2168 patients undergoing 2168 TKAs (mean age 69.3 years (sd 9.6; 22 to 91); BMI 30.8 kg/m2 (sd 5.8; 13 to 57); 1244 female (57%)) were recorded. Univariate analysis was used to identify variables associated with an EQ-5D score < 0: age, BMI, sex, deprivation quintile, comorbidities, and joint-specific function measured using the Oxford Hip Score (OHS) or Oxford Knee Score (OKS). Multivariate logistic regression was performed. EQ-5D and OHS/OKS were repeated one year following surgery in 1555 THAs and 1700 TKAs.