Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1060 - 1066
1 Sep 2022
Jin X Gallego Luxan B Hanly M Pratt NL Harris I de Steiger R Graves SE Jorm L

Aims. The aim of this study was to estimate the 90-day periprosthetic joint infection (PJI) rates following total knee arthroplasty (TKA) and total hip arthroplasty (THA) for osteoarthritis (OA). Methods. This was a data linkage study using the New South Wales (NSW) Admitted Patient Data Collection (APDC) and the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), which collect data from all public and private hospitals in NSW, Australia. Patients who underwent a TKA or THA for OA between 1 January 2002 and 31 December 2017 were included. The main outcome measures were 90-day incidence rates of hospital readmission for: revision arthroplasty for PJI as recorded in the AOANJRR; conservative definition of PJI, defined by T84.5, the PJI diagnosis code in the APDC; and extended definition of PJI, defined by the presence of either T84.5, or combinations of diagnosis and procedure code groups derived from recursive binary partitioning in the APDC. Results. The mean 90-day revision rate for infection was 0.1% (0.1% to 0.2%) for TKA and 0.3% (0.1% to 0.5%) for THA. The mean 90-day PJI rates defined by T84.5 were 1.3% (1.1% to 1.7%) for TKA and 1.1% (0.8% to 1.3%) for THA. The mean 90-day PJI rates using the extended definition were 1.9% (1.5% to 2.2%) and 1.5% (1.3% to 1.7%) following TKA and THA, respectively. Conclusion. When reporting the revision arthroplasty for infection, the AOANJRR substantially underestimates the rate of PJI at 90 days. Using combinations of infection codes and PJI-related surgical procedure codes in linked hospital administrative databases could be an alternative way to monitor PJI rates. Cite this article: Bone Joint J 2022;104-B(9):1060–1066


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 90 - 101
1 Jan 2020
Davis ET Pagkalos J Kopjar B

Aims

The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA).

Methods

Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad).


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1158 - 1166
14 Sep 2020
Kaptein BL den Hollander P Thomassen B Fiocco M Nelissen RGHH

Aims

The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs).

Methods

A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.