Advertisement for orthosearch.org.uk
Results 1 - 5 of 5
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 55 - 55
11 Apr 2023
Raina D Markeviciute V Arvidsson L Törnquist E Stravinskas M Kok J Jacobson I Liu Y Tengattini A Sezgin E Vater C Zwingenberger S Isaksson H Tägil M Tarasevicius S Lidgren L
Full Access

Majority of osteoporosis related fractures are treated surgically using metallic fixation devices. Anchorage of fixation devices is sometimes challenging due to poor osteoporotic bone quality that can lead to failure of the fracture fixation.

Using a rat osteoporosis model, we employed neutron tomography and histology to study the biological effects of implant augmentation using an isothermally setting calcium sulphate/hydroxyapatite (CaS/HA) biomaterial with synthetic HA particles as recruiting moiety for systemically administered bisphosphonates. Using an osteoporotic sawbones model, we then provide a standardized method for the delivery of the CaS/HA biomaterial at the bone-implant interface for improved mechanical anchorage of a lag-screw commonly used for hip fracture fixation. As a proof-of-concept, the method was then verified in donated femoral heads and in patients with osteoporosis undergoing hip fracture fixation.

We show that placing HA particles around a stainless-steel screw in-vivo, systemically administered bisphosphonates could be targeted towards the implant, yielding significantly higher peri-implant bone formation compared to un-augmented controls. In the sawbones model, CaS/HA based lag-screw augmentation led to significant increase (up to 4 times) in peak extraction force with CaS/HA performing at par with PMMA. Micro-CT imaging of the CaS/HA augmented lag-screws in cadaver femoral heads verified that the entire length of the lag-screw threads and the surrounding bone was covered with the CaS/HA material. X-ray images from fracture fixation surgery indicated that the CaS/HA material could be applied at the lag-screw-bone interface without exerting any additional pressure or risk of venous vascular leakage.: We present a new method for augmentation of lag-screws in fragile bone. It is envisaged that this methodcould potentially reduce the risk of fracture fixation failure especially when HA seeking “bone active” drugs are used systemically.


Bone & Joint Research
Vol. 8, Issue 2 | Pages 49 - 54
1 Feb 2019
Stravinskas M Nilsson M Vitkauskiene A Tarasevicius S Lidgren L

Objectives

The aim of this study was to analyze drain fluid, blood, and urine simultaneously to follow the long-term release of vancomycin from a biphasic ceramic carrier in major hip surgery. Our hypothesis was that there would be high local vancomycin concentrations during the first week with safe low systemic trough levels and a complete antibiotic release during the first month.

Methods

Nine patients (six female, three male; mean age 75.3 years (sd 12.3; 44 to 84)) with trochanteric hip fractures had internal fixations. An injectable ceramic bone substitute, with hydroxyapatite in a calcium sulphate matrix, containing 66 mg of vancomycin per millilitre, was inserted to augment the fixation. The vancomycin elution was followed by simultaneously collecting drain fluid, blood, and urine.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_17 | Pages 44 - 44
1 Dec 2018
Stravinskas M Tarasevicius S Vitkauskiene A Nilsson M Lidgren L
Full Access

Aim

In vivo studies have shown a preventive and curative effect of using an injectable vancomycin containing biphasic ceramic in an osteomyelitis model. No clinical long term pharmacokinetic release study has been reported. Inadequate concentration in target tissues results in treatment failure and selection pressure for antibiotic-resistant organisms.

Our hypothesis was that vancomycin in the first week would reach high local concentrations but with low systemic levels.

Method

9 patients (6 women, 3 men) with trochanteric hip fractures classified as A1 and A2 according to the AO-classification all had internal fixations. The mean age was 75.3 years (± S.D. 12.3 years, range 44–84y). An injectable ceramic with hydroxyapatite embedded in a calcium sulphate matrix containing 66mg vancomycin per mL augmented the fixation. A mean of 9.7 mL (± S.D. 0.7 mL, range 8–10mL) was used. The elution of vancomycin was followed by collecting drain fluid, blood (4 days) and urine (4 weeks)


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_23 | Pages 17 - 17
1 Dec 2016
Stravinskas M Horstmann PF Hettwer W Nilson M Tarasevicius S Petersen MM Lidgren L
Full Access

Aim

The demand for a synthetic bone substitute that can build bone and at the same time kill bacteria is high. The aim of this study was to compare the elution of gentamicin from a new synthetic bone substitute in vitro with the performance in clinical applications.

Method

Gentamicin release was measured from a synthetic bone graft substitute, comparing in vitro and clinical conditions:

elution in Ringers solution. The bone graft substitute contained 175mg gentamicin per 10mL. The material was introduced either as paste or as pre-set beads with a high or low surface areas, >100cm2 and 24cm2 respectively. The gentamycin release was measured by daily collection of samples.

elution in patients treated for trochanteric hip fractures(n=6) or uncemented hip revisions(n=5) 7,3±1,1mL of substitute was implanted and drainage was collected at 6h,12h,24h,30h,36h post-op. Blood serum was collected every hour for the first 6h and thereafter every 6h until 4 days post-op, urine – daily for the first 7 days post-op.

elution in patients treated after bone tumor resection(n=8), 12,1±5,5mL of substitute was implanted and both drainage and blood serum were collected daily until 2 days post-op.

Gentamicin concentrations were analyzed using antibody technique.


Bone & Joint Research
Vol. 5, Issue 9 | Pages 427 - 435
1 Sep 2016
Stravinskas M Horstmann P Ferguson J Hettwer W Nilsson M Tarasevicius S Petersen MM McNally MA Lidgren L

Objectives

Deep bone and joint infections (DBJI) are directly intertwined with health, demographic change towards an elderly population, and wellbeing.

The elderly human population is more prone to acquire infections, and the consequences such as pain, reduced quality of life, morbidity, absence from work and premature retirement due to disability place significant burdens on already strained healthcare systems and societal budgets.

DBJIs are less responsive to systemic antibiotics because of poor vascular perfusion in necrotic bone, large bone defects and persistent biofilm-based infection. Emerging bacterial resistance poses a major threat and new innovative treatment modalities are urgently needed to curb its current trajectory.

Materials and Methods

We present a new biphasic ceramic bone substitute consisting of hydroxyapatite and calcium sulphate for local antibiotic delivery in combination with bone regeneration. Gentamicin release was measured in four setups: 1) in vitro elution in Ringer’s solution; 2) local elution in patients treated for trochanteric hip fractures or uncemented hip revisions; 3) local elution in patients treated with a bone tumour resection; and 4) local elution in patients treated surgically for chronic corticomedullary osteomyelitis.