Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 52 - 52
1 Mar 2017
Navruzov T Riviere C Van Der Straeten C Harris S Aframian A Iranpour F Cobb J Auvinet E
Full Access

Background

The accurate positioning of the total knee arthroplasty affects the survival of the implants(1). Alignment of the femoral component in relation to the native knee is best determined using pre- and post-operative 3D-CT reconstruction(2). Currently, the scans are visualised on separate displays. There is a high inter- and intra-observer variability in measurements of implant rotation and translation(3). Correct alignment is required to allow a direct comparison of the pre- and post-operative surfaces. This is prevented by the presence of the prostheses, the bone shape alteration around the implant, associated metal artefacts, and possibly a segmentation noise.

Aim

Create a novel method to automatically register pre- and post-operative femora for the direct comparison of the implant and the native bone.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 53 - 53
1 Mar 2017
Navruzov T Van Der Straeten C Riviere C Jones G Cobb J Auvinet E
Full Access

Introduction

Hip resurfacing arthroplasty (HRA) is currently regaining positive attention as a treatment of osteoarthritis in young, active individuals[1]. The procedure is complex and has low tolerance for implant malpositioning [2]. ‘Precision tools', such as imageless navigation and patient specific instruments, have been developed to assist with implant positioning but have not been shown to be fully reliable [3]. The aim of this study is to present and validate the first step of novel quality control tool to verify implant position intra-operatively. We propose that, before reaming of the femoral head, a handheld structured light 3D scanner can be used to assess the orientation and insertion point of femoral guide wire.

Methods

Guide wires were placed into the heads of 29 solid foam synthetic femora. A specially designed marker (two orthogonal parallelepipeds attached to a shaft) was inserted into the guide wire holes. Each bone (head, neck and marker) was 3D scanned twice (fig 1). The insertion point and guide wire neck angle were calculated from the marker's parameters. Reference data was acquired with an optical tracking system. The measurements calculated with the 3D scans were compared to the reference ones to evaluate the precision. The comparison of the test retest measurements done with the new method are used to evaluate intra-rater variability.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 100 - 100
1 Jan 2017
Navruzov T Rivière C Van Der Straeten C Harris S Cobb J Auvinet E Aframian A Iranpour F
Full Access

The accurate positioning of the total knee arthroplasty affects the survival of the implants(1). Alignment of the femoral component in relation to the native knee is best determined using pre- and post-operative 3D-CT reconstruction(2). Currently, the scans are visualised on separate displays. There is a high inter- and intra-observer variability in measurements of implant rotation and translation(3). Correct alignment is required to allow a direct comparison of the pre- and post-operative surfaces. This is prevented by the presence of the prostheses, the bone shape alteration around the implant, associated metal artefacts, and possibly a segmentation noise.

The aim is to create a novel method to automatically register pre- and post-operative femora for the direct comparison of the implant and the native bone.

The concept is to use post-operative femoral shaft segments free of metal noise and of surgical alteration for alignment with the pre-operative scan. It involves three steps. Firstly, using principal component analysis, the femoral shafts are re-oriented to match the X axis. Secondly, variants of the post-operative scan are created by subtracting 1mm increments from the distal femoral end. Thirdly, an iterative closest point algorithm is applied to align the variants with the pre-operative scan.

For exploratory validation, this algorithm was applied to a mesh representing the distal half of a 3D scanned femur. The mesh of a prosthesis was blended with the femur to create a post-operative model. To simulate a realistic environment, segmentation and metal artefact noise were added. For segmentation noise, each femoral vertex was translated randomly within +−1mm,+−2mm,+−3mm along its normal vector. To create metal artefact random noise was added within 50 mm of the implant points in the planes orthogonal to the shaft. The alignment error was considered as the average distance between corresponding points which are identical in pre- and post-operative femora.

These preliminary results obtained within a simulated environment show that by using only the native parts of the femur, the algorithm was able to automatically register the pre- and post-operative scans even in presence of the implant. Its application will allow visualisation of the scans on the same display for the direct comparison of the perioperative scans.

This method requires further validation with more realistic noise models and with patient data. Future studies will have to determine if correct alignment has any effect on inter- and intra-observer variability.