Patients treated with limb salvage surgery for bone sarcomas of the extremities (upper and lower) may have physical disability as a result of treatment. Goal of this study was to evaluate the quality of life after treatment (chemotherapy and conservative surgery) at long term. 208 patients resected for a bone sarcoma and with prosthetic reconstruction (45 in the upper and 163 in the lower limb) were evaluated. Assessment of results was done using the Karnofsky Scale (K.S.). Patients were followed in the clinic and functional results assessed according to the Musculoskeletal Tumour Society (MSTS) system. Moreover the Toronto Extremity Salvage Score questionnaire (TESS) was mailed to 144 patients.Aim
Method
Objective of this study was to analyse results of two stage revisions in infected megaprostheses in lower limb.
Infection occurred in 80 cases (7.7%) at mean time of 4 yrs (min 1 month, max 19 yrs) in 18 KMFTR®, 47 HMRS®, 5 HMRS® Rotating Hinge, 10 GMRS®. Sites: 51 distal femurs, 21 proximal tibias, 6 proximal femurs, 1 total femur and 1 extrarticular knee resection. Most frequent bacteria causing infection were: Staphilococcus Epidermidis (39 cases), Staphilococcus Aureus (17) and Pseudomonas Aeruginosa (5). Infection occurred postoperatively within 4 weeks in 9 cases, early (within 6 months) in 12 cases, late (after 6 months) in 59 cases. Usual surgical treatment was “two stage” (removal of implant, one or more cement spacers with antibiotics, new implant), with antibiotics according with coltures. One stage treatment was used for immediate postoperative infections, only since 1998. Functional results after treatment of infection were assessed using the MSTS system.
In 3 cases an amputation was primarily performed, to proceed with chemotherapy. Revisions for infection were successful in 63 pts (79%), while 17 pts were amputated (21%). Functional results evaluated in 53 revised cases were good or excellent in 43 (81.1%).
Patients included 44 females and 38 males, ranging in age from 14 to 74 years. Mean follow-up was 9.5 years (min. 3, max. 27). Histopathological findings included chordomas in 55 cases and giant cell tumor (GCT) in 27. Most pts. had iv antibiotic therapy with amikacin and teicoplanin. Surgery of chordoma was resection, surgery of GCT was intralesional excision. In 6 sacral resections a miocutaneous transabdominal flap of rectus abdominis was used for posterior closure.
Mean surgical time was 14 hours for resections and 6 hours for excisions. No significant difference was found comparing deep wound infections with levels of resection (15/33 resections proximal to S3-45% and 8/19 resections below or at S3-42%), previous intralesional surgery elsewhere (4/9 patients previous treated elsewhere-44% and 19/46 primarily treated patients-41%) and age at surgery.
S. Epidermidis and S. Aureo were the most frequent bacteria causing infection (45%). Two stage treatment of infection was chosen: removal of the implant and temporary substitution with cement spacer with antibiotics (usually vancomycin) until infection healed. But a new prostheses was actually implanted in 3 cases only (at mean time of 5.7 mos), while in 17 the spacer was never removed by patients choice due to the acceptable result with the spacer. Systemic antibiotics were associated according to cultural results. Infection healed in all patients.
Purpose of this study was to analyse the different techniques of prosthetic reconstruction of the humerus (also in association with bone grafts) after resection of primary tumors, discussing indications and evaluating implant survival.
Histologically 24 were benign tumors and 253 primary malignant tumors. All patients were periodically followed in the clinic, imaging studies and histology were reviewed and special attention given to prostheses-related complications and implant survival. Univariate analysis through actuarial Kaplan Meier curves was used in evaluating implant survival to major complications. Functional results were assessed using the MSTS system.
Major complications of the implants included 19 cases of deep infection (6.8%), 8 aseptic loosenings (2.9%), 4 breakages (1.4%) causing failure of the implants requiring revisions. Further complications were observed in revised cases. Actuarial curve of implant survival to major complications showed over 80% at 10 years and over 70% at 20 years. Functional results according to the MSTS system were good or excellent (over 50%) in more than 90% of the patients, with an average score of 79%.
Purpose of this study is to analyze the results of a modular reconstructive tumor prosthesis for the lower limb (GMRS®) with a comparative statistical analysis of primary and secondary implants.
Gorham-Stout disease (GSD) is rare, characterized by proliferation of vascular channels resulting in progressive distruction of bone. In the Rizzoli files we found 15 cases of GSD from 1968 to 2008. Two were excluded for insufficient documentation. For 13 cases clinical data, imaging and histology were analysed. Histopathologic features included benign vascular proliferation, vascular pattern of osteolytic angioma, fibro-connective tissue component and bony destruction. A final diagnosis was established based on clinical, radiological and histopathologic features. Imaging included X-rays in 11 cases and CT or MRI in 5. All lesions were lytic, with associated sclerosis in two cases. There was one lesion only in 4 cases, multiple lesions in the same bone in 1 and multiple bones involved in 6. Primary sites were proximal femur in 7 cases, pelvis in 2, hip and knee, calcaneus, humerus and cervical spine in 1 case each. Two patients had no treatment, 2 conservative treatment (cast or brace), 5 surgery, 6 medical treatment (byphosphonates, calcitonin, zoledronic acid, interferon, steroids), 1 radiotherapy, 2 selective arterial embolization. Surgery consisted of internal fixation of pathologic fractures in 4 patients and reconstruction of the entire humerus with a double composite allograft in 1. Treatment was surgery only in 2 patients, medical treatment in 4 (1 also embolization), surgery and medical treatment in 2 (1 also embolization), radiotherapy only in 1, conservative treatment in 2. Four patients were lost at follow up. Mean follow up was 17 ys.(min 2, max 30) in 9 patients: 2 dead, 3 healed, 3 with stable disease, 1 alive with disease at 24 ys. No conclusive treatment recommendations are possible; surgery is indicated in pathologic fractures or reconstruction of massively destroyed bones, medical treatment and selective embolization are helpful. In literature prosthetic reconstruction is preferred due to the risk of allografts resorption.
Osteosarcoma is a common primary bone sarcoma and distal femur its most frequent site. Between 2003 and 2008 at Rizzoli, 66 patients with osteosarcoma of the distal femur had neoadjuvant chemotherapy, resection and reconstruction with modular uncemented mega-prostheses. Series included 37 males and 29 females. Mean follow up was 2 years. To measure “subjective” outcome Karnofsky scale (KPS) was assessed for each patient pre and post-treatment. Also a functional evaluation according to the MSTS system was performed. To find out the current quality of life, a questionnaire on life at work, study and sport before and after treatment was sent to 64 alive patients. Before treatment 7 patients had a Karnofsky index (KI) of 60%, 31 of 50%, 25 of 40% and 3 of 30%. After treatment 19 patients had a Karnofsky index performance of 90%, 28 of 80%, 11 of 70%, 5 of 50% and 1 of 40%. Two patients died of disease. The most represented index of KPS after teatment was “Able to carry on normal activity; minor symptoms”. Poor results were related with amputation (2), knee stiffness (3), infection (2), aseptic loosening (1). After treatment 91% of patients had a KI over 70%, while 89% a KI lower than 50% pre-treatment. MSTS system showed excellent or good results in 85% and fair or poor in 15% of the patients. Average score at MSTS evaluation was 22 (73%). Questionnaires (some still pending) confirm previous analysis. KPS is simple and effective in evaluating quality of life in patients treated for distal femur osteosarcoma. In this study it confirmed the satisfactory MSTS assessed results. It is an easy method, useful and accessible for patients. The reported analysis shows that patients treated for osteosarcoma of the distal femur can have a good quality of life.
Between 1983 and 2006 at Rizzoli 669 knee modular uncemented prostheses were implanted after resection of the distal femur, total femur or proximal tibia. These prostheses include 126 KMFTR prostheses and 543 second generation HMRS prostheses. Patients were followed periodically in the clinic. Data was obtained from clinical charts and imaging studies with special attention to major complications requiring revision surgery. Revision for polyethylene wear was considered a minor complication, since it did not imply failure of the implant. Functional results were assessed according to the MSTS system. Since data could be misleading due to deaths in an oncologic population (although 2/3 of patients were cured or long survivors), to censore the implant unrelated events Kaplan-Meyer curves of implant survival were studied. In 126 KMFTR group infection rated 13.5%, stem breakage 13%, aseptic loosening 9.5%; change of polyethylene rated 44%. In 543 HMRS prostheses infection rated 8.6%, stem breakage 3%, aseptic loosening 4.8%; revision for polyethylene wear rated 9.6%. Techniques of revisions and their outcome analysed showed about 2/3 of good results, but increased risk of further complications in revised implants. Functional results (MSTS system) were good or excellent in 80% of KMFTR prostheses and in 90% of HMRS. Decrease of major complications in newer prosthetic design was statistically significant and possibly due to newer materials and modified stem design. Polyethylene wear also decreased significantly. Function was satisfactory in most patients without complications for both groups. Revision surgery is technically demanding and appropriate timing of revision is crucial, since early treatment can improve final outcome.
Purpose of this paper was to review the Rizzoli experience in prosthetic reconstruction of the knee after resection of bone tumors with special attention to major complications and functional outcome.
Since abrupt data could be misleading due to the oncologic population and related deaths (although 2/3 of the patients were cured or long survivors), to censore the implant unrelated events a statistical analysis of the implant survival was performed and Kaplan-Meyer curves of implant survival were studied.
In 543 HMRS prostheses major complications were infection 8%, stem breakage 2%, aseptic loosening 4%, while revision for polyethylene wear components rated 7.4%. Techniques of revisions were analyzed, as well as the outcome of revised cases, which showed that about 2/3 of the patients treated for major complications do well, although the risk for further complications is significantly incresed in revised implants.
Discussion: The reduction of major complication rate with the newer designs of the modular prosthesis was statistically significant and this seems to confirm that newer materials and a modified stem design positively affected the implant survival. Also the wear of polyethylene component was dramatically decreased in the newer prosthetic design. Functional results were satisfactory in most of the patients that did not experience major complications, with a trend to improvement in the newer design. Revision surgery is technically demanding and sometimes newer ‘hybrid’ techniques are suggested, implying the use of bone allografts. Appropriate timing of revisions is crucial. An early treatment of complications can improve the final outcome.