MRI experiments were undertaken at 25.0 ± 0.1 °C on a Bruker Avance NMR spectrometer (Bruker Bio-Spin, Rheinstetten, Germany) using a 7.0T vertical bore magnet system, equipped with a 1.1 T m-1 (110 G cm-1) gradient set and 15 mm ‘birdcage’ RF resonator. Specimens for testing were immersed in physiological saline inside a 15 mm NMR. Both multi echo and diffusion weighted images were acquired with a recycle time TR = 2 s and 8 averages using a 0.7 mm slice thickness, a field of view (FOV) of ca. 15 × 15 mm and a 128 × 128 matrix. For multi echo experiments the echo time was 5 ms with 64 echoes and for DT imaging a diffusion gradient duration δ = 2 ms and diffusion delay Δ = 12 ms. The diffusion tensor was calculated from the seven requisite diffusion-weighted images using in-house Matlab® code (The Mathworks, Natick, MA) written for the purpose.