Twenty-nine patients (30 hips) with uncemented acetabular impaction allografting contained behind a metal backed component screwed to the pelvis at revision hip arthroplasty were reviewed at 12 to17 years (average 15.3 years) follow up. Five patients had died with the prosthesis in situ, 4 patients were lost to follow up, 13 patients had failed requiring further revision (only one failed prior to 5 years) and 9 survivors were minimally asymptomatic. The mean time to failure warranting further surgery was 9 years. Analysis of available serial radiographs (24 cases) demonstrated signs of loosening (migration, progressive radiolucent lines, screw breakage) in 54% of the latest radiographs of all cases. Removing those lost to follow up or deceased, 72% were radiologically loose (in the intact asymptomatic group 57% could be defined as loose). Additionally, in 70% of the cases the acetabular component tended to fail in a varus manner as the medial wall remodelled. These results indicate as previously published, short term results for this technique are satisfactory but in the long term they are not. Factors associated with this include the pressfit nature of the polyethylene liner which has been implicated in disappointing long term results for this prosthesis in primary applications, but not of the level of failure encountered in the current series. This experience suggests that the reliance on screw fixation over a bed of allograft in the absence of cement supplementation does not provide sufficient stability for reliable bone graft incorporation. The method reported above should be abandoned.