Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties.Objectives
Patients and Methods
Birmingham Metal-on-metal total hip resurfacing (BHR) is a bone-conserving option for patients with advanced articular damage. While the outcomes of Total Hip Replacement (THR) are well documented, there is a paucity of literature comparing the patient reported outcomes of BHR versus THR. This study aims to compare the patient reported outcomes for an impact on quality of life between patients who had a BHR vs. THR after correcting for selection bias. Patients who underwent a BHR or THR between July 2003 and December 2006 were included. Patient questionnaires included demographic details, co-morbidities, WOMAC, SF-36 Scores. In addition, a 4 point Likert scale was used to measure satisfaction post-operatively. The above data was collected pre-operatively and at 1, 2, 3 and 5 years post-operatively. Data was analysed with SPSS (version 19) software package. All analysis was adjusted for Age, gender, Co-morbidity and pre-operative score by using Multivariate regression analysis using a General Liner Model to rule out the effect of these predictors on outcomes between groups. 337 patients were included (205 for THR and 132 for BHR). BHR patients were younger than THR patients (49 vs. 67 years, p<0.01), were more likely to be male (68% vs. 42% of THR, p<0.01), reported fewer co-morbidities (1.06 vs. 1.59, p<0.01). BHR patients reported better WOMAC pain and function scores at 5 years (p<0.05). For SF-36 scores, BHR patients reported higher scores for all 8 domains at 1 year and 5 year follow up (p<0.05). BHR patients reported higher satisfaction than THR group for return to Activities of Daily Living and Recreational activities at 1 and 5 years (p<0.05) After correction for patient variability, BHR patients reported better improvement in pain and function and enjoyed a better quality of life in relation to return to ADL and recreational activities over to 5 years post-surgery.
Wear debris released from bearing surfaces has been shown to
provoke negative immune responses in the recipient. Excessive wear
has been linked to early failure of prostheses. Analysis using coordinate
measuring machines (CMMs) can provide estimates of total volumetric
material loss of explanted prostheses and can help to understand
device failure. The accuracy of volumetric testing has been debated,
with some investigators stating that only protocols involving hundreds
of thousands of measurement points are sufficient. We looked to
examine this assumption and to apply the findings to the clinical
arena. We examined the effects on the calculated material loss from
a ceramic femoral head when different CMM scanning parameters were
used. Calculated wear volumes were compared with gold standard gravimetric
tests in a blinded study. Objectives
Methods