Introduction: Knotless Suture Anchors provide numerous advantages in arthroscopic rotator-cuff (RC) repair such as, reducing the difficulties of knot tying, reducing surgical exposure, thus decreasing morbidity. The purpose of this in-vitro study was to compare the pull-out strength of three new knotless suture anchors in correlation with bone quality using the following anchors: Opus Magnum 2 -ArthroCare Co., USA; Push Lock and Swivel Lock -Arthrex Inc., USA.
Material &
Methods: Ten healthy and ten osteopenic macroscopically intact humeri with an average age of 51.7 and 79.5 years, respectively, were loaded with the three knotless suture anchors according to the manufacturers’ description. The healthy humeri had a mean trabecular BMD of 152.77 mgCa-HA/ml. The osteopenic humeri had a mean trabecular bone mineral density of 54.02 mgCa-HA/ml. The humeri were positioned in a custom - engineered adjustable fixation device, stabilising the direction of the pull of the sutures at an angle of 135° to the axis of the humeral shaft (Universal testing device Z010/TN2A – Zwick GmbH, Ulm, Germany). The anchors were cyclically loaded to simulate postoperative conditions.
The ultimate pull-out strenngth, the initial displacement in millimeters after the first pull with 75 N and the modes of failure were recorded.
Results: The mean ultimate failure loads of the Opus Magnum 2, PushLock and SwiveLock anchors in osteopenic humeri were 135.0, 102.5 and 130.0 N (p>
0.05), respectively, and in healthy humeri 142.5 N, 182.5 N and 202.5 N (p>
0.05), respectively. The initial system displacement in osteopenic humeri of the Opus Magnum 2, PushLock and SwiveLock anchors were 3.53 mm, 16.11 mm and 3.23 mm (p<
0.01), respectively, and in healthy humeri 3.71 mm, 1.98 mm, and 1.96 mm (p>
0.05), respectively.
Discussion: The results of this study show that in osteopenic humeri, the Opus Magnum 2 and SwiveLock anchors display significant superiority in system displacement with an initial pull of 75N compared to the PushLock anchor in osteopenic bone, but all three anchors fail to provide significance in the ultimate failure load. PushLock anchor might cause a greater gap formation between the RC-tendon and greater tubercle interface in osteopenic humeri due to inferior gripping and therefore should not be used solely for RC repair. Due to a manufacturing flaw the suture holding fixture of the Opus Magnum 2 anchor “breaks” when a mean force of 138.75 N is applied to the system, regardless of the bone quality, thus disabling the anchor to unfold it’s properties in healthy bone. Whereas the results in osteopenic bone are comparable to the other two anchors. SwiveLock provides the best support of all three anchors in healthy humeri.