Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 267 - 268
1 Jul 2008
ADAM P PHILIPPOT R COUMERT S FARIZON F FESSY M
Full Access

Purpose of the study: The double-mobility concept was introduced for clinical applications for total hip arthroplasty in 1976. The concept preserves joint range of motion while increasing stability. In this study we evaluated the consequences of these advantages in terms of polyethylene wear, measuring wear both on the concave and convex surfaces and volumetrically.

Material and methods: Forty polyethylene inserts were explanted and analyzed. Explantation had been performed for mechanical or septic failure after eight years implantation on average. Mean age of patients at implantation was 46 years. After examining the gross aspect of the insert, surface analysis was performed with direct measurement of changes in the curvature using a BHN 706 position sensor for the inner concave surface and lateral projection for the outer convex surface. Estimated measurement error was ±5μm for each method; the manufacturer's tolerance for production of the inserts was 50μm. Volumetric wear was determined by reference to the manufacturer’s data. Student’s t test for paired series was applied.

Results: At gross inspection, all inserts had lost the strips originally present on the convex surface; 40% presented visible wear of the retaining ring. Mean annual wear (± standard deviation, SD) obtained with the measuring system was 9±9 μm/yr) for the convex surface and 73 ± 69 μm/yr for the concave surface. Total annual wear, the sum of inner and outer surface wear, was 82±72 μm/yr. The mean volumetric wear was 28±28 mm3/year for the convex surface and 25±23 mm3/year for the concave surface and 53.4±40 mm3/year for total wear.

Discussion: Total wear for these 40 double-mobility inserts which had functioned in vivo was not greater than the values reported for the metal-polyethylene bearing with 22.2 mm femoral heads. The double mobility is not associated with greater wear. While there was no significant difference between the wear volume of the convex versus the concave surfaces, the differentials wear were widespread, which can be considered to result from functional differences.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 268 - 268
1 Jul 2008
ADAM P PHILIPPOT R DARGAI F COUMERT S FARIZON F FESSY M
Full Access

Purpose of the study: Double mobility prostheses are increasingly popular. Evidence in the literature demonstrates greater efficacy for the treatment and prevention of prosthesis instability. Ten-year survival is to the order of 95% (Aubriot, Philippot). One of the drawbacks is the risk of prosthetic head displacement outside the retaining polyethylene ring, i.e. intraprosthetic dislocation. We searched for factors causing this complication.

Material and methods: We reviewed retrospectively 67 files concerning intraprosthetic dislocation among a series of Novae cups (Serf) implanted from 1982. Head diameter was 22.2 mm for 59 cases, with a Pro stem (Serf) for 31 cases and a PF stem (Serf) for 36. Each type of stem has a specific neck design. All patients underwent revision surgery; the retaining function of the explanted pieces was analyzed.

Results: Mean time to the complication was 91 months; mean patient age at implantation was 54 years. Early cases exhibited macroscopically intact retaining capacity. Intermediary and late cases exhibited macroscopic wear with an oval shaped retaining ring. For three cases, intraprosthetic dislocation followed an episode of dislocation reduced under sedation. The cups measured 53 mm on average. The rate of calcification was high in this population (15 cases of Brooker grade 3 or 4). Mean survival was significantly different between the Pro and PF stems.

Discussion: Early dislocations were related to insufficient retaining capacity of the initially inserted ring. After a corrective measure by the manufacturer, this type of early complication has disappeared. Late dislocations resulted from impingement wear. Dislocation of a prosthesis with a double-mobility cup increases the risk of intraprosthetic dislocation after reduction; reduction procedures should thus be performed under general anesthesia with curare treatment. We analyzed the different parameters involved: head-neck relation, activity, periprosthetic calcification, cup diameter, resurfacing of the prosthetic neck. Observations were compared with data in the literature.