header advert
Results 1 - 13 of 13
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 111 - 111
1 Mar 2017
Reynolds R Walker P Buza J Borukhov I
Full Access

INTRODUCTION

Understanding the biomechanics of the anatomical knee is vital to innovations in implant design and surgical procedures. The anterior – posterior (AP) laxity is of particular importance in terms of functional outcomes. Most of the data on stability has been obtained on the unloaded knee, which does not relate to functional knee behavior. However, some studies have shown that AP laxity decreases under compression (1) (2). This implies that while the ligaments are the primary stabilizers under low loads, other mechanisms come into play in the loaded knee. It is hypothesized this decreased laxity with compressive loads is due to the following: the meniscus, which will restrain the femur in all directions; the cartilage, which will require energy as the femur displaces across the tibial surface in a plowing fashion; and the upwards slope of the anterior medial tibial plateau, which stabilizes the knee by a gravity mechanism. It is also hypothesized that the ACL will be the primary restraint for anterior tibial translation.

METHODS

A test rig was designed where shear and compressive forces could be applied and the AP and vertical displacements measured (Figure 1). The AP motion was controlled by the air bearings and motor, allowing for the accurate application of the shear force. Position and force data were measured using load cells, potentiometers, and a linear variable differential transducer.

Five knee specimens less than 60 years old and without osteoarthritis (OA), were evaluated at compressive loads of 0, 250, 500, 750 N, with the knee at 15° flexion. Three cycles of shear force at ±100 N constituted a test. The intact knee was tested, followed by testing after each of the following resections: LCL, MCL, PCL, ACL, medial meniscus, and lateral meniscus.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 75 - 75
1 Mar 2017
Walker P Borukhov I Bosco J Reynolds R
Full Access

INTRODUCTION

Most total knees today are CR or PS, with lateral and medial condyles similar in shape. There is excellent durability, but a shortfall in functional outcomes compared with normals, evidenced by abnormal contact points and gait kinematics, and paradoxical sliding. However unicondylar, medial pivot, or bicruciate retaining, are preferred by patients, ascribed to AP stability or retention of anatomic structures (Pritchett; Zuiderbaan). Recently, Guided Motion knees have been shown to more closely reproduce anatomic kinematics (Walker; Willing; Amiri; Lin; Zumbrunn). As a design approach we proposed Design Criteria: reproduce the function of each anatomic stabilizing structure with bearing surfaces on the lateral and medial sides and intercondylar; resected cruciates because this is surgically preferred; avoid a cam-post because of central femur bone removal, soft tissue entrapment, noises, and damage (Pritchett; Nunley). Our hypothesis was that these criteria could produce a Guided Motion design with normal kinematics.

METHODS & MATERIALS

Numerous studies on stability and laxity showed the ACL was essential to controlling posterior femoral displacement on the tibia whether the knee was loaded or unloaded. Under load, the anterior upwards slope of the medial tibial plateau prevented anterior displacement (Griffen; Freeman; Pinskerova; Reynolds). The posterior cruciate and the downward lateral tibial slope produced lateral rollback in flexion. The Replica Guided Motion knee had 3 bearings (Fig 1). The lateral side was shallow and sloped posteriorly, with a posterior lip to prevent excess displacement. The medial anterior tibial and femoral slopes were increased as in the anatomic knee. In the intercondylar region, a saddle bearing replaced ACL function by controlling posterior femoral displacement. For testing, a typical PS design was used as comparison. A Knee Test Machine (Fig 2) flexed the knee, and applied axial compression, shear and torque to represent a range of functions. Bone shapes were reproduced by 3D printing and collaterals by elastomeric bands. Motion was recorded with a digital camera, and Geomagic to process data.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 118 - 118
1 May 2016
Walker P Arno S Borukhov I Bell C Salvadore G
Full Access

Introduction

The major function of the medial meniscus has been shown to be distribution of the load with reduction of cartilage stresses, while its role in AP stability has been found to be secondary. However several recent studies have shown that cartilage loss in OA occurs in the central region of the tibia while the meniscus is displaced medially. In a lab study (Arno, Hadley 2013) it was confirmed that the AP laxity was greatly reduced with a compressive force across the knee, while the femur shifted posteriorly and the AP laxity was increased after a partial meniscetomy of the posterior horn. It is therefore possible that under load, the compression of the meniscus and the cartilage, 2–3mm in total, allows load transmission on the central tibial plateau, and causes radial expansion and tension of the meniscus providing restraint to femoral displacements. This leads to our hypotheses that the highest loading on the medial meniscus would be at the extremes of motion, rather than in the mid-range, and that the meniscus would provide the majority of the restraint to anterior-posterior femoral displacements throughout flexion when compressive loads were acting.

Methods & Materials

MRI scans were taken of ten knee specimens to verify the absence of pathology and produce computer models. The knees were loaded in combinations of compressive and shear loading over a full flexion range. Tekscan sensors were used to measure the pressure distribution across the joint as the knee was flexed continuously. A digital camera was used to track the motion, from which femoral-tibial contacts were determined by computer modelling. Load transmission was determined from the Tekscan for the anterior horn, central body, posterior horn, and the uncovered cartilage in the center of the meniscus. An analysis was carried out (Fig 2) to determine the net anterior or posterior shear force carried by the meniscus.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 120 - 120
1 May 2016
Walker P Meere P Borukhov I Bell C
Full Access

PURPOSE

Soft tissue balancing can be achieved by using spacer blocks, by distractors which measure tensile forces, or by instrumented devices which measure the forces on the lateral and medial condyles. However there is no quantitative method for assessment of balancing at clinical follow-up; to address this, we developed a Smart Knee Fixture (SKF) which measured the varus and valgus angles for a moment of 10 Nm. Our purpose was to determine if varus and valgus angles measured at clinical follow-up, was equivalent to the balancing parameters of distraction forces or contact forces measured at surgery.

METHODS

The SKF, which measured VV angles using stretch sensors on each side of the knee, was validated by cadaver studies, fluoroscopy, and emg. The balancing parameters were:

The lateral and medial contact forces at surgery, expressed as FL/FM

The distraction tensions in the collateral ligaments at surgery, expressed as TL/TM

The moments to cause lift-off when a varus or valgus moment is applied, MVAR/MVAL

The varus and valgus angles measured at post-op follow-up, VAR/VAL

A force analysis, and measurements on 101 surgical cases & clinical follow-up in an IRB study, were carried out to determine the relationship between these parameters.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 36 - 36
1 May 2016
Meere P Walker P Schneider S Salvadore G Borukhov I
Full Access

Introduction

The role of soft tissue balancing in optimizing functional outcome and patient satisfaction after total knee arthroplasty surgery is gaining interest. This is due in part to the inability of pure alignment to demonstrate excellent functional outcomes 6. Consistent soft tissue balancing has been aided by novel technologies that can quantify loads across the joint at the time of surgery 4. In theory, compressive load equilibrium should be correlated with ligamentous equilibrium between the medial and lateral collateral ligaments. The authors propose to use the Collateral Ligaments Strain Ratio (CLSR) as a functional tool to quantify and track surgical changes in laxity of the collateral ligaments and correlate this ratio to validated functional scores and patient reported outcomes. The relationship with intra-operative balancing of compartmental loads can then be scrutinized. The benefits of varus-valgus balancing within 2o include increased range of motion 7, whereas pressure imbalance between the medial and lateral joint compartments has been linked to condylar liftoff and abnormal kinematics post-TKA 8.

Methods

The study is a prospective IRB approved clinical study with three cohorts of 50 patients each: (1) a surgical prospective study group (2) a matched control group of non-operated high function patients; (3) a matched control group of high function knee arthroplasty recipients. Standard statistical analysis method is applied. The testing of the CLSR is performed using a validated Smart Knee Brace developed by the authors and previously reported 1. The output variables consist of the maximum angular change of the knee in the coronal plane at 10 degrees of flexion produced by a controlled torque application in the varus and valgus (VV) directions. This creates measureable strain on the lateral and medial collateral ligaments, which is reported as a ratio (CLSR). The New Knee Society Score is used to track outcomes. The intra-operative balance is achieved by means of an instrumented tibial tray (OrthoSensor, Inc).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 37 - 37
1 May 2016
Meere P Schneider S Borukhov I Walker P
Full Access

Introduction

The mechanical classical method of knee surgical instrumentation by alignment is based on built-in compromises and is considered insufficient to ensure consistent success. Soft tissue balancing is thus now seen as necessary for optimal functional outcomes and patient satisfaction. (Matsuda 2005, Winemaker 2002). The authors have previously demonstrated that balancing can be achieved through specific strategic moves. In this study, the goal was to determine the efficacy of a given surgical algorithm and to define predictors of improved outcome. The surgical target is equilibrium of contact loads. The mechanical axis remains in neutral, however subtle variation in the joint line obliquity and posterior slope are tolerated within the literature established boundaries of +/− 3 degrees and less than 10 degrees respectively.

Methods

Data was obtained from 101 consecutive primary procedures from a single surgeon (PAM) using a PCL-retaining device. For all cases the testing methodology consisted of a sag test, heel push, drawer testing at 90 degrees, and varus-valgus laxity testing at 10 degrees of flexion. Instrumented tibial trials were used to measure the contact forces on the lateral and medial sides at 10, 30, 60 and 90 degrees of flexion. Specific releases were identified and noted based on matrix profiling after each test. Re-iteration loops were enacted until balance within 15 lbs. of difference was achieved. The data was expressed as the ratio of medial/total force (total=medial + lateral), with 0.5 being equal lateral and medial forces. This was named the Contact Load Ratio (CLR). The load distribution was expressed as a scatter graph of lateral v. medial compartmental loads (Figure 1)


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 116 - 116
1 May 2016
Walker P Meere P Borukhov I Bell C
Full Access

PURPOSE

Soft tissue balancing can be achieved by using spacer blocks, by distractors which measure tensile forces, or by instrumented devices which measure the forces on the lateral and medial condyles. However there is no quantitative method for assessment of balancing at clinical follow-up; to address this, we developed a Smart Knee Fixture (SKF) which measured the varus and valgus angles for a moment of 10 Nm. Our purpose was to determine if varus and valgus angles measured at clinical follow-up, was equivalent to the balancing parameters of distraction forces or contact forces measured at surgery. METHODS: The SKF, which measured VV angles using stretch sensors on each side of the knee, was validated by cadaver studies, fluoroscopy, and emg. The balancing parameters were:

The lateral and medial contact forces at surgery, expressed as FL/FM

The distraction tensions in the collateral ligaments at surgery, expressed as TL/TM

The moments to cause lift-off when a varus or valgus moment is applied, MVAR/MVAL

The varus and valgus angles measured at post-op follow-up, VAR/VAL

A force analysis, and measurements on 101 surgical cases & clinical follow-up in an IRB study, were carried out to determine the relationship between these parameters.

RESULTS

The ratio TL/TM was approx. equal to FL/FM, especially near to a balanced state

The ratio MVAR/MVAL (lift-off moments) was equal to FL/FM

The ratio VAR/VAL was approx. equal to FL/FM only if the collateral stiffnesses were equal;

otherwise the ratio was approx. proportional to the collateral stiffnesses.

In the clinical follow-ups, there was no significant linear relation between VAR/VAL and FL/FM.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_5 | Pages 6 - 6
1 Feb 2016
Meere P Schneider S Borukhov I Walker P
Full Access

Introduction

The role of soft tissue balancing in optimising functional outcome and patient satisfaction after total knee arthroplasty surgery is gaining interest. Consistent soft tissue balancing has been aided by novel technologies that can quantify loads across the joint at the time of surgery. Based on free body diagram theory, compressive load equilibrium should be correlated with ligamentous equilibrium between the medial and lateral collateral ligaments. The authors propose to use the Collateral Ligaments Strain Ratio (CLSR) as a functional tool to quantify and track the effectuated surgical change in laxity of the medial and lateral collateral ligaments and correlate this ratio to validated functional scores and patient reported outcomes. The relationship with intra-operative balancing of compartmental loads can then be scrutinised.

Methods

The study is a prospective clinical study with three cohorts of 50 patients each: (1) a surgical prospective study group with ligamentous testing pre-operatively, at 4 weeks, 3 months and 6 months post-operatively; (2) a matched control group of non-operated high function patients; (3) a matched control group of high function knee arthroplasty recipients. Standard statistical analysis method is applied. The testing of the CLSR is performed using a validated Smart Knee Brace developed by the authors and previously reported. The output variables consist of the maximum angular change of the knee in the coronal plane at 10 degrees of flexion produced by a controlled torque application in the varus and valgus (VV) directions. This creates measureable strain on the lateral and medial collateral ligaments, which is reported as a ratio (CLSR). The New Knee Society Score is used to track outcomes. The intra-operative balance is achieved by means of an instrumented tibial tray (OrthoSensor, Inc). The applied torque was standardised to 10Nm with a handheld wireless dynamometer.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_5 | Pages 3 - 3
1 Feb 2016
Meere P Schneider S Borukhov I Walker P
Full Access

Introduction

Balancing at surgery is important for clinical outcome in terms of pain relief, flexion range, and function. The methodology usually involves making bone cuts to achieve correct leg alignment, and then obtaining equal gaps in extension and flexion using spacer blocks or tensor devices. In this study, we describe a method for quantifying balancing throughout the flexion range and show the effect of different surgical corrections from an unbalanced to a balanced state. In this way, we quantified how accurately balancing could be achieved within the practical time frame of a surgical procedure.

Methods

Data was obtained from 80 primary procedures using a PCL-retaining device. Initial bone cuts were made using navigation. Instrumented tibial trials were used to measure the contact forces and locations on the lateral and medial sides. Video/audio recordings were made of all aspects of the surgeries. The initial balancing was recorded during the Heel Push Test, namely the lateral and medial contact forces for the flexion range. The data was expressed as medial/total force ratio (total=medial + lateral), with 0.5 being equal lateral and medial forces. Surgical corrections to correct the specific imbalance pattern, determined from previous research, were carried out. The Heel Push Test was repeated after each correction and at final balancing.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 35 - 35
1 Jan 2016
Bell C Meere P Borukhov I Walker P
Full Access

Introduction

Evaluation of post-operative soft tissue balancing outcomes after Total Knee Arthroplasty (TKA) and other procedures can be measured by stability tests, with Anterior-Posterior (AP) drawer tests and Varus-Valgus (VV) ligamentous laxity tests being particularly important. AP stability can be quantified using a KT1000 device; however there is no standard way of measuring VV stability. The VV test relies on subjective force application and perception of laxity. Therefore we sought to develop and validate a device and method for quantifying knee balancing by analyzing VV stability.

Materials and Methods

Our team developed a Smart Knee Fixture to measure VV angular changes using two dielectric elastomer stretch sensors, placed strategically over the medial and lateral collateral ligaments (see Figure 1). The brace is secured in position with the leg in full extension and the sensors locked with pre-tension. Therefore, contraction and elongation of either sensor is measured and the VV angular deviation of the long axis of the femur relative to that of the tibia is derived and displayed in real time using custom software. EMG muscle activity was previously investigated to confirm there is no resistive activity during the VV test obstructing ligamentous evaluations.

The device was validated in two ways:

A bilateral lower body cadaver specimen, secured in a custom test rig, was used to compare the Smart Knee Fixture's readings to those measured from an optical surgical navigation system. Abduction and adduction force was gradually applied as varus and valgus moments with a wireless hand-held dynamometer up to 50N (19.8Nm) at 0 and 15° flexion.

Two male volunteers were used to compare the Smart Knee Fixture's readings to those measured from fluoroscopic images. An arthroscopic distal thigh leg immobilizer was used to prevent rotation and lateral movements of the thigh when moments were applied at the malleoli. A C-arm Fluoroscope was then positioned focusing on the center of the joint. The tests were performed at full extension, 10 and 20° of flexion and force was gradually applied to 50N.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 36 - 36
1 Jan 2016
Bell C Meere P Borukhov I Rathod P Walker P
Full Access

The use of smart trial components is now allowing a better assessment of soft tissue balancing at the time of total knee replacement surgery. A balanced knee can be defined as one that possesses symmetry, ie. equal and centered lateral and medial forces through the full range of flexion. There is still a need for a standard reproducible surgical test to quickly confirm optimized balancing at surgery with such devices. The Heel Push test is the established standard, by pushing the foot in a cephalad direction while supporting the thigh and keeping the leg stable in the vertical plane. A common variation of this test is the Thigh Pull test where the foot is actively assisted during the cephalad pull of the thigh through deep flexion. The test is an open chain test. The Thigh Pull test may be an improvement since the weight of the leg is alleviated and no supplemental compressive forces are introduced. The directional changes of the lower extremity are thus a result of ligamentous tension and balances. The purpose of this study is to compare the two tests using a standard testing methodology and observe the variation in kinetic parameters in a controlled biomechanical setting.

A custom l rig was developed, which independently controls all six degrees of freedom about the knee joint. In addition a commercial navigation system was used to derive instantaneous alignment values and flexion angles between the tibia and femur. The pelvis was fixed to the table and the foot was fitted onto a low friction carriage along a slide rail. The knee design used was cruciate retaining. The pressure mapping system was a wireless tibial trial that provided magnitude of load per compartment.

The study is a preliminary cadaveric study reporting the data from two. In this experiment the leg was then tested with the Heel Push and Thigh Pull tests after obtaining optimum soft tissue balance of the cadaveric specimen. From this standard neutral state a series of single surgical variables were introduced to mimic common intra-operative surgical corrections. This was achieved through custom tibial liner and angle shims.

The results defied theoretical anticipation. Though the total contact forces with heel push were generally higher than with thigh pull, the relative load distribution between compartments did not follow a trend (see Figure 1). Furthermore in deeper flexion the persistence of relatively high contact pressures would suggest that ligaments still generate intra-articular forces despite the much weaker gravitational effect. The clinical relevance lies in the asymmetry of the load distribution between medial and lateral compartment for the two methods tested. The load asymmetry as tested by the Thigh Pull test may correspond to an open chain in swing phase. This asymmetry would force some axial rotation and tibial femoral alignment deviation that can significantly affect the forces at the time of heel strike. The Heel Push test would be more representative of the compressive forces in a closed chain mode as seen during the stance phase of gait.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 34 - 34
1 Jan 2016
Bell C Meere P Borukhov I Rathod P Walker P
Full Access

Soft tissue balancing in total knee replacement may well be the determining factor in raising the fair patient satisfaction. The development of intelligent implants allows quantification of reactive loads to applied pressures. This can be tested in dynamic mode such as heel push test at surgery, or in static mode such as when testing for varus/valgus (VV) laxity of the collateral ligaments of the knee. We postulate that a well-balanced knee will have comparable if not equal load distribution across compartments in dynamic loading. When tested for laxity, we anticipate an equal or comparable response to VV applied loads under physiologic load range of 10–50N. This study sought to analyze the relationship between the kinematic (joint motion) and kinetic (force) effects to VV testing in the 0–15 degrees range of flexion. One goal was to demonstrate that testing the knee in locked extension (Screw Home effect) is unreliable and should be abandoned in favor of the more reliable VV testing at 10–15 degrees of flexion.

This is a preliminary cadaveric study utilizing data from two hemibodies. The pelvis was fixed in a custom test rig with open or closed chain lower leg testing capability along a sliding rail with foot VV translational. Forces were applied at the malleoli with a wireless hand held dynamometer. Kinematic analysis of the hip-knee-ankle (HKA) tibiofemoral angle was derived from a commercial navigation system with mounted infrared trackers. Kinetic analysis was derived from a commercially available sensor imbedded in a tibial trial liner. Balance was optimized by conventional methods with the use of the sensor feedback until loads were roughly symmetrical and VV testing yielded symmetrical rise in opposite compartments. The VV testing was then performed with the knees locked at the femoral side in axial rotation and translational motion in any plane. Sagittal flexion was pre-set at 0, 10, and 15 degrees and progressive load was applied.

Results

From the graphs one can observe significant differences between VV testing at 0 degrees (locked Screw Home), 10 degrees, and 15 degrees of flexion. The shaded area corresponds to the common range of VV stress testing loading pressure, typically less than 35N. The HKA deviates from neutrality no sooner than by the middle of the physiologic test zone. By 35N, the magnitude of the effect is also much less than that observed at 10 and 15 degrees (unlocked from Screw Home). From the kinetic analysis one can also note the significant difference in the High-Low spread throughout the testing range of applied pressure.

If the surgeon tests in the low range of applied loads, he/she may not observe the kinematic joint opening effect. The kinetic effect seems more reliable as sensed loads are detectable earlier on. It is clear however that testing at 10–15 degrees offers a much better sensitivity to the VV laxity or stiffness as exemplified in the bottom portions of the figure. Therefore testing in locked Screw Home full extension may lead to underestimation of the true coronal laxity of the joint.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 110 - 110
1 Jan 2016
Walker P Lowry M Arno S Borukhov I Bell C
Full Access

Total Knee Arthroplasty (TKA), has now become a reliable, successful, and widely used treatment for osteoarthritis. Numerous reports indicate that for the majority of patients, the TKA lasts a lifetime with pain relief and the ability to perform most everyday activities. However there are a number of ways in which the procedure can be further improved, the focus here being on function. One of the problems in evaluating function is that it depends upon the inherent ability, motivation, and expectation of the patients. There are several well-used questionnaire systems which capture functional ability objectively. In the effort to simplify evaluation, a ‘forgotten knee’ evaluation has been introduced, the concept being that ‘the ideal TKA design’ would feel and function like a normal knee. Such a measure would include factors such as surgical technique, alignment, and rehabilitation, as well as the TKA design itself. Another approach to evaluation is to measure biomechanical parameters, such as in gait analysis and fluoroscopy, which evaluate kinematic or kinematic parameters, using normal controls for comparison. Nevertheless, such evaluations still include factors other than the TKA design itself, and do not apply to new designs.

The approach taken here for the evaluation of a new TKA design independent of other factors, is to measure the neutral path of motion and the laxity boundaries of the loaded knee on the application of shear and torque over a full range of flexion. The benchmark is the same kinematic data from the normal intact knee. The rationale has some analogy to the ‘forgotten knee’ in that if the laxity response of a design of TKA is the same as that of the anatomic knee itself, the behavior of that implanted knee in any functional condition will be indistinguishable from that of the anatomic knee itself. Such a testing concept has some similarities to the constraint test described in the ASTM standard. In this paper, a novel design algorithm is proposed for creating different design concepts. First, a general morphological form is formulated for each design concept, a Cam-Post PS, a Saddle-Ramp, and a Converging Condyle, all with overall anatomic-like surfaces. Each femoral component is then designed, which is then moved through the normal neutral path and laxity paths, which creates the tibial surface. The concepts are evaluated using a Desktop Knee Machine configured to move the knee dynamically through full flexion while applying combinations of compression, shear and torque; kinematic data being captured optically and plotted using custom software. The normal benchmark was obtained from 10 normal knee specimens, which showed the restraint of the medial femoral condyle to anterior displacement and the overall rollback and laxity laterally. Compared with standard CR and PS designs, the Guided Motion designs were seen to more closely resemble normal. It is proposed that this approach can result in designs which will more likely reproduce a ‘forgotten knee’ and achieve the optimal function for a given patient.