Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_19 | Pages 8 - 8
1 Nov 2017
Annan J Murray A
Full Access

Simulation in surgical training has become a key component of surgical training curricula, mandated by the GMC, however commercial tools are often expensive. As training budgets become increasingly pressurised, low-cost innovative simulation tools become desirable. We present the results of a low-cost, high-fidelity simulator developed in-house for teaching fluoroscopic guidewire insertion.

A guidewire is placed in a 3d-printed plastic bone using simulated fluoroscopy. Custom software enables two inexpensive web cameras and an infra-red led marker to function as an accurate computer navigation system. This enables high quality simulated fluoroscopic images to be generated from the original CT scan from which the bone model is derived and measured guidewire position. Data including time taken, number of simulated radiographs required and final measurements such as tip apex distance (TAD) are collected.

The simulator was validated using a DHS model and integrated assessment tool. TAD improved from 16.8mm to 6.6mm (p=0.001, n=9) in inexperienced trainees, and time taken from 4:25s to 2m59s (p=0.011). A control group of experienced surgeons showed no improvement but better starting points in TAD, time taken and number of radiographs.

We have also simulated cannulated hip screws, femoral nail entry point and SUFE, but the system has potential for simulating any procedure requiring fluoroscopic guidewire placement e.g. pedicle screws or pelvic fixation. The low cost and 3D-printable nature have enabled multiple copies to be built. The software is open source allowing replication by any interested party. The simulator has been incorporated successfully into a higher orthopaedic surgical training program.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_16 | Pages 19 - 19
1 Oct 2016
Griffin M Annan J Hamilton D Simpson A
Full Access

3D imaging is commonly employed in the surgical planning and management of bony deformity. The advent of desktop 3D printing now allows rapid in-house production of specific anatomical models to facilitate surgical planning. The aim of this pilot study was to evaluate the feasibility of creating 3D printed models in a university hospital setting.

For requested cases of interest, CT DICOM images on the local NHS Picture Archive System were anonymised and transferred. Images were then segmented into 3D models of the bones, cleaned to remove artefacts, and orientated for printing with preservation of the regions of interest. The models were printed in polylactic acid (PLA), a biodegradable thermoplastic, on the CubeX Duo 3D printer.

PLA models were produced for 4 clinical cases; a complex forearm deformity as a result of malunited childhood fracture, a pelvic discontinuity with severe acetabular deficiency following explantation of an infected total hip replacement, a chronically dislocated radial head causing complex elbow deformity as a result of a severe skeletal dysplasia, and a preoperative model of a deficient proximal tibia as a result of a severe tibia fracture. The models materially influenced clinical decision making, surgical intervention planning and required equipment. In the case of forearm an articulating model was constructed allowing the site of impingement between radius and ulnar to be identified, an osteotomy was practiced on multiple models allowing elimination of the block to supination. This has not previously been described in literature. The acetabulum model allowed pre-contouring of a posterior column plate which was then sterilised and eliminated a time consuming intraoperative step.

While once specialist and expensive, in house 3D printing is now economically viable and a helpful tool in the management of complex patients.