Introduction. The most common bearing couple used in total knee arthroplasty (TKA) is ultra-high molecular weight polyethylene (UHMWPE) articulating against a CoCrMo alloy femoral component. Although this couple has demonstrated good clinical results, UHMWPE wear has been identified as one of the principal causes for long-term failure of total knee joint replacements. 1. indicating a need for improvements in TKA bearings technology. The wear resistance of UHMWPE can be improved by radiation crosslinking; however, in order to get the full benefit of this improved wear resistance, an abrasion resistant ceramic counterface is necessary. 2. Since the radiation crosslinking degrades mechanical properties, it is also important to have an optimized radiation dose and subsequent processing. The purpose of this study was to evaluate the
The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation.Objectives
Methods
Although suction force may have a benefit in reducing the risk of dislocation, it may prevent lubricant recovery between the bearings and will influence the sliding resistance. If the suction force is too high, the head and cup can be held together such that the recovery of synovial fluid is restricted or impossible, even when the hip is not loaded during the swing phase. Both the clearance and the viscosity have a significant effect to determine the suction force and the lubrication of MOM hip joints. It is concluded that suction force is a factor to be considered during the selection of MOM hip joint clearance.
Oxidized zirconium (Oxinium) and highly cross-linked polyethylene (HXLPE) were developed with the purpose of minimizing wear, and subsequent osteolysis, in Total Hip Arthroplasty (THA). However, few articles have been published on long-term results of Oxinium on highly cross-linked polyethylene. The purpose of this investigation is to report minimum 10-year HXLPE wear rates and the clinical outcome of patients in this group and compare this population to a control group of cobalt chrome and ceramic. One hundred forty THAs were performed for 123 patients using an Oxinium head with an HXLPE liner. Ninety-seven had 10 years of clinical follow-up (avg. 14.5). Harris Hip Scores (HHS) were collected preoperatively and at the most recent follow-up. Radiographs of 85 hips were available for a minimum 10-year follow-up (avg. 14.5) and used to calculate wear using PolyWare software. Control groups of cobalt chrome and ceramic articulation on HXLPE with a minimum 10-year follow-up were studied. Clinical follow-up of the Oxinium group showed a statistical improvement compared to preoperative and was similar to the control group of patients. Radiographic evaluation found the linear and volumetric wear rates for the Oxinium group of 0.03 mm/year (range 0.00–0.08) and 3.46 mm3/year (range 1.0 to 15.0) respectively. There was no statistically significant difference in linear or volumetric wear rate between the groups (P-value 0.92 and 0.55 respectively). None of these patients underwent revision of their hip for any reason. Oxinium on highly cross-linked polyethylene has performed exceptionally with wear rates comparable to those of cobalt chrome or ceramic on HXLPE.
Aims. Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the
Acetabular implant position is important for the stability, function, and
Acetabular implant position is important for the stability, function, and
Background. Distal femoral replacements (DFR) are used in children for limb-salvage procedures after bone tumor surgery. These are typically modular devices involving a hinged knee axle that has peripheral metal-on-polyethylene (MoP) and central metal-on-metal (M-M) articulations. While modular connections and M-M surfaces in hip devices have been extensively studied, little is known about
There are pros and cons of all bearing surface options for our young patients. I pick the bearing surface for my young patients trying to maximise durability and minimise risks. For the ultra-young, ≤30 years of age patient, I use ceramic-on-ceramic. The pros of this are the best wear couple available and a favorable track record (with well designed implants). The risks can be minimised: fracture risk now decreased, runaway wear minimised with good surgical technique, impingement problems minimised with good technique and well designed implants, as well as squeaking is minimised with good design (majority of reported squeakers are of one designed socket). I don't use metal-on-metal because I am not willing to subject young patients to potentially 50+ years of high metal ion exposure. I also don't use HCLPE. This would be okay from a biologic standpoint but I still have concerns about
Acetabular implant position is important for the stability, function, and
We live in an era where younger, fitter, more active patients are presenting with the symptoms and signs of degenerative joint disease and require total knee and total hip arthroplasty at a young age. At the same time, this population of patients is living longer and longer and is likely to create new and more complex failure modes for their implants. The ideal solution is a biological one, whereby we can either prevent joint degradation or catch it in its early stages and avoid further deterioration. There may also be advances along the way in terms of partial arthroplasty and focal resurfacing that will help us prevent the need for total joint arthroplasty. There are several tensions that need to be considered. Should we resurface / replace early, particularly now that we have access to navigation and robotics and can effectively customise the implants to the patient's anatomy and their gait pattern? This would allow good function at a young age. Or should we wait as long as possible and risk losing some function for the sake of preserving the first arthroplasty for the lifetime of the patient?. There are some key issues that we still do not fully understand. The lack of true follow-up data beyond 20 or 30 years is worrying. The data available tends to be from expert centers, and always has a dramatic loss to follow-up rate. We worry about bearing surfaces and how those materials will behave over time but we really do not know the effect of chronic metal exposure over several decades, nor do we really understand what happens to bone as it becomes more and more osteopenic and fragile around implants. We have largely recorded but ignored stress shielding, whereas this may become a very significant issue as our patients get older, more fragile, more sarcopaenic and more neurologically challenged. All the fixation debates that we have grappled with, may yet come back to the fore. Can ingrowth lead to failure problems later on? Will more flexible surfaces and materials be required to fit in with the elasticity of bone?. We have failed dramatically at translating the in vitro to the in vivo model. It seems that the in vitro model tells us when failure is going to occur but success in vitro does not predict success in vivo. We, therefore, cannot assume that
To explore the clinical efficacy of using two different types of articulating spacers in two-stage revision for chronic knee periprosthetic joint infection (kPJI). A retrospective cohort study of 50 chronic kPJI patients treated with two types of articulating spacers between January 2014 and March 2022 was conducted. The clinical outcomes and functional status of the different articulating spacers were compared. Overall, 17 patients were treated with prosthetic spacers (prosthetic group (PG)), and 33 patients were treated with cement spacers (cement group (CG)). The CG had a longer mean follow-up period (46.67 months (SD 26.61)) than the PG (24.82 months (SD 16.46); p = 0.001).Aims
Methods
Articulation of the polyethylene (PE) insert between the metal femoral and tibial components in total knee replacements (TKR) results in wear of the insert which can necessitate revision surgery. Continuous PE advancements have improved wear resistance and durability increasing implant longevity. Keeping up with these material advancements, this study utilises model-based radiostereometric analysis (mbRSA) as a tool to measure in vivo short-term linear PE wear to thus predict
Introduction. The
Wear testing of THR has chaperoned generations of improved UHMWPE bearings into wide clinical use. However, previous in vitro testing failed to screen many metal-on-metal hips which failed. This talk tours hip wear testing and associated standards, giving an assortment of THR wear test results from the author's laboratory as examples. Two international hip wear-simulator standards are used: ISO-14242-1 (anatomic configuration) and ISO-14242-3 (orbital-bearing). Both prescribe 5 million (MC) force-motion cycles involving cross-shear synchronized with compression simulating walking gate of ideally aligned THRs. ISO-14242-1 imposes flexion (flex), abduction-adduction (ad-ab) and internal-external (IE) rotations independently and simultaneously. An orbital-bearing simulator more simply rotates either a tilted femoral head or acetabular component, switching from flexion-dominated to ad-ab-dominated phases in each cycle with some IE. In the latter, the acetabular component is typically placed below the femoral head to accentuate abrasive conditions, trapping third-body-wear debris. Wear is measured (ISO-14242-2) gravimetrically (or volumetrically in some hard-on-hard bearings). Wear-rate ranges from negligible to >80mg/MC beyond what causes osteolysis. This mode-1 adhesive wear can therefore “discriminate” to screen hip designs-materials in average conditions. Stair-climbing, sitting, squatting and other activities may cause THR edge-loading and even impingement with smaller head-to-neck ratios or coverage angle, naturally worse in metal-on metal hips. Deformation of thin acetabular components during surgical impaction may cause elevated friction or metal-metal contact, shedding more metal-ions and accelerating failure. Surgical misalignments in inclination angle, version and tilt can make this worse, even during modest activities in hard-on-hard bearings. Abrasive particulate debris from bone or bone-cement, hydroxyapatite, neck-impingement, normal wear, or corrosion can compound the above. Such debris can scratch the femoral head surface, or embed in the UHMWPE liner compromising the wear of even metal-on-plastic hips. Much of the belated standardization activity for higher demand hip testing is in response to the metal-metal failures. ASTM F3047M is a recent non-prescriptive guide for what more rigorous testing can generally be done. Third-body particulate can be intentionally introduced or random scratching of the femoral component surface in extra abrasion testing. Also, the compressive load can be increased, more frequent start-stops to disrupt lubrication, and steepening acetabular shell-liner angles to reduce contact area and cause edge-loading, made harsher when combined with version misalignment. Transient separation can occur between head and liner during the swing phase in a lax THR joint with low coverage angle and misalignments; the separated head impacts the liner rim when reseating. An edge-loading ISO test is currently being discussed where (so-called) “microseparation” to a known distance is directly imposed by a lateral spring force in a hip simulator. Friction testing of a THR in a pendulum-like setup undergoing flexion or abduction swings is being discussed in the ASTM, and so have multi-dimensional THR friction measurements during a
Despite 46 years clinical experience with ceramic-on-ceramic (COC) hip bearings, there is no data on what constitutes a successful
Introduction. Cross-linked polyethylene in total hip arthroplasty has demonstrated excellent
The authors entered patients into a randomised trial to compare the results of the use of cemented and cementless acetabular prostheses between 1993 and 1995. The results of mid-term wear studies at average follow up of eight years were reported in the journal in 2004. We now present long-term results to show the eventual fate of the hip replacements under study. The initial study group of 162 patients was randomly assigned to a modular titanium cup with a polyethylene liner or an all polyethylene cemented cup. All patients received a cemented stem with a 26 mm head and a standardised surgical technique. The polyethylene wear was estimated via head penetration measurement and the mid-term results showed a significantly higher wear rate in the cementless cups compared to the cemented cups (0.15mm/yr vs. 0.07mm/yr p<0.0001). The prediction was that this would lead to a higher rate of aseptic loosening in the cementless group. Patients have now been re-examined at an average of 15 years with the main emphasis on prosthesis survival. Wear studies were also performed. There were exclusions from the initial study because of death and reoperation for reasons other than aseptic loosening. The number of patients in this longer-term study had decreased as a result of death and loss to follow up. Revisions for aseptic loosening did not follow the path as suggested by the mid term wear studies. There were five cup revisions in the cemented group and one cup revision in the cementless group for aseptic loosening. No femoral stem was revised for aseptic loosening. Details of the
Introduction. Wear of polyethylene tibial inserts has been cited as being responsible for up to 25% of revision surgeries, imposing a very significant cost burden on the health care system and increasing patient risk. Accurate measurement of material loss from retrieved knee bearings presents difficult challenges because gravimetric methods are not useful with retrievals and unworn reference dimensions are often unavailable. Geometry and the local anatomy restrict in vivo radiographic wear analysis, and no large-scale analyses have illuminated
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods