header advert
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Bone & Joint Open
Vol. 3, Issue 8 | Pages 656 - 665
23 Aug 2022
Tran T McEwen P Peng Y Trivett A Steele R Donnelly W Clark G

Aims

The mid-term results of kinematic alignment (KA) for total knee arthroplasty (TKA) using image derived instrumentation (IDI) have not been reported in detail, and questions remain regarding ligamentous stability and revisions. This paper aims to address the following: 1) what is the distribution of alignment of KA TKAs using IDI; 2) is a TKA alignment category associated with increased risk of failure or poor patient outcomes; 3) does extending limb alignment lead to changes in soft-tissue laxity; and 4) what is the five-year survivorship and outcomes of KA TKA using IDI?

Methods

A prospective, multicentre, trial enrolled 100 patients undergoing KA TKA using IDI, with follow-up to five years. Alignment measures were conducted pre- and postoperatively to assess constitutional alignment and final implant position. Patient-reported outcome measures (PROMs) of pain and function were also included. The Australian Orthopaedic Association National Joint Arthroplasty Registry was used to assess survivorship.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 48 - 48
1 Mar 2009
Bray R Steele R Newman J Hutabarat S Ackroyd C
Full Access

Purpose of Study: Fixed bearing unicompartmental knee replacement (UKR) has become popular since several series have now shown good 10 year survivorship and excellent function. However little is known about survival during the second decade.

Method: From the Bristol database of over 4000 knee replacements 203 St. Georg Sled UKR’s which had already survived 10 years were identified. The mean age at surgery was 67 years (48–85), with 64% being female. This cohort has been further reviewed at an average of 14.8 years (10–30) from surgery to determine survivorship and function.

Results: Survivorship during the second decade was 87.5%. 58 patients (69 knees) had died with implant in situ and only 2 after revision. A further 15 UKR’s have been revised at an average of 13 years post op; 7 for progression of disease in the opposite compartment, 4 for tibial loosening, 3 for polyethylene wear, 2 for femoral component fracture and 2 for infection.

99 knees were followed for 15 years, 21 knees for 20 years and four for 25 or more years. The average Bristol knee score of the surviving knees fell from 86 to 79 during the second decade, largely as a result of aging.

A previous study of the St Georg. Fixed bearing UKR showed an 89% 10 year survivorship and this is now extended to 82% at 15 years and 76.5% at 20 years.

Conclusion: Satisfactory survival of fixed bearing UKR can be achieved in the second decade suggesting that the indications for mobile bearings require careful definition since there is a higher incidence of complications in many people’s hands.