header advert
Results 1 - 2 of 2
Results per page:
Applied filters
Content I can access

General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 126 - 126
1 Sep 2012
Moroni A Hoque M Micera G Orsini R Nocco E
Full Access

Introduction

Metal-on-polycarbonate urethane (MPU) is a cutting-edge new bearing technology for hip arthroplasty. The acetabular component consists of a 2.7mm-thick polycarbonate-urethane liner inserted into a specially manufactured uncemented titanium shell coated with hydroxyapatite [(HA) Fig. 1]. The liner is pliable and biomechanically mimics human cartilage. In vitro studies have shown minimal wear, fluid film lubrication, physiological load transmission and shock absorption capacity equal to the normal hip. This system includes prosthetic heads of a diameter 12mm less than the socket diameter. The aim of this study was to clinically assess patients treated with this novel technology in a retrospective single centre study.

Methods

Twenty-seven patients with osteoarthritis treated with MPU bearing arthroplasty were included. Mean patient age was 67.9±10.35 years (44–84). Sixteen patients were female and 11 were male. Twenty-four of these had an uncemented HA-coated stem while 3 had a hip resurfacing metal femoral component. All patients were operated on by a single surgeon using a postero-lateral approach.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 151 - 151
1 Jun 2012
Moroni A Hoque M Micera G Sinapi F
Full Access

A retrospective single-center review has been performed to gather clinical data on the use of polycarbonate-urethane (PCU) as an articulating bearing material inside a cobalt-chrome (Co-Cr) press-fit acetabular shell.

As of January 2010, the Co-Cr shell and PCU liner have been implanted into 25 total hip patients which were retrospectively followed. The indications for use were in 24 cases of osteoarthritis, and 1 revision case. No patient was lost to follow-up. The average follow-up time was 17.6 months (range 8-27). The average age of these patients was 67.9 (range 44-84), the sex distribution was 14 female and 11 male patients, of whom 15 were right and 10 left side. 24 patients received a total hip replacement with the metal acetabular system and a cementless femoral stem and 1 patient received the metal acetabular shell coupled to a cemented resurfacing head.

None of the cases has had a dislocation, revision, dislodgement, or infection. At follow-up, the mean Harris hip score was 98 points (80-99). X-rays showed good bone-implant contact without any osteolysis or bone rarefaction.

A detailed review of the clinical data of these patients shows that a PCU liner inserted into a Co-Cr acetabular shell is as safe and effective as other commonly used acetabular shells in other total hip systems currently available. No new or unintended adverse or device-related events were discovered with the clinical use of PCU in a Co-Cr acetabular shell.