header advert
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Hip

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 2 - 2
1 Jun 2016
MacLeod A Sullivan N Whitehouse M Gill R Harinderjit S
Full Access

Introduction

The majority of primary total hip arthroplasty (THA) procedures performed throughout the world use modular junctions, such as the trunnion-head interface; however, the failure of these press-fit junctions is currently a key issue that may be exacerbated by the use of large diameter heads. Several factors are known to influence the strength of the initial connection, however, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection.

Methods

Ti-6Al-4V trunnions (n = 60) and two different sizes of Co-Cr heads (28 mm and 36 mm) were used in the study. Three different levels of assembly force were considered; 4, 5 and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. Finite element and analytical models were also developed to better understand the mechanics of the problem.