header advert
Results 1 - 3 of 3
Results per page:
Bone & Joint Research
Vol. 7, Issue 7 | Pages 485 - 493
1 Jul 2018
Numata Y Kaneuji A Kerboull L Takahashi E Ichiseki T Fukui K Tsujioka J Kawahara N

Objective

Cement thickness of at least 2 mm is generally associated with more favorable results for the femoral component in cemented hip arthroplasty. However, French-designed stems have shown favorable outcomes even with thin cement mantle. The biomechanical behaviors of a French stem, Charnley-Marcel-Kerboull (CMK) and cement were researched in this study.

Methods

Six polished CMK stems were implanted into a composite femur, and one million times dynamic loading tests were performed. Stem subsidence and the compressive force at the bone-cement interface were measured. Tantalum ball (ball) migration in the cement was analyzed by micro CT


Bone & Joint Research
Vol. 6, Issue 5 | Pages 351 - 357
1 May 2017
Takahashi E Kaneuji A Tsuda R Numata Y Ichiseki T Fukui K Kawahara N

Objectives

Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep.

Methods

We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 75 - 75
1 May 2016
Kaneuji A Takahashi E Tsuda R Numata Y Matsumoto T Hirosaki K Takano M
Full Access

Introduction

The French paradox regarding cemented femoral components has not been resolved, so we compared the mechanical behavior of a French stem, the CMK stem (Biomet, Warsaw, IN, USA), with a collarless, polished, tapered stem (CPT, Zimmer, Warsaw, IN, USA) using an original biomechanical instrument.

Materials and Methods

Two size-3 CPT stems and two size-302 CMK stems stems were fixed with bone cement into a composite femur soaked in vegetable oil to simulate wet condition. The composite femur was attached to a biomechanical testing instrument after stem implantation, and a 1-Hz dynamic sine wave load (3000 N) was applied to the stems for a total of 1 million cycles. An 8-hour unload period was set after every 16 hours of load. Femur temperature was maintained at 37°C during testing. The femoral canal was prepared for the CPT stems by standard rasping; for the CMK stems, however, the French method was used, in which cancellous bone was removed with a reamer. One CMK stem (CMK-1) was inserted into a femur without collar contact (>2 mm above the calcar), and the other (CMK-2) was inserted into a femur with collar contact. Stem subsidence was measured at the stem shoulder. Compressive force and horizontal cement movement were measured via rods set at the cement–bone interface on the medial, lateral, anterior, and posterior sides of the proximal and distal portions of the composite femurs.