Osteolysis was found in 36 cups, 18 of them have been revised. The average linear wear was 1.1 (+/− 0.9) mm, the average wear volume 798.7 (+/−622.3) mm3, the linear wear rate 0.07 (+/−0.06) mm/year and the volumetric wear rate 54.5 (+/− 43.2) mm3/year. Younger patients had increased linear wear rates (p=0.035). Osteolysis of the cup, cup migration and cup revision were correlated with linear and volumetric wear (all p=<
0.001). There was no correlation of the cup position, inclination and anteversion with osteolysis, loosening and any of the wear parameters. Volume calculation did not provide further information.
Aim of our study was to analyze cortical bone changes in prosthetic hips with time and compare those changes with the contra lateral non operated femur.
10 patients were not operated on the contralateral hip and were measured in standardized manner in the same locations as in THA femurs.
Clinical and radiological follow-up was planned at 4 months 1, 2, 5, 10, 15 and 20 years. Clinical follow-up included a standardised examination and the completion of an IDES form. Cumulative survival rates were calculated by Kaplan-Meier analysis. Radiographs were analysed for osteolysis according to Gruen et al. (zones 1–7) and radiolucent lines.
The median HHS at the last follow-up was 80 points (range 30–98 points). 31% of the non-revised 41 stems showed osteolysis, most of them in zone 7 (21.9%). Two stems with isolated cup revision had a continuous radiolucent line. Incomplete radiolucent lines appeared most frequently in zone 1. Cup revision was associated with increased radiological changes on the femoral side (p=0.094, Mann-Whitney U-test).