Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 520 - 520
1 Oct 2010
Jibodh S Bragdon C Greene M Kandil A Malchau H
Full Access

Highly cross-linked polyethylene liners in total hip replacement (THR) have allowed the use of larger diameter femoral heads. Larger heads allow for increased range of motion, decreased implant impingement, and protection against dislocation. The purpose of this study is to report the clinical and radiographic outcomes of patients with large femoral heads with HXLPE at 5 years post-op.

A group of 124 patients (132 THRs) who had a primary THR with a 36mm or larger cobalt-chrome femoral head and a Durasul or Longevity liner (Zimmer; Warsaw, IN) were prospectively enrolled in this study. 93 THRs (88 patients) had minimum 5 year follow-up. All patients received a cementless acetabular shell (Trilogy or Inter-op, Zimmer Inc, Warsaw IN) and a highly cross-linked polyethylene liner with an inner diameter of 36 or 38mm. The median radiographic follow-up was 5.6 years (range 5.0–8.0), and patients were assessed clinically by Harris Hip score, UCLA activity score, EQ-5D, and SF-36 functional scores. Femoral head penetration was measured using the Martell Hip Analysis Suite.

No osteolysis was seen in the pelvis or proximal femur, and no components failed due to aseptic loosening. Four patients have questionable signs of bone changes around the acetabular shell with future CT scans scheduled to help reach a final determination. The median acetabular shell abduction and anteversion were 44° (30–66°) and 13° (3–33°) respectively. There was no evidence of cup migration, screw breakage, or eccentric wear on the liner. Regarding the femoral component, there were no episodes of loosening, migration, osteolysis, or fracture. There was no significant difference in the median penetration rate from post-op to longest follow-up between the 36mm (24 patients) and 38mm (4 patients) femoral head groups (0.056±0.10mm/yr and 0.060±0.05mm/yr respectively). Therefore, the data were pooled into one group. Using every post-op to follow-up comparison, the linear regression penetration rate of this combined group was 0.003 mm/yr which is within the error detection of the Martell method. The median femoral head penetration rate during the first post-op year measured 0.59±1.04 mm/yr. In contrast, the median steady state wear rate from the 1yr film to the longest follow-up measured -0.009±0.15mm/yr. A linear regression steady state wear rate from the 1 year film to every follow-up of −0.031 mm/yr indicated no correlation between the magnitude of polyethylene wear and time.

The mid-term results on this series of patients with THRs with a 36 or 38mm femoral head articulating with highly cross-linked polyethylene showed excellent clinical, radiographic, and wear results. The lack of early signs of osteolysis with the use of these large diameter femoral heads is encouraging. Continued and longer-term follow-up is needed to provide survivorship data.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 511 - 511
1 Oct 2010
Bragdon C Burke D Ekeledo A Freiberg A Greene M Harris W Malchau H
Full Access

Highly cross-linked polyethylene (HXLPE) is one of the most widely utilized bearing surfaces for total hip arthroplasty (THA). The first patients to receive XLPE will be 10 years post-op as of December 31, 2008. The purpose of this study is to report the long-term clinical and radiographic outcomes of patients implanted with HXLPE.

A group of 247 primary total hip replacements (224 patients) using HXLPE liners (Longevity or Durasul, Zimmer Inc.) with 22, 26, 28, or 32mm femoral heads were implanted between 1999 and 2001. Clinical evaluation measures included the Harris hip, EQ-5D, SF-36 functional scores, and UCLA activity scores. In addition to plain radiograph assessment, the computerized Martell method was used to measure head penetration over time. A matched group of 241 primary total hip replacements (201 patients) with the same head sizes using conventional polyethylene (PE) with a minimum of 7 years follow-up was used as a Martell method control group. The steady state penetration rate was defined as the slope of the linear regression line of the plot of head penetration from the 1 year film to each subsequent film to discount the early bedding-in process. A student’s t-test was used to compare wear rates between head sizes in each group, and a repeated-measures mixed model ANOVA was used to compare the groups for the 28mm head size.

There were no osteolytic lesions around the cup or stem, and no revisions were performed for polyethylene wear or liner fracture. Clinical outcome scores were averaged: Harris Hip 88.1±11.97, EQ-5D 74.0±27.0, SF-36 physical activity scores 53.3±8.4, SF-36 mental score 46.9±11.1, and UCLA activity 6.4±2.1. The steady state wear of the conventional polyethylene patients increased with time for both 26 and 28mm head sizes (0.144 and 0.127mm/year, respectively). No significant difference was found between the head sizes coupled with conventional polyethylene (p=0.14). Femoral head penetration in the highly cross-linked polyethylene did not increase over time after the first year. The steady state wear rates of HXLPE liners with 28mm or 32mm femoral heads were not significantly different than a slope of zero (p=0.54 for both head sizes).

Clinical follow-up results are typical of a primary THR patient population, and the radiographic results are excellent with no signs of peri-prosthetic osteolysis. Patients with PE show wear rates that are significantly different than zero indicating significant wear of the material. Conversely, patients with HXLPE display no measureable wear at 7–9 years as the wear rates were within the error detection of the Martell method. This long-term clinical and radiographic follow-up study for this new bearing material shows excellent clinical outcomes with very low in vivo wear.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 295 - 295
1 May 2010
Bragdon C Greene M Goldvasser D Freiberg A Malchau H
Full Access

Introduction: Electron beam irradiated highly cros-slinked polyethylene has been used in total hip arthroplasty for over 8 years. Due to its low wear characteristics, the use of femoral heads that are greater than 32mm in diameter is now available, allowing for an increase in range of motion and increased stability against dislocation when necessary. The purpose of this study is to provide a summary report on the radiographic analysis of the longest term data available on primary THR patients receiving highly cross-linked polyethylene and to compare the results of two methods of measuring femoral head penetration.

Methods: Three prospective clinical studies involving electron beam irradiated highly cross-linked polyethylene have been initiated at our center. To date, the results of: 200 hips with a minimum of 6 year follow-up with conventional sized femoral heads (primarily 28 and 32mm); 45 hips with minimum 5 year follow-up with larger sized femoral heads (primarily 36 and 38mm); and 30 hips with 5 year follow-up enrolled in a Radiostereometric analysis (RSA) study (15 patients with 28mm and 15 patients with 36mm diameter femoral heads); were available for this summary report. Data from patient administered questionnaires (Harris Hip score, UCLA activity score, and WOMAC), radiographic review, and wear analysis using RSA or the Martell Hip Analysis Suite™. In addition, for comparative purposes, wear was measured in a subset of patients using the Devane Polyware™ program.

Results: All hips had good clinical outcome at longest follow-up regardless of which femoral head size was used. There were no revisions due to polyethylene wear and no evidence of peri-prosthetic osteolysis. In general, after the bedding in period, there was no significant increase in femoral head penetration regardless of head size.

For RSA, the wear rate for the 28mm femoral head group was 0.05±0.02 while the 36mm femoral head group was 0.03±0.02, p=0.13.

For the Martell analysis, the average steady-state wear rate was −0.002 ± 0.01 mm per year and −0.026 ± 0.13 mm per year for 28mm and 32mm head sizes, respectively, p=0.62. There was no correlation between wear and time in situ or femoral head size for any of the clinical studies.

In comparing the Martell and Devane programs, the total average wear rates were significantly different, 0.07±0.05 and 0.03±0.06mm/year respectively, p=0.01. However, when the absolute values of the Martell results were used, there was no difference, p=0.22.

Conclusion: The mid-term follow-up of a large group of primary THR patients receiving highly cross-linked polyethylene components have shown no problems related to the new bearing material. Extremely low wear and lack of peri-prosthetic osteolysis are encouraging results requiring further long-term study.