Vol. 94-B, Issue SUPP_XL | Pages 108 - 108
1 Sep 2012
Meccia B Spencer E Zingde S Sharma A Lesko F Mahfouz M Komistek R
Full Access

INTRODUCTION

Total shoulder arthroplasty (TSA) implants are used to restore function to individuals whose shoulder motions are impaired by osteoarthritis. To improve TSA implant designs, it is crucial to understand the kinematics of healthy, osteoarthritic (OA), and post-TSA shoulders. Hence, this study will determine in vivo kinematic trends of the glenohumeral joints of healthy, OA, and post-TSA shoulders.

Methods

In vivo shoulder kinematics were determined pre and post-operatively for five unilateral TSA subjects with one healthy and a contralateral OA glenohumeral joint. Fluoroscopic examinations were performed for all three shoulder categories (healthy, OA, and post-TSA) for each subject shoulder abduction and external rotation. Then, three-dimensional (3D) models of the left and right scapula and humerus were constructed using CT scans. For post-operative shoulders, 3D computer-aided design models of the implants were obtained. Next, the 3D glenohumeral joint kinematics were determined using a previously published 3D to 2D registration technique. After determining kinematics, relative Euler rotation angles between the humerus and scapula were calculated in MATLABĀ® to determine range of motion (ROM) and kinematic profiles for all three shoulder categories. The ROMs for each category were compared using paired t-tests for each exercise.

Also, the location of the contact point of the humerus on the glenoid was found. This allowed the vertical translation from the most superior to most inferior contact point (SI contact range) to be calculated as well as the horizontal translation from the most anterior to most posterior contact point (AP contact range). The SI and AP contact ranges for all shoulder categories were compared using paired t-tests for each exercise.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 39 - 39
1 Sep 2012
Leszko F Zingde S Argenson J Dennis D Wasielewski R Mahfouz M Komistek R De Bock T
Full Access

Anterior knee pain is one of the most frequently reported musculoskeletal complaints in all age groups. However, patient's complaints are often nonspecific, leading to difficulty in properly diagnosing the condition. One of the causes of pain is the degeneration of the articular cartilage. As the cartilage deteriorates, its ability to distribute the joint reaction forces decreases and the stresses may exceed the pain threshold. Unfortunately, the assessment of the cartilage condition is often limited to a detailed interview with the patient, careful physical examination and x-ray imaging. The X-ray screening may reveal bone degeneration, but does not carry sufficient information of the soft tissues' conditions. More advanced imaging tools such as MRI or CT are available, but these are expensive, time consuming and are only suitable for detection of advanced arthritis. Arthroscopic surgery is often the only reliable option, however due to its semi-invasive nature, it cannot be considered as a practical diagnostic tool. However, as the articular cartilage degenerates, the surfaces become rougher, they produce higher vibrations than smooth surfaces due to higher friction during the interaction. Therefore, it was proposed to detect vibrations non-invasively using accelerometers, and evaluate the signals for their potential diagnostic applications.

Vibration data was collected for 75 subjects; 23 healthy and 52 subjects suffering from knee arthritis. The study was approved by the IRB and an Informed Consent was obtained prior to data collection. Five accelerometers were attached to skin around the knee joint (at the patella, medial and lateral femoral condyles, tibial tuberosity and medial tibial plateau). Each subject performed 5 activities; (1) flexion-extension, (2) deep knee bend, (3) chair rising, (4) stair climbing and (5) stair descent. The vibration and motion components of the signals were separated by a high pass filter. Next, 33 parameters of the signals were calculated and evaluated for their discrimination effectiveness (Figure 1). Finally the pattern recognition method based on Baysian classification theorem was used for classify each signal to either healthy or arthritic group, assuming equal prior probabilities.

The variance and mean of the vibration signals were significantly higher in the arthritic group (p=2.8e-7 and p=3.7e-14, respectively), which confirms the general hypothesis that the vibration magnitudes increase as the cartilage degenerates. Other signal features providing good discrimination included the 99th quantile, the integral of the vibration signal envelope, and the product of the signal envelope and the activity duration. The pattern classification yielded excellent results with the success rate of up to 92.2% using only 2 features, up to 94.8% using 3 (Figure 2), and 96.1% using 4 features.

The current study proved that the vibrations can be studied non-invasively using a low-cost technology. The results confirmed the hypothesis that the degeneration of the cartilage increases the vibration of the articulating bones. The classification rate obtained in the study is very encouraging, providing over 96% accuracy. The presented technology has certainly a potential of being used as an additional screening methodology enhancing the assessment of the articular cartilage condition.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 118 - 118
1 Sep 2012
Nakamura S Nakamura T Kobayashi M Ito H Ikeda N Nakamura K Komistek R
Full Access

Introduction

Achieving high flexion after total knee arthroplasty is very important for patients in Asian countries where deep flexion activities are an important part of daily life. The Bi-Surface Total Knee System (Japan Medical Material, Kyoto, Japan), which has a unique ball-and-socket mechanism in the mid-posterior portion of the femoral and tibial components, was designed to improve deep knee flexion and long-term durability after total knee arthroplasty (Figure 1). The purpose of this study was to determine the in vivo three dimensional kinematics of Bi-Surface Total Knee System in order to evaluate and analyze the performance of this system with other conventional TKA designs currently available in the market today.

Materials and Methods

Three dimensional kinematics were evaluated during a weight-bearing deep knee bend activity using fluoroscopy and a 2D-to-3D registration technique for 66 TKA. Each knee was analyzed to determine femorotibial kinematics, including weight-bearing range of motion, anterior/posterior contact position, and tibio-femoral rotation.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 26 - 26
1 Sep 2012
Carr C Cheng J Sharma A Mahfouz M Komistek R
Full Access

Introduction

Numerous studies have been conducted to investigate the kinematics of the lumbar spine, and while many have documented its intricacies, few have analyzed the complex coupled out-of-plane rotations inherent in the low back. Some studies have suggested a possible relationship between patients having low back pain (LBP) or degenerative conditions in the lumbar region and various degrees of restricted, excessive, or poorly-controlled lumbar motion. Conversely, others in the orthopedic community maintain there has been no distinct correlation found between spinal mobility and clinical symptoms. The objective of this study was to evaluate both the in-plane and coupled out-of-plane rotational magnitudes about all three motion axes in both symptomatic and asymptomatic patients.

Methods

Ten healthy, 10 LBP, and 10 degenerative patients were CT scanned and evaluated under fluoroscopic surveillance while performing flexion/extension of the lumbar spine. Three-dimensional, patient-specific bone models were created and registered to fluoroscopic images using a 3D-to-2D model fitting algorithm. In vivo kinematics were derived at specified increments and the overall in-plane flexion/extension and coupled out-of-plane rotations were analyzed using two techniques. The first method derived the maximal absolute rotational magnitude (MARM) at each level by subtracting the rotational motion in the increment exhibiting the most negative or least amount of rotation from the increment having the greatest amount of rotation. The second method was designed to isolate the path of rotation (POR) of the vertebrae at each level while performing the prescribed flexion/extension activity. By tracking the rotational path of the cephaled vertebrae as it articulated upon the more caudal vertebrae and summing the absolute rotation between each increment about each axis the POR was calculated over the entire flexion/extension activity.