Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 3 - 3
1 Jun 2016
Laura AD Whittaker R Hothi H Kwon Y Skinner J Hart A
Full Access

Introduction

Dual-mobility bearings increase the stable range of motion of total hip arthroplasty (THA) but are limited by the mechanical effects of a large diameter metal on polyethylene bearing which may cause high rates of wear from the surfaces of the polyethylene bearing and the head-stem taper. Improved polyethylene (PE) has reduced concern over bearing wear but the effects on the taper junction are unknown.

We aimed to better understand the effect of dual mobility bearings on fretting-corrosion damage to the taper junction by comparison to standard bearings.

Materials and Methods

We collected and analysed retrieved hips of one design with either dual mobility (n= 39) or standard bearings (n=30). The bearing size in the dual mobility group was 42mm whereas in the standard bearing group it had a median of 36mm. Stem trunnions had V40 tapers. Time of implantation and body mass index were comparable between the two groups.

Fretting and corrosion at the stem trunnions was quantified by: 1) visual scoring and 2) surface profilometry.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 25 - 25
1 Jun 2016
Ferguson D Henckel J Holme T Berber R Matthews W Carrington R Miles J Mitchell P Jagiello J Skinner J Hart A
Full Access

Introduction

Surgical simulation and ‘virtual’ surgical tools are becoming recognised as essential aids for speciality training in Trauma & Orthopaedics, as evidenced by the BOA T&O Simulation Curriculum 20131,2. The current generation of hip arthroplasty simulators, including cadaveric workshops, offers the trainee limited exposure to reproducible real life bony pathology. We developed and implemented a novel training course using pathological dry bone models generated from real patient cases to support senior orthopaedic trainees and new consultants in developing knowledge and hands on skills in complex total hip arthroplasty.

Patient/Materials & Methods

A two-day programme for 20 delegates was held at a specialist centre for hip arthroplasty. Three complex femoral and three complex acetabular cases were identified from patients seen at our centre. 3D models were printed from CT scans and dry bone models produced (using a mold-casting process), enabling each delegate to have a copy of each case at a cost of around £30 per case per delegate (Figure 1). The faculty was led by 4 senior Consultant revision hip surgeons. A computerised digitising arm was used to measure cup positioning and femoral stem version giving candidates immediate objective feedback (Figure 2). Candidate experience and satisfaction with the course and models was evaluated with a standardised post-course questionnaire.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 27 - 27
1 Nov 2015
Berber R Khoo M Carrington R Miles J Skinner J Hart A
Full Access

Introduction

Uncertainties in the management of patients with MOM hip implants continue to be a problem for all surgeons. Guidelines vary and do not fully define or quantify thresholds. We aimed to assess the differences in decision-making amongst an international community of six specialist orthopaedic institutions.

Methods

Five international tertiary referral orthopaedic units (one UK, two USA, and two European) were invited to participate. Each unit organised an MDT panel consisting of 2 or more hip surgeons and a musculoskeletal radiologist. All units discussed the same 10 patients. A full clinical dataset was provided including blood test and all imaging. Differences in the interpretation of findings, management decision and rationale for decisions were compared between institutions.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 4 - 4
1 Nov 2015
Osman K Panagiotidou A Meswania J Skinner J Hart A Haddad F Blunn G
Full Access

Introduction

Recent studies on large diameter femoral head hip replacements have implicated the modular taper junction as one of the significant sources of wear and corrosion products and this has been attributed to increased torque and bending on the taper interface. The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper junction and to investigate whether different material combinations also had an effect.

Patients/Materials & Methods

We examined 1) Cobalt Chromium (CoCr) heads on CoCr stems 2) CoCr heads on Titanium alloy (Ti) stems and 3) Ceramic heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the femoral stem in the anterior posterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm where the force generated was equivalent to 0Nm, 9Nm, 14Nm and 18Nm. In Test 2 we investigated the effect of increasing bending moment by offsetting the application of axial load from the midline in the medial-lateral (ML). Offset increments equivalent to +0, +7 and +14 heads were used. For each test we used n=3 for each different material combination.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 26 - 26
1 Nov 2015
Skinner J Sabah S Henckel J Cook E Hothi H Hart A
Full Access

Introduction

The National Joint Registry (NJR) for England, Wales and Northern Ireland contributes important information on the performance of implants and surgeons. However, the quality of this data is not known. This study aimed to perform an independent validation of primary metal-on-metal hip procedures recorded on the NJR through linkage to the London Implant Retrieval Centre (LIRC).

Patients/Materials & Methods

Primary, metal-on-metal hip arthroplasties performed between 1st April 2003 and 5th November 2013 were recruited from the NJR (n=67045). Retrieved, metal-on-metal components were recruited from the LIRC (n=782). Data linkage and validation checks were performed.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 2 - 2
1 Nov 2015
Govind G Henckel J Hothi H Sabah S Skinner J Hart A
Full Access

Introduction

Retrieved metal-on-metal acetabular components are invaluable resources from which to investigate the wear behaviour of failed hip implants. New forensic and investigative techniques continue to be developed to help the surgeon further understand factors which contribute to early failure. We have developed a novel technique to locate the in vivo location of the primary wear scar of an explanted cup.

Patients/Materials & Methods

Thirteen (13) patients with failed metal hip resurfacings were recruited and their acetabular components retrieved. A 3D wear map was generated and the precise location of the primary wear scar in each cup was identified using a coordinate measuring machine (CMM). This wear scar position and location was noted in relation to standard landmarks on the acetabular cup. All patients underwent a computerised tomography (CT) scan prior to revision surgery. The 3D positional map from the CMM was then co-registered with the implant on the patient's pelvic 3D CT scan.