Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Bone & Joint Research
Vol. 7, Issue 3 | Pages 226 - 231
1 Mar 2018
Campi S Mellon SJ Ridley D Foulke B Dodd CAF Pandit HG Murray DW

Objectives

The primary stability of the cementless Oxford Unicompartmental Knee Replacement (OUKR) relies on interference fit (or press fit). Insufficient interference may cause implant loosening, whilst excessive interference could cause bone damage and fracture.

The aim of this study was to identify the optimal interference fit by measuring the force required to seat the tibial component of the cementless OUKR (push-in force) and the force required to remove the component (pull-out force).

Materials and Methods

Six cementless OUKR tibial components were implanted in 12 new slots prepared on blocks of solid polyurethane foam (20 pounds per cubic foot (PCF), Sawbones, Malmo, Sweden) with a range of interference of 0.1 mm to 1.9 mm using a Dartec materials testing machine HC10 (Zwick Ltd, Herefordshire, United Kingdom) . The experiment was repeated with cellular polyurethane foam (15 PCF), which is a more porous analogue for trabecular bone.


The Bone & Joint Journal
Vol. 98-B, Issue 10_Supple_B | Pages 3 - 10
1 Oct 2016
Hamilton TW Pandit HG Lombardi AV Adams JB Oosthuizen CR Clavé A Dodd CAF Berend KR Murray DW

Aims

An evidence-based radiographic Decision Aid for meniscal-bearing unicompartmental knee arthroplasty (UKA) has been developed and this study investigates its performance at an independent centre.

Patients and Methods

Pre-operative radiographs, including stress views, from a consecutive cohort of 550 knees undergoing arthroplasty (UKA or total knee arthroplasty; TKA) by a single-surgeon were assessed. Suitability for UKA was determined using the Decision Aid, with the assessor blinded to treatment received, and compared with actual treatment received, which was determined by an experienced UKA surgeon based on history, examination, radiographic assessment including stress radiographs, and intra-operative assessment in line with the recommended indications as described in the literature.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 345 - 349
1 Mar 2014
Liddle AD Pandit HG Jenkins C Lobenhoffer P Jackson WFM Dodd CAF Murray DW

The cementless Oxford unicompartmental knee replacement has been demonstrated to have superior fixation on radiographs and a similar early complication rate compared with the cemented version. However, a small number of cases have come to our attention where, after an apparently successful procedure, the tibial component subsides into a valgus position with an increased posterior slope, before becoming well-fixed. We present the clinical and radiological findings of these six patients and describe their natural history and the likely causes. Two underwent revision in the early post-operative period, and in four the implant stabilised and became well-fixed radiologically with a good functional outcome.

This situation appears to be avoidable by minor modifications to the operative technique, and it appears that it can be treated conservatively in most patients.

Cite this article: Bone Joint J 2014;96-B:345–9.