Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
Bone & Joint Open
Vol. 3, Issue 9 | Pages 716 - 725
15 Sep 2022
Boulton C Harrison C Wilton T Armstrong R Young E Pegg D Wilkinson JM

Data of high quality are critical for the meaningful interpretation of registry information. The National Joint Registry (NJR) was established in 2002 as the result of an unexpectedly high failure rate of a cemented total hip arthroplasty. The NJR began data collection in 2003. In this study we report on the outcomes following the establishment of a formal data quality (DQ) audit process within the NJR, within which each patient episode entry is validated against the hospital unit’s Patient Administration System and vice-versa. This process enables bidirectional validation of every NJR entry and retrospective correction of any errors in the dataset. In 2014/15 baseline average compliance was 92.6% and this increased year-on-year with repeated audit cycles to 96.0% in 2018/19, with 76.4% of units achieving > 95% compliance. Following the closure of the audit cycle, an overall compliance rate of 97.9% was achieved for the 2018/19 period. An automated system was initiated in 2018 to reduce administrative burden and to integrate the DQ process into standard workflows. Our processes and quality improvement results demonstrate that DQ may be implemented successfully at national level, while minimizing the burden on hospitals.

Cite this article: Bone Jt Open 2022;3(9):716–725.


Bone & Joint Research
Vol. 12, Issue 6 | Pages 362 - 371
1 Jun 2023
Xu D Ding C Cheng T Yang C Zhang X

Aims

The present study aimed to investigate whether patients with inflammatory bowel disease (IBD) undergoing joint arthroplasty have a higher incidence of adverse outcomes than those without IBD.

Methods

A comprehensive literature search was conducted to identify eligible studies reporting postoperative outcomes in IBD patients undergoing joint arthroplasty. The primary outcomes included postoperative complications, while the secondary outcomes included unplanned readmission, length of stay (LOS), joint reoperation/implant revision, and cost of care. Pooled odds ratios (ORs) and 95% confidence intervals (CIs) were calculated using a random-effects model when heterogeneity was substantial.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


Bone & Joint Research
Vol. 3, Issue 3 | Pages 60 - 68
1 Mar 2014
Langton DJ Sidaginamale RP Holland JP Deehan D Joyce TJ Nargol AVF Meek RD Lord JK

Objectives

Wear debris released from bearing surfaces has been shown to provoke negative immune responses in the recipient. Excessive wear has been linked to early failure of prostheses. Analysis using coordinate measuring machines (CMMs) can provide estimates of total volumetric material loss of explanted prostheses and can help to understand device failure. The accuracy of volumetric testing has been debated, with some investigators stating that only protocols involving hundreds of thousands of measurement points are sufficient. We looked to examine this assumption and to apply the findings to the clinical arena.

Methods

We examined the effects on the calculated material loss from a ceramic femoral head when different CMM scanning parameters were used. Calculated wear volumes were compared with gold standard gravimetric tests in a blinded study.