Aims. The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre. Methods. All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery. Results. A total of 829 stems in 745 patients were included in the dataset; 155 patients (20.8%) died within ten years, and of the remaining 664 stems, 648 stems (97.6%) remained in situ beyond ten years. For the 21 patients (2.5%) undergoing revision surgery, 16 femoral stems (1.9%) were revised and 18 acetabular components (2.2%) were revised. Indications for revision in order of decreasing frequency were infection (n = 6), pain (n = 6), aseptic component loosening (n = 3), periprosthetic fracture (n = 3),
Periprosthetic hip and knee infection remains one of the most severe complications following arthroplasty, with an incidence between 0.5% to 1%. This study compares the outcomes of revision surgery for periprosthetic joint infection (PJI) following hip and knee arthroplasty prior to and after implementation of a specialist PJI multidisciplinary team (MDT). Data was retrospectively analyzed from a single centre. In all, 29 consecutive joints prior to the implementation of an infection MDT in November 2016 were compared with 29 consecutive joints subsequent to the MDT conception. All individuals who underwent a debridement antibiotics and implant retention (DAIR) procedure, a one-stage revision, or a two-stage revision for an acute or chronic PJI in this time period were included. The definition of successfully treated PJI was based on the Delphi international multidisciplinary consensus.Aims
Methods
The COVID-19 pandemic posed significant challenges to healthcare systems across the globe in 2020. There were concerns surrounding early reports of increased mortality among patients undergoing emergency or non-urgent surgery. We report the morbidity and mortality in patients who underwent arthroplasty procedures during the UK first stage of the pandemic. Institutional review board approval was obtained for a review of prospectively collected data on consecutive patients who underwent arthroplasty procedures between March and May 2020 at a specialist orthopaedic centre in the UK. Data included diagnoses, comorbidities, BMI, American Society of Anesthesiologists grade, length of stay, and complications. The primary outcome was 30-day mortality and secondary outcomes were prevalence of SARS-CoV-2 infection, medical and surgical complications, and readmission within 30 days of discharge. The data collated were compared with series from the preceding three months.Aims
Methods
Wear debris released from bearing surfaces has been shown to
provoke negative immune responses in the recipient. Excessive wear
has been linked to early failure of prostheses. Analysis using coordinate
measuring machines (CMMs) can provide estimates of total volumetric
material loss of explanted prostheses and can help to understand
device failure. The accuracy of volumetric testing has been debated,
with some investigators stating that only protocols involving hundreds
of thousands of measurement points are sufficient. We looked to
examine this assumption and to apply the findings to the clinical
arena. We examined the effects on the calculated material loss from
a ceramic femoral head when different CMM scanning parameters were
used. Calculated wear volumes were compared with gold standard gravimetric
tests in a blinded study. Objectives
Methods