Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


Bone & Joint Open
Vol. 6, Issue 2 | Pages 186 - 194
13 Feb 2025
Battaglia AG D'Apolito R Ding BTK Tonolini S Ramazzotti J Zagra L

Aims

Revision hip arthroplasty for femoral stem loosening remains challenging due to significant bone loss and deformities requiring specialized revision stems. The aim of this study was to evaluate the clinical and radiological outcomes, and survival, of a consecutive series of femoral revisions performed using a primary cementless stem with tapered geometry and rectangular cross-section at medium-term follow-up.

Methods

We retrospectively evaluated 113 patients (115 hips) with intraoperative Paprosky type I (n = 86) or II (n = 29) defects, who underwent femoral revision with Alloclassic Zweymüller SL stem for one-stage aseptic revision or two-stage septic revision from January 2011 to December 2020. The mean follow-up was 77.9 months (SD 33.8). Nine patients were lost to follow-up (deceased or not available), leaving 104 patients (106 hips) for the clinical and radiological analysis. Clinical assessment was performed with Harris Hip Score (HHS) and visual analogue scale (VAS) before surgery and at final follow-up.