Bone & Joint Research
Vol. 6, Issue 1 | Pages 52 - 56
1 Jan 2017
Hothi HS Kendoff D Lausmann C Henckel J Gehrke T Skinner J Hart A

Objectives

Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces.

Methods

In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons.


The Bone & Joint Journal
Vol. 97-B, Issue 10_Supple_A | Pages 20 - 29
1 Oct 2015
Gehrke T Alijanipour P Parvizi J

Periprosthetic joint infection (PJI) is one of the most feared and challenging complications following total knee arthroplasty. We provide a detailed description of our current understanding regarding the management of PJI of the knee, including diagnostic aids, pre-operative planning, surgical treatment, and outcome.

Cite this article: Bone Joint J 2015;97-B(10 Suppl A):20–9.