Aims

The aim of this investigation was to compare risk of infection in both cemented and uncemented hemiarthroplasty (HA) as well as in total hip arthroplasty (THA) following femoral neck fracture.

Methods

Data collection was performed using the German Arthroplasty Registry (EPRD). In HA and THA following femoral neck fracture, fixation method was divided into cemented and uncemented prostheses and paired according to age, sex, BMI, and the Elixhauser Comorbidity Index using Mahalanobis distance matching.


Bone & Joint Research
Vol. 5, Issue 1 | Pages 18 - 25
1 Jan 2016
Sims AL Parsons N Achten J Griffin XL Costa ML Reed MR

Background. Approximately half of all hip fractures are displaced intracapsular fractures. The standard treatment for these fractures is either hemiarthroplasty or total hip arthroplasty. The recent National Institute for Health and Care Excellence (NICE) guidance on hip fracture management recommends the use of ‘proven’ cemented stem arthroplasty with an Orthopaedic Device Evaluation Panel (ODEP) rating of at least 3B (97% survival at three years). The Thompsons prosthesis is currently lacking an ODEP rating despite over 50 years of clinical use, likely due to the paucity of implant survival data. Nationally, adherence to these guidelines is varied as there is debate as to which prosthesis optimises patient outcomes. Design. This study design is a multi-centre, multi-surgeon, parallel, two arm, standard-of-care pragmatic randomised controlled trial. It will be embedded within the WHiTE Comprehensive Cohort Study (ISRCTN63982700). The main analysis is a two-way equivalence comparison between Hemi-Thompson and Hemi-Exeter polished taper with Unitrax head. Secondary outcomes will include radiological leg length discrepancy measured as per Bidwai and Willett, mortality, re-operation rate and indication for re-operation, length of index hospital stay and revision at four months. This study will be supplemented by the NHFD (National Hip Fracture Database) dataset. Discussion. Evidence on the optimum choice of prosthesis for hemiarthroplasty of the hip is lacking. National guidance is currently based on expert opinion rather than empirical evidence. The incidence of hip fracture is likely to continue to increase and providing high quality evidence on the optimum treatment will improve patient outcomes and have important health economic implications. Cite this article: A. L. Sims. The World Hip Trauma Evaluation Study 3: Hemiarthroplasty Evaluation by Multicentre Investigation – WH. I. TE 3: HEMI – An Abridged Protocol. Bone Joint Res 2016;5:18–25. doi: 10.1302/2046-3758.51.2000473


Bone & Joint Research
Vol. 9, Issue 12 | Pages 848 - 856
1 Dec 2020
Ramalhete R Brown R Blunn G Skinner J Coathup M Graney I Sanghani-Kerai A

Aims

Periprosthetic joint infection (PJI) is a debilitating condition with a substantial socioeconomic burden. A novel autologous blood glue (ABG) has been developed, which can be prepared during surgery and sprayed onto prostheses at the time of implantation. The ABG can potentially provide an antimicrobial coating which will be effective in preventing PJI, not only by providing a physical barrier but also by eluting a well-known antibiotic. Hence, this study aimed to assess the antimicrobial effectiveness of ABG when impregnated with gentamicin and stem cells.

Methods

Gentamicin elution from the ABG matrix was analyzed and quantified in a time-dependent manner. The combined efficiency of gentamicin and ABG as an anti-biofilm coating was investigated on titanium disks.


Bone & Joint Research
Vol. 7, Issue 5 | Pages 357 - 361
1 May 2018
Shin T Lim D Kim YS Kim SC Jo WL Lim YW

Objectives

Laser-engineered net shaping (LENS) of coated surfaces can overcome the limitations of conventional coating technologies. We compared the in vitro biological response with a titanium plasma spray (TPS)-coated titanium alloy (Ti6Al4V) surface with that of a Ti6Al4V surface coated with titanium using direct metal fabrication (DMF) with 3D printing technologies.

Methods

The in vitro ability of human osteoblasts to adhere to TPS-coated Ti6Al4V was compared with DMF-coating. Scanning electron microscopy (SEM) was used to assess the structure and morphology of the surfaces. Biological and morphological responses to human osteoblast cell lines were then examined by measuring cell proliferation, alkaline phosphatase activity, actin filaments, and RUNX2 gene expression.


Bone & Joint Research
Vol. 7, Issue 1 | Pages 103 - 104
1 Jan 2018
Young PS Patil S Meek RMD


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.