Abstract
Total knee arthroplasty (TKA) is widely accepted as a successful treatment option for the pain and limitation of function associated with severe joint disease. The ideal knee arthroplasty implant should provide reliable pain relief and normal levels of functional strength and range of motion. However, there are still a number of implant-specific problems following knee arthroplasty, such as irregular kinematics, polyethylene wear and poor range of motion.
MRI and cadaveric studies have highlighted important kinematics during movement of the native knee. In particular, flexion of the joint results in a phenomenon referred to as “roll back and slide”. This essentially describes posterior translation of the femur on the tibia which in turn has a two-fold biomechanical function: to increase the lever arm of the quadriceps and allow clearance of the femur from the tibia in deep flexion. During extension of the joint, the femur rolls forward increasing the lever arm of the hamstrings to act as a brake on hyperextension.
Additional rotation of the joint arises in the axial plane. This is attributed to the concave tibial plateau and relatively fixed meniscus on the medial compartment of the joint in comparison to a lateral convex plateau with a mobile meniscus. This asymmetry allows axial rotation of the lateral compartment over the medial compartment by up to 30 degrees. Subsequently, from extension to full flexion the tibia rotates internally on the femur and vice versa. The external rotation of the tibia on the femur that occurs during the terminal degrees of knee extension is often referred to as the “screw home mechanism” and results in tightening of both the cruciate ligaments locking the knee such that the tibia is in a position of maximum stability on the femur.
Numerous studies over the past two decades have characterised the in-vivo motions of knee replacements. Major conclusions from these studies are that the motion after knee arthroplasty generally does not replicate normal knee motions. In particular, many kinematic studies of unconstrained devices have demonstrated the femur sliding forwards rather than backwards with flexion. This paradoxical movement is also seen in many posterior cruciate retaining knees. This in turn has a negative outcome in range of movement, particularly in light of fluoroscopic studies highlighting strong positive correlations in weight-bearing flexion with femoral roll back. In contrast knee arthroplasties that retain both cruciate ligaments come closest to replicating normal knee motion and furthermore, provide greater stair climbing stability. It may therefore be presumed that this excessive AP motion in a well-designed prosthesis is attributed to a loss in the natural intrinsic stabilizing structures.
A number of studies to date have also highlighted close correlation between knee kinematics and functional strength. Generally, patients with knee replacement exhibit a significant loss of strength compared to normal. The common experimental findings is that knees with the highest intrinsic stability, whether provided by retained ligaments, conforming articular surfaces or post-cam substitution, exhibit the greatest functional strength in high-demand activities in TKA patients.
On the basis of this knowledge, it would be intuitive to choose a TKA design that attempts to restore natural knee joint stability. The medially conforming ‘ball and socket’ articulation of the medial tibio-femoral compartment is a design concept thought to provide stability through the complete arc of knee flexion. Clinical and biomechanical data from a number of centers suggests that this has been a successful evolution in TKA that will continue to benefit patients.