CONSTRAINED LINERS: YESTERDAY'S SOLUTION?

The Current Concepts in Joint Replacement (CCJR) Winter Meeting, 14 – 17 December 2016.



Abstract

Instability after total hip arthroplasty is the most common indication for revision arthroplasty and can be difficult to treat. The purpose of this study is to evaluate the outcomes associated with the use of a constrained acetabular component as a treatment for instability after hip arthroplasty. We reviewed the clinical and radiographic outcomes of 149 arthroplasties, that had been performed with use of a single design of constrained acetabular component between 2007 and 2012 at a single institution. Patient demographics and case specific data were collected The Mann-Whitney U test was used to assess continuous variables. Categorical variables were examined using the Chi-square test and Fisher's exact test when appropriate. Survival probability was calculated using the Kaplan-Meier method.

The mean age at time of index surgery was 70 years, 65% were female, and mean BMI was 26.3. The average number of previous surgeries was 3.6. The constrained liner was cemented into a well-fixed cup in 40 hips (20%). In eighty-two (55%) hips the constrained component was implanted for the treatment of recurrent instability, and in sixty-seven (45%) hips it was implanted because the hips demonstrate instability during revision surgery. At an average duration of follow-up of 4.2 (2–7) years the overall revision rate was 10.6 % The constrained acetabular device eliminated or prevented hip instability in all patients except five; 3.3% had a new dislocation and six (4.0%) had failure of the retentive ring. Three revisions (2%) were performed for deep infection, and 2 (1.3%) for acetabular component loosening. Radiographic analysis revealed a non-progressive radiolucent line around the cup in 19 hips (12.7%). When stratified by patient age, survivorship for patients less than 65 years of age versus those greater than 65 years was similar.

This study correlates with results of other papers in the literature looking at outcome of constrained tripolar type sockets. The focal constraint socket with a metal ring type design has a much greater failure rate (9–29%). Constrained liners remain an excellent option for hip instability in early to mid- term follow up.