Introduction

Shoulder arthroplasty is used to treat several common pathologies of the shoulder, including osteoarthritis, post-traumatic arthritis, and avascular necrosis. In replacement of the humeral head, modular components allow for anatomical variations, including retroversion angle and head-neck angle. Surgical options include an anatomic cut or a guide-assisted cut at a fixed retroversion and head-neck angle, which can vary the dimensions of the cut humeral head (height, anteroposterior (AP), and superoinferior (SI) diameters) and lead to negative long term clinical results. This study measures the effect of guide-assisted osteotomies on humeral head dimensions compared to anatomic dimensions.

Methods

Computed tomography (CT) scans from 20 cadaveric shoulder specimens (10 male, 10 female; 10 left) were converted to three-dimensional models using medical imaging software. An anatomic humeral head cut plane was placed at the anatomic head – neck junction of all shoulders by a fellowship trained shoulder surgeon. Cut planes were generated for each of the standard implant head-neck angles (125°, 130°, 135°, and 140°) and retroversion angles (20°, 30°, and 40°) in commercial cutting guides. Each cut plane was positioned to favour the anterior humeral head-neck junction while preserving the posterior cuff insertion. The humeral head height and diameter were measured in both the AP plane and the SI plane for the anatomic and guide-assisted osteotomy planes. Differences were compared using separate two-way repeated measures ANOVA for each dependent variable and deviations were shown using box plot and whisker diagrams.

Results

Guide-assisted cuts tend to be smaller than the anatomic humeral head dimensions. Retroversion angle showed a significant effect on head height, AP, and SI diameters (p=0.002). The effect of head-neck angle was only significant for SI diameter (p<0.001). The largest dimensional deviation was observed at 20 degrees of retroversion and resulted in a 2.5mm decrease in humeral head height, averaged over the range of head-neck values.

Conclusion

Where patient's natural anatomy falls outside the range of commercial cutting guides, resection according to the template may result in a deviation from the natural dimensions of the humeral head, which impacts the sizing of the implant head component. This has implications for both manufacturers to create a template that has a larger range of retroversion angles, as well as surgeons' choices in intra-operative planning.